DenTal CHaIr DIagnosTIC CoDes

InPuT
sTaTus
DIgIT
0
No input received
1
Foot switch button pressed
2
Touch pad button pressed
Foot switch and touch pad button pressed
3
Chair stopped due to loss of input signal
4
OUTPUT STATUS DIGIT
0
1
2
3
4
5
6
7
8
DIAGNOSTIC DISPLAY
00
FF
E_
RICHT DECIMAL
LEFT DECIMAL
MaIn ConTrol boarD
DIagnosTIC DIsPlaY
8
InPuT sTaTus
OUTPUT STATUS
No motion in progress
Backrest in motion
Base in motion
Backrest and loose in motion
Chair stopped due to no sensed motion
Backrest input received, out of soft limits
Base input received, out of soft limits
Backrest and base input received,
out of soft limits
Auto position in progress
OPERATING STATUS
Potentiometer value too low
Potentiometer value too high
Main control board failure (replace)
Error is for base potentiometer
Error is for back rest potentiometer
047657 r7
CorreCTIVe aCTIon
Examine input device
Switch may be stuck
Switch maybe stuck
Multiples switches stuck
Input device or connection faulty
CORRECTIVE ACTION
No output present
Normal operation
Normal operation
Normal operation
Examine potentiometer
Reprogram soft limits
Reprogram soft limits
Reprogram soft limits
Normal operation
CORRECTIVE ACTION
Calibrate potentiometer
Calibrate potentiometer
Replace board
Examine potentiometer
Examine potentiometer