Intended Use The P280, MRS (Mattress Replacement System) model with foam base or Overlay model and P280 Air, is a therapeutic mattress designed by Hill-Rom.Its benefits are an efficient automatic redistribution of pressure, especially in the sacrum and heel zones, as part of a protocol for prevention and appropriate treatment of pressure ulcers.
• Two functions: Over-Inflation and Maximum Inflation (Pmax). The P280 Overlay model must be installed on a standard foam mattress, or on the MRS foam base. The MRS foam base can be ordered separately; See “MRS Foam Base” on page 39 to upgrade P280 Overlay into P280 MRS.
(See “Disinfecting and Servicing” on page 31). • Check the safe positioning of the device in its working environment (See “Before Placing the Patient on the Device” page 19). 187692(10) - EN - P280 Mattress Instructions for Use Page 5...
Labels on the Device” on page 15). The device should be connected to a power system equipped with a maximum 30 mA earth leakage circuit breaker, in compliance with IEC 60364-5-53. Page 6 187692(10) - EN - P280 Mattress Instructions for Use...
Protected against light and damp; • At least 10 cm above floor level to prevent fluid ingress; • Protected against dust; • Outside passageways. Never stack more than 5 mattresses. 187692(10) - EN - P280 Mattress Instructions for Use Page 7...
Specifications are subject to change. Essential performances of the device The P280 is a therapeutic mattress. It has two operating modes: continuous low pressure (CLP) and alternating low pressure (ALP). The mattress firmness is adjustable from the control unit.
(IEC 60529) water drops Protection against inflammable Not for use with flammable anesthetics. anesthetic mixtures Front and Back Views of the P280 Control Uni Control Panel Hose Connector Fuses Handle Attachment Main Switch...
Detachable Upper Cover Pneumatic Mattress Heel Zone Foam Bottom Cover Power Cord Hose MRS Foam Base (P280 MRS Mattress Configuration only) Mattress (P280 MRS) or Mattress Topper (P280 Overlay) Bottom Cover Page 10 187692(10) - EN - P280 Mattress Instructions for Use...
Page 17
Item Name: P280 Air Control Unit Detachable Upper Cover Pneumatic Upper Mattress Pneumatic Lower Mattress Bottom Cover Power Cord 187692(10) - EN - P280 Mattress Instructions for Use Page 11...
Do not walk on or run over the power cord Foot End Zone for Notes Top of the Mattress Install on a standard foam type mattress (P280 Overlay only) Adjust the head and foot end straps Page 12 187692(10) - EN - P280 Mattress Instructions for Use...
- P+6 Figures = Model, - A = Device Version Letter. See “Locating the Labels on the Device” on page 15. Page 14 187692(10) - EN - P280 Mattress Instructions for Use...
(A) Conditions of Use and Specifications of the Device P005856A P005987A P006172A P005858A P006052A P006173A (B) Serial Number P005858A/P006052A/ P005856A/P005987A/P006172A P006173A (C) Mandatory Certification only for specific countries See “Symbols on the Labels” on page 14. 187692(10) - EN - P280 Mattress Instructions for Use Page 15...
Accessing to the Upper and Bottom Covers Identification Labels Open the zip fastener on the mattress. Upper Cover P280 Upper Cover P280 MRS Fire Bottom Cover Bottom Cover P280 P280 MRS Page 16 187692(10) - EN - P280 Mattress Instructions for Use...
P280 Air See details of the symbols for cleaning and disinfection in section: “Disinfecting and Servicing” on page 31. Serial number of the mattress P005856A P005858A P005987A P006052A 187692(10) - EN - P280 Mattress Instructions for Use Page 17...
P006172A P006173A Checking the Model of the Device on the Packaging Label P005856A P005858A P005987A P006052A P006173A P006172A Page 18 187692(10) - EN - P280 Mattress Instructions for Use...
IEC 60601-2-52 standard and the ’’Hospital Bed Safety Workgroup’’ guide, and that the resulting combination does not alter the performance of the devices, their safety nor their usability. 187692(10) - EN - P280 Mattress Instructions for Use Page 19...
Take care not to use it directly on the mattress support platform. The P280 MRS mattress, delivered with a foam basis and the P280 Air mattress, are conceived to be used directly on a sleep deck. Be careful to remove the foam mattress, if present, before installing.
Page 27
11. Make sure that the hose cable does not become entangled in the barriers of the bed. Where appropriate, place the cable on the handrail panel footboard. 12. Connect the power cord to the control unit. 187692(10) - EN - P280 Mattress Instructions for Use Page 21...
Page 28
18. Note the position of the CPR valve. The CPR valve must remain accessible for use in the event of emergency deflation. Placing a cotton sheet on the mattress will improve patient comfort and make it easier to administer care. Page 22 187692(10) - EN - P280 Mattress Instructions for Use...
The pressure in the air bladders of the device can be adjusted according to the condition of the patient in all the operating modes using the comfort settings buttons: • To decrease the pressure • To increase the pressure 187692(10) - EN - P280 Mattress Instructions for Use Page 23...
Page 30
Maximum Inflation (P-Max) When connected to the mains power supply for the first time, the P280 automatically inflates to P-Max and the indicator light near the symbol comes on.
1. Connect the hose connector to the control unit. The device returns to the previous comfort settings. 2. If necessary, place the bed frame in the appropriate position (refer to the instructions for the bed frame). 187692(10) - EN - P280 Mattress Instructions for Use Page 25...
If the problem continues, the audible indicator sounds again after 3 minutes. Mains Power Fault Indicator If the P280 is disconnected or if there is a mains power failure, an audible indicator sounds and the light near the symbol comes on.
If necessary, use the attachments provided with the device. 6. Safely stow the power cord. 187692(10) - EN - P280 Mattress Instructions for Use Page 27...
Page 34
The CPR function remains available in transport mode. If the hose is connected, simply disconnect it from the control unit. Otherwise, remove the transport cap to allow the air to escape. Page 28 187692(10) - EN - P280 Mattress Instructions for Use...
6. Transfer the mattress on to the other bed. In the case of P280 Overlay model, if there is no mattress on the other bed, transfer the base mattress with the device. 7. To install the mattress on the other bed, refer to steps 2 through 13 of the Installation procedure (page 20).
Never fold the mattress lengthways in order to conserve the life time of the footrest foam. 10. Store the mattress in its transport bag (See “Complying with Conditions for Transport, Storage and Use” page 7). Page 30 187692(10) - EN - P280 Mattress Instructions for Use...
The following instructions are not designed to replace more appropriate cleaning and disinfection protocols drawn up by the hygiene officer or by other bodies for your hospital in the event of particular infectious situations. 187692(10) - EN - P280 Mattress Instructions for Use Page 31...
Do not use highly acidic detergents or disinfectants (pH<4). Do not use highly basic detergents or disinfectants (pH>10). Never use abrasive cleaning products or materials, such as scouring pads or powders. Page 32 187692(10) - EN - P280 Mattress Instructions for Use...
In the event of visible soiling, Hill-Rom recommends that the mattress should be disinfected with an intermediate disinfectant (tuberculocidal) that complies with the applicable regulations (e.g. with the requirements of directive 93/42/EEC). 187692(10) - EN - P280 Mattress Instructions for Use Page 33...
• Do not brush and apply reduced pressure to labels and markings. • Dry carefully and look for signs of water ingress. • Test the device before reusing it. Page 34 187692(10) - EN - P280 Mattress Instructions for Use...
Always remove all traces of the products used when washing or disinfecting the device. • Make sure that all the parts of the device are perfectly dry before installing to avoid any risk of condensation forming inside. 187692(10) - EN - P280 Mattress Instructions for Use Page 35...
The expected lifetime of the device in normal conditions of use, maintenance and service is 5 years and 2 years for the cover. Please refer to the service manual for more information on lifetime of wear parts and accessories. Page 36 187692(10) - EN - P280 Mattress Instructions for Use...
The air filter can be ordered separately. For more information, refer to the Spare Parts List of the P280 Mattress. If the problem is not resolved, contact your local Hill-Rom representative (see back of manual). Indicate the serial number of the device, which can be found on the control unit, when requesting repairs.
(as per Directive 2006/96/EC). Refer to the P280 Mattress Service Manual for detailed battery replacement instructions. The device is designed for easy dismantling so that it can be destroyed or reused in accordance with the applicable recycling regulations (e.g.
Seat Cushion (M04177)** ™ The optional seat cushion is an accessory that can be inflated using the P280 Reflex adaptor. In combination with the air mattresses, the ReFlex™ seat cushion provide 24- hour pressure relieving care and help to prevent pressure ulcers for semi-mobile patients up to high.
Medical Device conformity mark Medical Device Cushion Inflate Cushion strapping Locate the seat cushion labels On the seat cushion cover. Open the zip fastener of the device. Page 40 187692(10) - EN - P280 Mattress Instructions for Use...
The device automatically inflates to the maximum (P-max) so as to fully develop the air cushions. When he returns to the default mode "Alternating Pressure Low", the cushion is ready. 187692(10) - EN - P280 Mattress Instructions for Use Page 41...
All the parts can be ordered separately. Refer to the Spare Parts List for the product part numbers. Contact your local Hill-Rom representative or your local distributor (see back of manual). Page 42 187692(10) - EN - P280 Mattress Instructions for Use...
Electromagnetic Compatibility Compliance CE conformity mark The CE conformity mark applicable to class IIa Medical Devices was applied to the P280 mattress for the first time in 2011. Standards Name EN 60601-1 (2006) & A1 (2013) IEC 60601-1: 2005 & A1 (2012),...
Manufacturer's recommendations and declaration – electromagnetic emissions The P280 Mattress is intended for use in the electromagnetic environment specified below. The customer or the user of the P280 should make sure it is used in such an environment. Emission test...
Page 51
Manufacturer's recommendations and declaration – electromagnetic immunity The P280 Mattress is intended for use in the electromagnetic environment specified below. The customer or the user of the P280 Mattress should make sure it is used in such an environment. Immunity Test...
Page 52
FRS 460 (limited to 0,6 W ERP*), PMR 446 UMTS, DECT (limited to 0.25 W ERP*) Bluetooth, WLAN 802.11 b/g 2450 , RFID 2450 (limited to 0.1 W EPR*) *:Effective Radiated Power Page 46 187692(10) - EN - P280 Mattress Instructions for Use...
Page 53
The P280 device is designed for use in an electromagnetic environment in which interference due to radiated RF is monitored. The customer or user of the P280 device can contribute to the prevention of electromagnetic interference by keeping the P280 device at the recommended distances from portable and mobile RF equipment (transmitters) as shown below, according to the maximum power output of the communication equipment.
Page 54
Page 48 187692(10) - EN - P280 Mattress Instructions for Use...