6.2 Device Error Messages

If an error occurs and the LED on the device's cover is blinking, you should take the actions
specified below. If the problem persists you may contact Itamar or an authorized
representative directly.
Device LED's status
No light
Red LED blinks (every 10 seconds)
Red LED blinks (every 10 seconds)
Red LED blinks fast (5 times per
second)
Red LED is on
WatchPAT
ONE System
Possible Reason
registered
No charger plugged into a
mobile phone
The Application could have
been suspended by your
Phone before completion of
the data upload.
The device is not in proximity
or the battery has been
extracted
Possible Reason
Battery depleted, or battery
is placed backward or
hardware error.
The Mobile phone ran out of
power
The Mobile phone and the
device are not close enough
There is a hardware failure in
the device
The device battery is running
out of power
38
Action
Plug in a charger
Open the WatchPAT ONE
Application and follow the guidance
provided in its screens.
Bring the device closer to the phone
or insert the battery
Action
Check the battery. If it is good and
placed properly there is a hardware
error. Return device to Itamar.
Connect a charger, rerun the
Application and keep the device close
by until all the stored data has been
sent to the Application
Make sure the phone and the device
are nearby, until all the stored data
has been sent to the Application
Return the device to the provider,
and a new one will be shipped in
return.
Replace the battery with a fresh
Alkaline AAA battery
Operation Manual

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the WatchPAT ONE and is the answer not in the manual?

Questions and answers

Ken Wiggs
February 7, 2025

Devise software update at 0% for 20 minutes, I have changed the battery three times, my phone is connected to a power source

1 comments:
Mr. Anderson
February 10, 2025

Open the WatchPAT ONE Application and follow the guidance provided on its screens. Ensure the device is in proximity to the phone. If the issue persists, restart the phone and reopen the application. If the problem continues, contact Itamar Medical or an authorized representative for support.

This answer is automatically generated

Frank Krieger
March 1, 2025

I need to reset the pin timer

Table of Contents

Save PDF