R&S ZNA26 User Manual page 505

Vector network analyzers
Hide thumbs Also See for ZNA26:
Table of Contents

Advertisement

®
R&S
ZNA
Start Cal Sweep
Starts the calibration sweeps for the selected port and power calibration settings and
closes the dialog. The calibration is performed as described in
on page 168.
Open the
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect[:ACQuire]
Calibrate Only Port Frequency
For
Frequency Conversion
brated for all resulting frequencies – even if they are currently not relevant for some
ports. Using this switch, you can limit the calibrated frequencies to those frequencies
that are required for the related port, which results in shorter calibration times.
Power Cal Dialog (Meas. Receiver)
In "Meas. Receiver" mode, the "Power Cal" dialog guides you through the calibration of
a measurement receiver. To ensure accurate source signal powers, a source power
calibration must be performed before the measurement receiver calibration.
Background information
Refer to
Access:
User Manual 1178.6462.02 ─ 16
Pwr Cal Settings Tab
Measurements, by default the reference receivers are cali-
Chapter 5.5.6.2, "Measurement Receiver
Power Cal Wizard
> "Meas. Receiver"
if you wish to modify the calibration procedure.
Calibration",
GUI Reference
Cal Softtool
"Calibration procedure"
on page 169.
505