Model 4200A-SCS Pulse Card (PGU and PMU) User's Manual
2. Copy the waveforms into the Sequencer for Channel 1, Channel 2, or both:
▪
The order that two or more waveforms appear in a channel sequencer is the order that the
waveforms are output by that channel.
▪
Refer to
details.
3. Save the waveforms in the Sequencer as a .kaf file. See
waveforms into Sequencer
4. Load the .kaf waveform file into a pulse generator (using the appropriate pulse card tab): Refer
to
Custom Arb file operation: Load waveform and turn on output
5. Turn on the output for enabled channels. See
on output
Custom Arb file operation: Select and configure waveforms
To select and configure waveforms:
1. Select the Arb Generator tab.
2. Select New Waveform to open the Waveform Generator dialog box, shown in the following
figure.
4200A-PMU-900-01 Rev. A December 2020
Custom Arb file operation: Copy waveforms into Sequencer
(on page 5-10).
(on page 5-12).
Figure 106: Custom Arb file operation: Select and configure waveforms
Section 5: KPulse (for Keithley Pulse Cards)
Custom Arb file operation: Copy
(on page 5-12) for details.
Custom Arb file operation: Load waveform and turn
(on page 5-10) for
5-9