Serene
User's Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Serene and is the answer not in the manual?

Questions and answers

Summary of Contents for Apex Medical Serene

  • Page 1 Serene User’s Manual...
  • Page 2: Table Of Contents

    CONTENTS ALTERNATING PRESSURE REDISTRIBUTION SYSTEM IMPORTANT SAFEGUARDS GROUNDING INSTRUCTIONS INSTRUCTION PRODUCT DESCRIPTION INSTALLATION OPERATION CLEANING STORAGE MAINTENANCE EXPECTED SERVICE LIFE TROUBLE SHOOTING TECHNICAL SPECIFICATION APPEDNIX A: EMC INFORMATION MODEL NO.: 9P-079040 Please read the manual before use...
  • Page 3: Important Safeguards

    Serene / English IMPORTANT SAFEGUARDS READ ALL INSTRUCTIONS BEFORE OPERATING THIS DEVICE. DANGER - To reduce the risk of electrocution: 1. Always unplug this product immediately after using. 2. Do not use while bathing. 3. Do not place or store this product where it can fall or be pulled into a tub or sink.
  • Page 4: Grounding Instructions

    Serene / English SAVE THESE INSTRUCTIONS GROUNDING INSTRUCTIONS DANGER - Improper use of the grounding plug can result in a risk of electric shock. If repair or replacement of the cord or plug is necessary, do not connect the grounding wire to either flat blade terminal. The wire with insulation having an outer surface that is green with or without yellow stripes is the grounding wire.
  • Page 5 Serene / English SYMBOLS Authorised representative in the European community Manufacturer Protected against solid foreign objects of 12,5 mm and greater; Protection against vertically falling water drops Temperature limitation/temperature range Consult operating instructions for use. "BF" symbol, indicate this product is according to the degree of...
  • Page 6: Instruction

    Serene / English Introduction This manual should be used for initial set up of the system and for reference purposes. 1.1 General Information EN 60601-1 EN 60601-1-2 EN 55011 Class B IEC61000-3-2 IEC 61000-3-3 EMC Warning Statement This equipment has been tested and found to comply with the limits for medical devices to the EN 60601-1-2.
  • Page 7 Serene / English 1.2 Intended Use This product is intended: to help and reduce the incidence of pressure ulcers while optimizing patient ■ comfort. for long term home care of patients suffering from pressure ulcers. ■ for pain management as prescribed by a physician.
  • Page 8: Product Description

    2.1 Pump & Mattress System 1. Quick Connector for CPR 2. Pump Unit 3. Mattress 4. Mattress Foam Pocket NOTE: An underlying mattress such foam must be used with Serene system. 2.2 Pump Unit Front 1. Quick Connector 2. Front Panel Rear 1.
  • Page 9 Serene / English 2.3 Front Panel 1. Comfort Setting The “Comfort Setting” controls the air pressure output. When the firmness is increased ), the output pressure will increase and vice versa ( ) for decreasing air pressure. A hand check is needed to determine if patient is bottoming out. When a...
  • Page 10 Serene / English 2. Therapy Modes A. Static All of the air cells are equally inflated at lower pressures to redistribute body mass over a greater surface area at a constant low pressure. B. Seat Inflate The Seat Inflation features additional supports to the patient during head raise position to prevent bottoming out.
  • Page 11 Serene / English 6. Alarm Mute Press alarm mute button to temporary suspend the alarm. Should the situation not resolved within 3 minutes, the low pressure and tech support alarm shall resume to notify the caregiver. A. Low Pressure Indicator...
  • Page 12: Installation

    WARNING: The Serene mattress must be applied on a underlying mattress or to insert an foam into the pocket of bottoming cover.
  • Page 13 Serene / English 3. Connect the Quick Connector from air mattress to the pump unit. Make sure the connector is in the right position as the inserted diagram below. When a "click" sound is felt or heard, the connection is completed and secured.
  • Page 14: Operation

    Serene / English Operation NOTE: Always read the operating instruction before use. 4.1 General Operation 1. Switch on the main power switch found from the side of the pump and press on the control panel to turn on the power.
  • Page 15: Cleaning

    Serene / English Cleaning It is important to follow the cleaning procedures to avoid cross contamination. Be sure to clean the surface in a dry and dust free environment. Wipe down the pump unit with a damp cloth pre-soaked with a mild detergent. Avoid contact with dust and proximity to dusty areas.
  • Page 16: Storage

    Serene / English Storage 1. To quickly deflate the mattress for storage, click on the release button on the Quick Connector 2. Roll from the foot end towards the head end with CPR valve open, and make sure the tubing is not kinked.
  • Page 17: Maintenance

    Serene / English Maintenance 7.1 General 1. Check main power cord and plug if there are abrasions or excessive wears. 2. Check mattress cover for signs of wear or damage. Ensure mattress cover and tubes are stubbed together correctly. 3. Check the air hoses for any kink or break. For replacement, please contact your local dealers.
  • Page 18: Expected Service Life

    The products are intended to offer safe and reliable operation when use or installed according to the instructions provided by Apex Medical. Apex Medical recommends that the system be inspected and serviced by authorized technicians if there are any signs of wear or concerns with device function and indication on products.
  • Page 19: Trouble Shooting

    Serene / English Trouble Shooting Q.1 Q1 Power is not ON ‧ Check if the plug is connected to mains. ‧ Check for a blown fuse. Q2 Low Pressure Alarm is on ‧ Check if the Quick Connector is tightly secured.
  • Page 20: Technical Specification

    Serene / English Technical specification PUMP Power Supply (Note: See rating label AC 220-240V, 50/60 Hz, 3A on the product) Fuse Rating T3.15AL 250V Dimension (L x W x H) 34 x 16.3 x 23.5 (cm) or 13.4” x 6.4” x 9.3”...
  • Page 21: Appednix A: Emc Information

    Serene / English Appendix A: EMC Information Guidance and Manufacturer’s Declaration- Electromagnetic Emissions: This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment.
  • Page 22 Serene / English at least 30 %. ±2kV for power Mains power quality should Electrical fast supply line ±2kV for power supply line be that of atypical transient/ burst ±1kV ±1kV for input/output line commercial or hospital IEC61000-4-4 input/output environment line ±...
  • Page 23 Serene / English Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an...
  • Page 24 Serene / English Recommended separation distances between portable and mobile RF communications equipment and this device: This device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of this device can help prevent...
  • Page 28 Apex Medical S.L. Elcano 9, 6 planta 48008 Bilbao. Vizcaya. Spain Apex Medical Corp. No.9, Min Sheng St., Tu-Cheng, New Taipei City, 23679, Taiwan Print-2019/All rights reserved 796004-0000 V1.2 www.apexmedicalcorp.com...