Move In X-, Y- & Z-Direction - Endress+Hauser Analytik Jena CyBio FeliX Operating Manual

Hide thumbs Also See for Analytik Jena CyBio FeliX:
Table of Contents

Advertisement

Operation 
7.6.4
Move in X-, Y- & Z-Direction
In software-supported operation control mode the pipetting tips can be moved into 
position directly over a plate or reservoir by triggering a "Move to deck position" 
command. 
The Z-axis will be in uppermost position. 
Select the reference point for the Z-axis as follows:
 Topmost vertical position (Oberste Vertikalposition)
 Deck position (Deckposition) or
 Plate adapter position (Plattenadapterposition)
Use a "Move to deck position" command to position the Z-axis in relation to the lab-
ware. Use a "Vertical drive" command to position only the Z-axis at a particular height.
Move to deck position
– Select position referenced to X- and Y-axis
– Z-axis: uppermost vertical position
– Shift selectable for each axis
Fig.
31: Positioning window
Move to labware
– Select position referenced to X-, Y- and Z-axis
– Z-axis: reference point is well bottom or Well top
– Shift can be selected for all axes
Vertical drive
The following points may be set as reference:
Topmost vertical position
Performs downward move starting at absolute zero-point of Z-axis – only negative 
values are available.
Current position
Performs upward move starting in current Z-axis position – negatively and positively 
signed values can be used as input.
Well bottom (Well-Boden)/Well top (Well-Oberkante)
Starting move from the well bottom (or well top) the last approached labware up / 
down. The input of negative / positive values is possible.
® 
1 Valid from CyBio
® 
2 Valid from CyBio
60
Composer software 2.55.
Composer software 2.55.