Overview EBI 300 series data logger is a battery powered temperature measurement and recording instrument for monitoring temperature sensitive products during transportation. They can be programmed indefinitely, but as soon as the measurement is running, they can´t be restarted. The logger has a USB-interface for connecting directly to a computer or USB-printer.
Page 4
1..... Eye 2..... LED 3..... Start/Stop button 4..... USB-connector 5..... Protective cap User manual EBI 330 V 1.1 Page 4 / 16...
Safety instructions Do not use the device in explosive areas. Danger of death! Do not use the device in an environment hotter than 60°C! The lithium battery may explode! Do not expose the unit to microwave radiation. The lithium battery may explode! To ensure the logger is water tight, the protective cap must be fit or an external probe must be connected.
Unpacking / Package Contents After receipt of delivery, please check the equipment for any damage. Furthermore, please verify that the delivery corresponds to your order. Scope of Delivery The delivery includes: 10 EBI 330 temperature data loggers, each with battery, protective cap and a label, on which the device type, serial number, bar code and temperature limits are printed.
Configuration The loggers can be ordered preconfigured. As long as the measurement hasn´t started yet, it is possible to reconfigure it any time. The configuration options are, among others: Measurement interval Upper temperature limit Lower temperature limit Alarm delay Start delay Time zone Language of the PDF-file You can apply these settings by downloading the free of...
Start / Stop The logger is started by pushing the start/stop-button for at least three seconds. CAUTION! The device can only be started once! Depending on the set start delay, the device starts to measure immediately or after a few minutes. It will continue to measure and store temperature data according to the set interval, until the data memory is full (up to 6000 values), the battery empty (lasts for at least 100 days), or the device is...
LEDs The logger has a green and a red LED. When pushing the start/stop button, the LED will flash as long as you keep the button pushed - red for start, green for stop. Start and stop, respectively, are confirmed by triple blinking of the LED.
As soon as the logger is plugged into a USB-slot, it automatically creates a PDF-file, which is saved on a USB- drive, similar to the ones of USB-sticks. The file can easily be copied from there. The file has a lot of information to the measurement – the logger configuration, statistic data to and a chart of the measurement, and possible alarms.
Battery The battery of the EBI 330-T30 loggers lasts for at least 100 days, independently from application temperature and the measurement interval. The battery of the EBI 330-T85 loggers lasts for at least 100 days. In neither logger version you can replace the battery. Cleaning and maintenance Please use a damp cloth to clean the logger.
Technical Data General Accuracy class: Resolution: 0.1° C IP rating: IP 65 Norms: DIN EN 12830 (S, T, C, D, 1), ATP Calibration: batch calibration Size: 80 x 28 x 12 mm Weight: 16 g (incl. battery) EBI 330-T30 Temperature: -30 °C…+60 °C / -22 °F…+140 °F Accuracy: ±...
Page 13
EBI 330-T85 Temperature: -85 °C … +50 °C / -121 °F … +122 °F Accuracy: ± 2.0 °C (-85 °C … -30 °C) ± 0.8° C (-30 °C … -20 °C) ± 0.5° C (-20 °C … +50 °C) Operating temp.: -85 °C …...
Approvals The conformity certificate confirms that this product complies to CE guidelines 2004/108 EG. The product EBI 330 complies to the guidelines in accordance with EN 12830. Suitability: S (Storage), T (Transport) Location: C (Food storage and distribution systems), D (Food Transport Vehicles) Accuracy classification: 1, for measurement of the air temperature...
Need help?
Do you have a question about the Ebro EBI 300 Series and is the answer not in the manual?
Questions and answers