Puritan Bennett 840 Service Manual

Puritan Bennett 840 Service Manual

Table of Contents

Advertisement

Quick Links

840
Service Manual
V e n t i l a t o r S y s t e m
4-070496-00 Rev K
August 2010

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 840 and is the answer not in the manual?

Questions and answers

GUSSIE WHITE
March 19, 2025

HOWTODISABLEO2SENSORONVENT840

1 comments:
Mr. Anderson
March 19, 2025

The context mentions that the oxygen (O2) sensor can be disabled in the settings of the Puritan Bennett 840 ventilator. To disable the O2 sensor, access the ventilator's settings and look for the option to disable the oxygen sensor. If the sensor is faulty, it may also be disabled automatically.

This answer is automatically generated

GUSSIE WHITE
March 19, 2025

DISABLE O2 SENSOR

1 comments:
Mr. Anderson
March 19, 2025

To disable the O₂ sensor on a Puritan Bennett 840 ventilator, go to the lower touch screen and navigate to OTHER SCREENS -> MORE SETTINGS. From there, disable the oxygen monitoring feature and press ACCEPT.

This answer is automatically generated

CALIBRACIÓN
March 20, 2025

Hello, what are the inputs required for the use of the ventilator?

1 comments:
Mr. Anderson
March 20, 2025

The Puritan Bennett 840 ventilator requires the following inputs:

1. Power Supply – It operates on AC mains power (100 V, 120 V, or 220–240 V, depending on the model) and has a Battery Power Source (BPS) for backup.
2. Compressed Air Source – It can use an external air supply (wall or bottled air) or an optional compressor unit.
3. Patient Circuit Components – This includes the inspiratory and expiratory limbs, a humidification device, a collector vial, and an expiratory filter.
4. Software Control – The ventilator software manages exhalation valve positioning and monitors pressures throughout the respiratory cycle.

These inputs ensure proper ventilation function and monitoring.

This answer is automatically generated

Summary of Contents for Puritan Bennett 840

  • Page 1 Service Manual V e n t i l a t o r S y s t e m 4-070496-00 Rev K August 2010...
  • Page 3 Nothing in this manual shall limit or restrict in any way Puritan Bennett’s right to revise or otherwise change or modify the equipment (including its software) described herein, without notice. In the absence of an express, written agreement to the contrary, Puritan Bennett has no obligation to furnish any such revisions, changes, or modifications to the owner or user of the equipment (including its software) described herein.
  • Page 4 In case of fire or a burning smell, immediately disconnect the ventilator from the oxygen supply, facility power, and BPS. • When handling any part of the Puritan Bennett 840 Ventilator System, always follow your hospital infection control guidelines for handling infectious material.
  • Page 5 • Patients on life-support equipment should be appropriately monitored by competent medical personnel and suitable monitoring devices. • The Puritan Bennett 840 Ventilator System is not intended to be a comprehensive monitoring device and does not activate alarms for all types of dangerous conditions for patients on life-support equipment.
  • Page 6 While this manual covers the ventilator configurations currently supported by Puritan Bennett, it may not be all-inclusive and may not be applicable to your ventilator. You should ensure you have the most current applicable version of this manual; if in doubt, contact Puritan Bennett or visit the Puritan Bennett product manual Web page at http://www.puritanbennett.com/serv/manuals.aspx...
  • Page 7: Table Of Contents

    2.3.3 Exhalation module..................... 2-38 2.3.3.1 Exhalation module components ............. 2-39 2.3.3.2 Exhalation module operation ..............2-40 2.3.4 806 Compressor Unit ..................2-43 2.3.4.1 806 Compressor unit components ............2-43 2.3.4.2 806 Compressor unit operation.............. 2-46 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 8 2.6.3 Pressure transducer autozero................2-115 2.6.4 Power monitoring and power fail handling ............. 2-117 2.6.4.1 Loss of power source ................2-117 2.6.4.2 Supply voltage monitoring ..............2-118 2.7.1 Safety valve open (SVO) state................2-118 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 9 4.2.10.2 External Test Control: Performing remote ventilator testing ..... 4-9 4.2.10.3 Exp Valve Calibration ................4-9 4.2.10.4 Vent Inop Test..................4-12 4.2.10.5 Flow Sensor Calibration................4-13 4.2.10.6 Atmospheric Pressure Transducer calibration.......... 4-15 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 10 6.12 How to troubleshoot GUI LCD screen messages ........... 6-109 6.12.1 Example 1: Missing data key information ............6-109 6.12.2 Example 2: No communication between the GUI and BDU......6-111 6.12.3 Example 3: POST or Background check errors ..........6-112 viii Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 11 8.13 Repairing the 10.4-inch GUI ..................8-8 8.13.1 Removing or installing the 10.4-inch GUI when mounted on RTA cart....8-8 8.13.2 Removing or installing the 10.4-inch GUI when mounted on Puritan Bennett 800 Series Ventilator Compressor Mount Cart or Puritan Bennett 800 Series Ventilator Pole Cart................
  • Page 12 8.15 Breath delivery unit (BDU) ..................8-50 8.15.1 Removing BDU from RTA cart ................. 8-53 8.15.2 Removing BDU from Puritan Bennett 800 Series Ventilator Compressor Mount Cart or Puritan Bennett 800 Series Ventilator Pole Cart..8-53 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 13 8.15.4 Installing BDU onto RTA cart ................8-55 8.15.5 Installing BDU onto Puritan Bennett 800 Series Ventilator Compressor Mount Cart or Puritan Bennett 800 Series Ventilator Pole Cart ..8-55 8.15.6 Analog interface (AI) PCB and breath delivery (BD) CPU PCB......8-56 8.15.6.1 Removing AI PCB or BD CPU PCB ............
  • Page 14 8.17.17 Compressor assembly ................. 8-121 8.17.17.1 Removing the compressor assembly ............ 8-121 8.17.17.2 Replacing the coalescing filter element ..........8-122 8.17.18 Replacing the compressor panels ..............8-123 8.17.19 Reinstalling the plenum assembly..............8-124 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 15 8.19.4.2 Installing casters on RTA cart ..............8-151 8.19.5 Removing casters from Puritan Bennett 800 Series Ventilator Compressor Mount Cart or Puritan Bennett 800 Series Ventilator Pole Cart ... 8-151 8.19.6 Installing casters on Puritan Bennett 800 Series Ventilator Compressor Mount Cart or Puritan Bennett 800 Series Ventilator Pole Cart ..8-152 8.19.7 Removing/installing GUI mount on all carts...........
  • Page 16 Contents Parts list 9.1 How to use this parts list ..................... 9-1 9.2 Puritan Bennett 840 Ventilator System patient system and accessories ....... 9-3 9.2.1 Puritan Bennett 840 Ventilator System NeoMode patient system and accessories ....................9-6 9.3 Flex arm assembly, oxygen and air hose assemblies, power cords ......9-8 9.3.1 Flex arm assembly ....................
  • Page 17 RTA cart, Puritan Bennett 800 Series Ventilator Compressor Mount Cart, and Puritan Bennett 800 Series Ventilator Pole Cart ......2-6 Figure 2-8.
  • Page 18 AC panel ............. 2-63 Figure 2-47. Puritan Bennett 840 Ventilator System interconnect diagram – AC panel ... . 2-64 Figure 2-48.
  • Page 19 /CAL In Progress indicator......... 4-20 Figure 5-1. Puritan Bennett 840 ventilator side view — PTS 2000 port ......5-9 Figure 5-2.
  • Page 20 Adding ferrites to air and oxygen transducer harnesses ......8-66 xviii Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 21 Removing existing brackets ..........8-128 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 22 Bennett 800 Series Ventilator Pole Cart ........8-154 Figure 9-1. Puritan Bennett 840 Ventilator System patient system and accessories ....9-5 Figure 9-2.
  • Page 23 Puritan Bennett 800 Series Ventilator Pole Cart ....... . 9-92...
  • Page 24 Figures This page is intentionally blank. xxii Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 25 Recommended separation distances between portable and mobile RF communications equipment and the Puritan Bennett 840 Ventilator System ......1-14 Table 1-7.
  • Page 26 Table 6-2. Puritan Bennett 840 Ventilator System diagnostic codes ......6-7 Table 6-3.
  • Page 27: General Information

    BPS, which is located in the BDU shelf. The Puritan Bennett 800 Series Ventilator Pole Cart can also be used with a one-hour battery or an optional four-hour battery, located in the cart base. The Puritan Bennett 800 Series Ventilator Pole Cart cannot be used with a compressor.
  • Page 28: Configuration Information

    An external air source is required when performing service mode calibrations and performance verification testing. 1.3 Configuration information The Puritan Bennett 840 Ventilator System is available in a variety of versions, intended to meet differing needs and regulations throughout the world. The major differences in configuration among ventilators are listed below: Electrical requirements: Available in 100 V, 50/60 Hz;...
  • Page 29: Accessories

    “Part numbers” appendix of the Puritan Bennett 800 Series Ventilator System Operator’s and Technical Reference Manual for patient circuit ordering information. Humidification device: The Puritan Bennett 840 Ventilator System supports the use of an optional humidification device, including a heated humidifier, heat and moisture exchanger (HME), or heated wire.
  • Page 30: Specifications

    (41 in. high x 27 in. wide x 33 in. deep with wheels in outermost position) Puritan Bennett 800 Series Ventilator Pole Cart: 1041mm high x 686 mm wide x 839 mm deep (41 in. high x 27 in. wide x 33 in. deep with wheels in outermost position) 804 Compressor: 417 mm high x 458 mm wide x 362 mm deep (16.4 in.
  • Page 31 , 50 cmH O; P , 3 cmH O. Input power specifications are for ventilators with Fisher & Paykel MR730 MEAN SENS humidifiers. (Humidifier connection only available on 100 – 120 V ventilators.) Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 32 Every 2 months when storage temperature is 41 to 50 C (105 to122 F) NOTE: BPS battery life specifications are approximate. To ensure maximum battery life, maintain full charge and minimize the number of complete discharges. Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 33: Figure 1-1. Remote Alarm (Nurse's Call) Port Pinout (View From Back Of Gui)

    The remote alarm port is a 4-pin female connector. Allowable current is 500 mA at 30 V DC (maximum). Signal Normally closed (NC) Relay common Normally open (NO) Not connected Figure 1-1. Remote alarm (nurse’s call) port pinout (view from back of GUI) Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 34: Figure 1-2. Puritan Bennett 840 Ventilator System Rs-232 Serial Port Pinout

    RS-232 (serial) port (Figure 1-2). A 9-pin male connector configured as data terminal equipment capabilities (cont) (DTE). Allowable current is 0.2 A at 10 V DC (maximum). Figure 1-2. Puritan Bennett 840 Ventilator System RS-232 serial port pinout Signal Not connected...
  • Page 35: Compliance And Approvals

    General information 1.6 Compliance and approvals The Puritan Bennett 840 Ventilator System was developed in accordance with pertinent North American and International standards (Table 1-2). The manufacturing facility for this product is certified to EN ISO 13485:2000 (ISO 13485:1996) Quality Systems - Medical Devices - Particular Requirements for the Application of ISO 9001:1994.
  • Page 36 220 – 240 V, 60 Hz EN 60601-1:1990 EN 60601-1 Amendment 1:1993 EN 60601-1 Amendment 11:1993 EN 60601-1 Amendment 12:1993 EN 60601-1 Amendment 2:1995 EN 60601-1 Amendment 13:1996 IEC 60601-2-12:2001 EN 60601-1-2:2001+A1:2006 Manufacturer self- certification 1-10 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 37: Manufacturer's Declaration

    Puritan Bennett 800 Series Ventilator System Operator’s and Technical Reference Manual. Warning • The Puritan Bennett 840 Ventilator System should not be used adjacent to or stacked with other equipment, except as specified in this manual and the Puritan Bennett 800 Series Ventilator System Operator’s and Technical Reference Manual.
  • Page 38: Table 1-4. Electromagnetic Immunity

    Table 1-4: Electromagnetic Immunity The Puritan Bennett 840 Ventilator System is intended for use in the electromagnetic environment specified below. The customer or the user of the Puritan Bennett 840 Ventilator System should ensure it is used in such an environment.
  • Page 39: Table 1-5. Electromagnetic Immunity - Conducted And Radiated Rf

    Table 1-5: Electromagnetic immunity – conducted and radiated RF The Puritan Bennett 840 Ventilator System is intended for use in the electromagnetic environment specified below. The customer or the user of the Puritan Bennett 840 Ventilator System should ensure it is used in such an environment. Immunity test...
  • Page 40: Table 1-6. Recommended Separation Distances Between Portable And Mobile Rf Communications Equipment And The Puritan Bennett 840 Ventilator System

    If the measured field strength in the location in which the Puritan Bennett 840 Ventilator System is used exceeds the applicable RF compliance level above, the Puritan Bennett 840 Ventilator System should be observed to ver- Puritan Ben- ify normal operation.
  • Page 41 General information Table 1-6: Recommended separation distances between portable and mobile RF communications equipment and the Puritan Bennett 840 Ventilator System (continued) For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
  • Page 42: Table 1-7. Compliant Cables

    General information Table 1-7: Compliant cables Puritan Bennett does not supply remote alarm (nurse call) or serial port cables. In order to maintain compliance to International Electromagnetic Compatibility (EMC) standards, Puritan Bennett recommends using shielded cables for these applications. Warning...
  • Page 43: Technical Information

    General information 1.7 Technical information Refer to Table 1-8 for Puritan Bennett 840 Ventilator System miscellaneous technical information. NOTE: When pressure units are set to hPa, pressure delivery and spirometry are subject to an additional 2% error. Table 1-8: Technical information Maximum limited pressure 127.5 cmH...
  • Page 44 Inspiratory pneumatics: 50 mL 5 mL Expiratory pneumatics: 1000 mL 25 mL (including expiratory filter and collector vial) The Puritan Bennett 840 Ventilator System automatically adjusts for volume losses due to gas compressibility (that is, automatic compliance compensation), subject to a maximum delivered volume of 2500 mL.
  • Page 45: Range, Resolution, Accuracy, And New Patient/Default Settings

    Pediatric: 15 s Adult: 20 s Apnea mandatory type Range: VC or PC Resolution: N/A Accuracy: N/A New patient: PC with NEONATAL patient circuit VC with PEDIATRIC or ADULT patient circuit 1-19 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 46 New patient: Maximum of 3* mL or (7.25 x IBW); with NEONATAL patient circuit (7.25 x IBW); with PEDIATRIC or ADULT patient circuit *Assumes NeoMode 2.0 software option is installed Normal (non-apnea) ventilation 1-20 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 47 Range: HME, non-heated expiratory tube, or heated expiratory tube Resolution: Not applicable Accuracy: Not applicable New patient value: previous setting Humidifier volume Range: 100 mL to 1000 mL Resolution: 10 mL New patient value: previous setting 1-21 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 48 VC with PEDIATRIC or ADULT patient circuit Mode Range: A/C, SIMV, SPONT, or BILEVEL Resolution: Not applicable Accuracy: Not applicable New patient: SIMV with NEONATAL patient circuit A/C with PEDIATRIC or ADULT patient circuit 1-22 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 49 Accuracy: 0.1 + 0.6% of setting)/min averaged over 60 seconds or 5 breaths, whichever occurs last New patient: 20/min with NEONATAL patient circuit 14/min with PEDIATRIC patient circuit 10/min with ADULT patient circuit 1-23 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 50 *Assumes NeoMode 2.0 software option is installed Trigger type Range: Flow (V-TRIG) NEONATAL patient circuit Pressure or flow (P-TRIG or V-TRIG) with PEDIATRIC or ADULT patient circuit Resolution: Not applicable Accuracy: Not applicable New patient: Flow (V-TRIG) 1-24 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 51 SUPP MEAN is rejected. Refer to the Technical Reference section of part of the Puritan Bennett 800 Series Ventilator System Operator’s and Technical Reference Manual for more details. Vent type Range: INVASIVE or NIV (non-invasive)
  • Page 52 Resolution: 1 mL for 1 to 100 mL; 5 mL for 100 to 400 mL; 10 mL for 400 to 2500 mL New patient: 7.25 x IBW x 0.70 when vent type is INVASIVE; OFF when vent type is NIV 1-26 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 53 O for PEEP  20 cmH 1.0 cmH Sensor Range: Enabled, Disabled, or Calibration Resolution: Not applicable New patient value: Enabled NOTE: The O sensor must be for the High/Low delivered O Enabled alarm to function. 1-27 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 54 O for 10 to 130 cmH MEAN (formerly P CIRC Accuracy: ± (3 + 4% of reading) cmH O relative to pressure measured at the exhalation side of the patient wye. 1-28 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 55: Tools, Equipment, And Service Materials

    1.9 Tools, equipment, and service materials The tools, equipment, and service materials listed in Table 1-10 are used to service the Puritan Bennett 840 Ventilator System. Refer to Chapter 5 for a list of required tools, equipment, and service materials specific to performance verification.
  • Page 56: Table 1-10. Tools, Equipment, And Service Materials

    Digital multimeter patch cords G-061567-00 Performance verification (optional) * These parts may be purchased individually or by ordering a Puritan Bennett 840 VTS Accessories Kit, P/N 4-076599-00. Disinfectant/cleaner The following solutions are acceptable General cleaning for disinfecting/cleaning the Puritan...
  • Page 57 Filter, expiratory bacteria filter DAR 351P19005 Performance verification * These parts may be purchased individually or by ordering a Puritan Bennett 840 VTS Accessories Kit, P/N 4-076599-00. Filter, expiratory bacteria filter, 4-070305-00 Performance verification reusable (Re/X800) Filter, inspiratory, Re/Flex...
  • Page 58 • One available TCP/IP network port with BNC connector (if software download desired) * These parts may be purchased individually or by ordering a Puritan Bennett 840 VTS Accessories Kit, P/N 4-076599-00. Pneumatic calibration analyzer or Puritan Bennett PTS 2000 Performance...
  • Page 59 Test lung, 4 L 4-075578-00 Performance verification Tie wraps 4-000003-00 (small) Various places 4-000004-00 (large) * These parts may be purchased individually or by ordering a Puritan Bennett 840 VTS Accessories Kit, P/N 4-076599-00. 1-33 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 60: Series Ventilator Compressor Mount Cart, And Puritan Bennett 800 Series Ventilator Pole Cart

    N-4800168 Replacing casters (RTA cart) Wye, patient circuit *4-000338-00 SST, performance verification * These parts may be purchased individually or by ordering a Puritan Bennett 840 VTS Accessories Kit, P/N 4-076599-00. 1-34 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 61: Periodic Maintenance

    Table 1-11 lists the periodic maintenance activities required for the Puritan Bennett 840 Ventilator System. See the Ventilator Information screen for total hours of operation for the ventilator and compressor. For details on patient system maintenance, refer to the Puritan Bennett 800 Series Ventilator System Operator’s and Technical Reference Manual.
  • Page 62 Use appropriate preventive maintenance kit (see Table 1-12). Every 15,000 hours NOTE: If any part found in a preventive maintenance kit requires replacement before the recommended interval elapses, consider installing the entire kit anyway. 1-36 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 63: Service Kits

    General information 1.11 Service kits Table 1-12 lists the Puritan Bennett 840 Ventilator System service kits. Chapter 9 lists the mounting kits available for the ventilator. Table 1-12: Service kits Interval Description Part no. Kit contents 10,000 hours 9.4“ GUI and BDU...
  • Page 64 Filter element for water trap assembly Main inlet filter Outlet filter for solenoid assembly Tie wrap, large, for air dryer assembly Tie wraps, small for fan harnesses (qty. 2) Tinnerman clips (qty. 6) 1-38 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 65: Controls And Indicators

    1.12 Controls and indicators Refer to Figure 1-3 through Figure 1-10 and Table 1-13 through Table 1-20 for ventilator controls and indicators. 10 11 Figure 1-3. Monochrome GUI front view (showing all keys) 1-39 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 66: Table 1-13. Gui Front View

    Color GUI (9.4 inch screen): Key present but inactive Monochrome GUI: Display brightness key. Allows you to adjust screen brightness when you hold down this key while turning the knob. (US version only) 1-40 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 67 You cannot reset a DEVICE ALERT alarm. Displays basic operating information about the ventilator. US version only 1-41 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 68 0.5 and 2 seconds. Pressing and holding this key initiates a manual inspiratory pause extending inspiration up to 7 seconds. Cancels a proposed setting. 1-42 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 69 Gray normal ventilator operation indicator. Indicator appears unilluminated when no ventilator inoperative condition exists. non-US version US version text is not visible when no ventilator inoperative condition exists. US version 1-43 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 70 Red safety valve open (SVO) indicator. Illuminates when the ventilator has entered its safe state and opened its safety valve to allow the patient to breathe unassisted from room air. non-US version US version 1-44 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 71 (battery symbol) is lit, ventilator is operating on BPS, and AC power is insufficient to support ventilator operation. During BPS operation, power to the compressor unit and the humidifier outlet (if available) is off. BATTERY US version only 1-45 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 72 Green normal ventilator operation indicator steadily lit. This indicator is off if the ventilator is not in a ventilation mode, for example, during service mode or short self test (SST). 1-46 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 73: Figure 1-4. 10.4-Inch Gui Rear View

    10.4-inch GUI only: Two serial ports with 9-pin male connector configured as data terminal equipment (DTE). NOTE: Allowable current is 0.2 A at 10 V DC (maximum). RS-232 NULL Null modem port (9.4-inch and 10.4-inch GUI) 1-47 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 74: Figure 1-5. Bdu Front View

    Index Labeling Function (Figure 1-5) EXHAUST port BDU exhaust port. Gas is vented to atmosphere. Ventilator operation indicators Red ventilator inoperative indicator. (See Table 1-13, item 15.) non-US version US version 1-48 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 75 Gray normal GUI indicator. Indicator appears unilluminated when no loss of GUI condition exists. non-US version US version text is not visible when no loss of GUI condition exists. US version 1-49 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 76 2.3 A (270 VA) with a maximum leakage current of 50 A. To patient port Ventilator outlet From patient port Expiratory limb connector on exhalation filter (Collector vial drain port) Collector vial drain port. Use to attach drainage bag. 1-50 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 77: Figure 1-6. Bdu I/O Panel

    BDU, the ventilator is placed into service mode (for example, to run EST). PTS 2000 Puritan Bennett PTS 2000 Performance Test System connection. Data key connection Caution Do not remove the data key. The data key cover can only be removed with a screwdriver.
  • Page 78 General information Table 1-16: BDU I/O panel (continued) Index Labeling Function (Figure 1-6) Compressor data cable connection Compressor US version GUI cable connection Display (GUI) US version 1-52 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 79: Figure 1-7. Bdu Right-Side Panel

    Labeling Function (Figure 1-7) Ventilator circuit breaker for compressor and humidifier NOTE: A humidifier connection is only available on 100 – 120 V ventilators. Compressor & humidifier circuit breaker US version 1-53 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 80 A common grounding point for the entire ventilator. Maximum allowed output to auxiliary mains socket (compressor electrical connection) Compressor outlet: 5.6 A max US version 1-54 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 81: Figure 1-8. Bdu Rear View

    (High-pressure air fitting) DISS male, DISS female, NIST, Air Liquide, or SIS fitting (High-pressure oxygen DISS male, DISS female, NIST, Air Liquide, or SIS fitting fitting) Figure 1-9. GUI rear view 1-55 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 82: Table 1-19. Gui Rear View

    Labeling Function (Figure 1-9) (Warning label) Warns user of hazards associated with the operation of the Puritan Bennett 840 Ventilator System and GUI (Serial number label) Unique assigned number. Must be the same as the GUI serial number stored on the data key.
  • Page 83: Figure 1-10. 802 Bps Controls And Indicators

    General information Figure 1-10. 802 BPS controls and indicators Figure 1-11. 803 BPS controls and indicators 1-57 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 84: Figure 1-12. Puritan Bennett 800 Series Ventilator Compressor Mount Cart Charging Status And

    General information Figure 1-12. Puritan Bennett 800 Series Ventilator Compressor Mount Cart charging status and battery indicator labels 1-58 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 85: Figure 1-13. Puritan Bennett 800 Series Ventilator Pole Cart Charging Status And

    General information Figure 1-13. Puritan Bennett 800 Series Ventilator Pole Cart charging status and battery indicator labels Table 1-20: BPS controls and indicators Labeling Function Index (Figure 1-10) 802 BPS charging status. When the ventilator is operating on mains power, the top symbol (green LED next to gray battery icon) indicates that the BPS is charged, and the bottom symbol (yellow LED next to gray battery icon) indicates that the BPS is charging.
  • Page 86 All countries Index (Figure 1-12) Charging status on Puritan Bennett 800 Series Ventilator Compressor Mount Cart: Indicates the charging status of the BPS. A yellow LED next to the partially full battery icon indicates the battery is charging. A green LED next to the full battery icon indicates the battery is charged.
  • Page 87: Onscreen Symbols And Abbreviations

    Manual to interpret these. 1.14 Ventilator serial numbers and software version The Puritan Bennett 840 Ventilator System serial numbers and software versions can be displayed on the GUI upper subscreen. On the GUI upper screen, select the VENT CONFIG button. The ventilator configuration subscreen displays the current software revisions for the BDU, GUI, compressor, and audible alarm subsystem.
  • Page 88 General information This page intentionally blank. 1-62 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 89: Theory Of Operation

    ECTION H A PT ER Theory of operation This chapter details the operational theory of the Puritan Bennett™ 840 Ventilator System and contains the following information: • description of major ventilator assemblies • overview of ventilator operation • description of the pneumatic system •...
  • Page 90: Breath Delivery Unit (Bdu)

    Ventilator Compressor Mount Cart, and Puritan Bennett 800 Series Ventilator Pole Cart 2.1.1 Breath delivery unit (BDU) The BDU, shown in Figure 2-2, is the core of the Puritan Bennett 840 Ventilator System. Its pneumatic system, under control of the breath delivery (BD) central processing unit (CPU), mixes oxygen and air and controls gas flow to the patient.
  • Page 91: Graphic User Interface (Gui)

    GUI. The GUI CPU monitors BDU communications. In turn, the GUI displays patient and ventilator performance information. Figure 2-3. 10.4-inch GUI Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 92: Compressor Unit

    803 BPS will operate the ventilator for up to four hours with new, fully charged batteries. Neither BPS supplies the compressor unit or the humidifier with electrical power. The ventilator automatically switches back to AC power when facility power returns within the required limits. Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 93: Cart

    The optional carts are shown in Figure 2-7. The cart mounts system components and accessories, including the compressor unit on the RTA and Puritan Bennett 800 Series Ventilator Compressor Mount Carts. They also provide mobility for the ventilator. Brakes on the front casters prevent the cart from rolling and turning.
  • Page 94: Patient System

    Theory of operation 8-01119 Figure 2-7. RTA cart, Puritan Bennett 800 Series Ventilator Compressor Mount Cart, and Puritan Bennett 800 Series Ventilator Pole Cart 2.1.6 Patient system The patient system (adult, reusable version shown in Figure 2-8) includes reusable or single- patient use tubing plus an inspiratory filter that prevents gas-borne particles from exiting the ventilator.
  • Page 95: Operational Overview

    GUI: high circuit pressure limit, breath phase, breath mode, autozero offsets, inspiratory time, apnea interval, target pressure for pressure controlled breaths, breath phase start, and time stamp. Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 96: Figure 2-10. Puritan Bennett 840 Ventilator System Block Diagram

    Figure 2-10. Puritan Bennett 840 Ventilator System block diagram The GUI logs an event in the diagnostic log and declares a ventilator inoperative condition if: • Any raw signal data from three BDU transmissions within 24 hours is corrupted.
  • Page 97 BDU CPU to control the PSOLs. As a result, the ventilator supplies mixed breathing gas to the patient according to operator-set variables. The mixed air and oxygen passes through the patient circuit external to the ventilator. Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 98 In case of a malfunction that prevents software from opening the safety valve, there is also an analog circuit that opens the safety valve if system pressure exceeds 100 to 120 cmH 2-10 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 99: Pneumatic System

    Theory of operation 2.3 Pneumatic system This section describes the Puritan Bennett 840 Ventilator System pneumatics, as follows: • Inspiratory module: Contains the following pneumatic subsystems: – Gas supply conditioning subsystem – Flow control subsystem – Safety valve/inspiration monitoring subsystem •...
  • Page 100: Figure 2-12. Pneumatic System Diagram

    Theory of operation Figure 2-12. Pneumatic system diagram 2-12 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 101: Table 2-1. Pneumatic Component Descriptions

    Measures oxygen or air flow before PSOL. Regulator, oxygen/air Reduces input supply pressure (35 to 100 psig, flow up to 200 L/min REG1 BTPS) to output pressure (9 psig minimum to 12 psig maximum). REG2 2-13 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 102 Measures exhalation flow. Solenoid, autozero, expiratory +6 V, three-way solenoid valve. Energized (common to normally SOL2 pressure transducer closed) when transducer is autozeroed. De-energized (common to normally open) all other times. 2-14 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 103 Water trap Collects condensate as the compressed air cools in the heat exchanger. Automatically drains collected water to the housing base where it evaporates. 2-15 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 104: Inspiratory Module

    The inspiratory module, which is part of the BDU, is shown in Figure 2-13 and Figure 2-14. It includes the gas supply conditioning, flow control, and safety valve/inspiration monitoring subsystems. Figure 2-15 is a diagram of the inspiratory module gas flow. Figure 2-13. Inspiratory module 2-16 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 105: Gas Supply Conditioning Subsystem

    It does this by sending signals to the electronics when a gas source falls below a minimum pressure. Figure 2-17 illustrates the gas flow through gas supply conditioning subsytem. 2-17 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 106: Figure 2-16. Gas Supply Conditioning Subsystem

    Figure 2-16. Gas supply conditioning subsystem Figure 2-17. Gas supply conditioning subsystem gas flow diagram 2.3.1.1.1 Gas supply conditioning subsystem components The oxygen side of the gas supply conditioning subsystem includes the following components: 2-18 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 107 • The oxygen pressure valve (TP1) checks the oxygen regulator setting (REG1). • The oxygen pneumatic noise filter (F7), housed in the flow sensor manifold, conditions gas flow by eliminating swirling of gas induced by elbows and restrictions. 2-19 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 108 • The air pressure valve (TP2) checks the oxygen regulator setting (REG2). • The air pneumatic noise filter (F6), housed in the flow sensor manifold, conditions gas flow by eliminating swirling of gas induced by elbows and restrictions. 2-20 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 109: Figure 2-18. Gas Supply Conditioning Subsystem Components

    Theory of operation Oxygen inlet filter (F3) Oxygen impact filter (F1) Oxygen Filters (F1 and F3) Air inlet filter (F2) Oxygen/air pressure switches (PS1/PS2) Figure 2-18. Gas supply conditioning subsystem components 2-21 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 110 (F1), traps particles down to 65 m (microns) and the subsequent filter, the oxygen inlet filter (F3), traps particles down to 0.3 m (micron). The oxygen transfer tube directs oxygen to the check valve manifold. There is no check valve for oxygen. 2-22 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 111: Flow Control Subsystem

    The flow control subsystem, shown in Figure 2-20 and Figure 2-19, controls the mixture and flow of oxygen and air to the patient. It is mounted on the gas supply conditioning subsystem. PSOL1 PSOL2 Figure 2-19. Flow control subsystem 2-23 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 112: Figure 2-20. Flow Control Subsystem Gas Flow Diagram

    • Q1 and Q2 use a hot film (Figure 2-21) within each sensor to measure gas velocity (flow). The hot film constitutes one leg of a bridge circuit. The bridge is supplied by a constant 2-24 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 113: Figure 2-22. Flow Control Subsystem Components

    Oxygen and air flow sensors (Q1 and Q2) with ferrites installed Oxygen and air proportional solenoid valves (PSOL1 and PSOL2) Figure 2-22. Flow control subsystem components 2-25 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 114: Safety Valve And Inspiration Monitoring Subsystem

    • A safety valve section that, under certain circumstances, vents excessive ventilator pressure and lets the patient breathe room air. • An inspiration monitoring section that monitors the pressure and oxygen concentration of the inspiratory gas. 2-26 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 115: Figure 2-23. Safety Valve And Inspiration Monitoring Subsystem

    The safety valve section includes the following components, shown in Figure 2-24: • The safety valve (SV) consists of a +12 V solenoid actuator that is normally energized (closed) while the ventilator operates. The closed valve prevents gas from escaping to the 2-27 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 116 • The absolute pressure transducer (PA), located on the inspiratory electronics PCB, measures atmospheric pressure (psia). One port is opened to the atmosphere and one port is connected to a sealed vacuum chamber. 2-28 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 117: Figure 2-25. Safety Valve And Inspiration Monitoring Subsystem Components

    Theory of operation Safety valve (SV) Seat Leaf Check valve housing Oxygen sensor Oxygen sensor port 8-0112 Inspiratory outlet manifold Figure 2-25. Safety valve and inspiration monitoring subsystem components Oxygen sensor (OS) 2-29 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 118 O (measured at PE), components in the safety valve subsystem work together to relieve excess pressure. If pressure at the safety valve reaches nominal 115 cmH O, the valve is forced open. 2-30 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 119: Figure 2-26. Safety Valve Open Gas Flow Diagram

    PEEP plus the effective working pressure for spontaneous, pressure-supported breaths. Ongoing diagnostics monitor ventilator pressures and check for severe occlusions and circuit disconnects by comparing pressure at PE and PI. 2-31 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 120: Inspiratory Module Operation

    Check valve manifold Air check valve (CV2) Compressor check valve (CV4) Air impact screen filter (F4) Air regulator (REG2) Flow sensor manifold Air pressure valve (TP2) Air pneumatic noise filter (F6) 2-32 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 121 Inspiration monitoring. The oxygen concentration of the delivered gas is measured by the oxygen sensor (OS). The ventilator alarms if the monitored oxygen concentration is not within ±7% of the oxygen percentage setting. Pressure measurements are made by two differential pressure transducers. 2-33 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 122: Figure 2-28. Oxygen Flow Diagram

    Check valve manifold Oxygen impact screen filter (F5) Oxygen regulator (REG1) Regulator vent Flow sensor manifold Oxygen pressure valve (TP1) Oxygen pneumatic noise filter (F7) Oxygen flow sensor (Q1) PSOL/SV manifold 2-34 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 123 REG1 reduces oxygen pressure from 35 – 100 psig to 9 – 12 psig (10.5 psig nominal). It vents at a maximum of 2 L/min (outside the BDU to prevent oxygen buildup). 2-35 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 124: Patient System

    A variety of optional accessories can be used in the patient circuit. Figure 2-29 shows a typical patient system configuration. Inspiratory filter (F8) Humidification device Water trap (2 PL) Figure 2-29. Patient system (minus exhalation collector vial and expiratory filter) 2-36 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 125: Patient System Components

    30 L/min with 99.99% efficiency. It has 22-mm ISO conical connections. The NeoMode DAR (PN 351919005) disposable expiratory filter and the Puritan Bennett Neo Re/x800 reusable expiratory filter are designed specifically for use with the NeoMode option.
  • Page 126: Patient System Operation

    • The mounting plate, used only with neonatal patient systems, allows the ventilator to accommodate the neonatal expiratory filter. Caution To protect the exhalation assembly, always use a Puritan Bennett expiratory filter or a recommended DAR expiratory filter. 2.3.2.2 Patient system operation A predetermined oxygen/air gas mixture flows from the ventilator, through F8, and the optional humidification device.
  • Page 127: Exhalation Module Components

    • The exhalation flow sensor (Q3) provides flow information on exhaled gas. Flow sensor measurements are used to determine net gas flow to the patient and spirometry. 2-39 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 128: Exhalation Module Operation

    The expiratory pressure transducer (PE), on the exhalation transducer PCB, and the exhalation flow sensor (Q3) provide readings used in breath delivery calculations. 2-40 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 129: Figure 2-33. Exhalation Module Components

    Theory of operation Exhalation flow sensor (Q3) Figure 2-33. Exhalation module components NOTE: Exhalation module parts are shown here for reference only. For assembly illustration, refer to Figure 9-18 on page 9-53. 2-41 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 130 Theory of operation Expiratory pressure transducer autozero solenoid (SOL2) Exhalation check valve (CV5) Exhalation heater (EXH HTR) Exhalation module components (continued) 2-42 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 131: Compressor Unit

    The optional, cart-mounted 806 compressor unit, shown in Figure 2-35, is the latest style compressor system available for use with the Puritan Bennett 840 Ventilator System. When AC power is present, the compressor system supplies air to the ventilator in the event that a hospital air source is unavailable or the source pressure drops below 26 psig.
  • Page 132: Figure 2-36. 806 Components

    (F11) to prevent contaminants from entering the transducer. Pressure relief valve To heat exchanger Motor/compressor Silencer Cooling fans 0.3 micron filter Silencer Intake filter Filter foam Figure 2-36. 806 components 2-44 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 133: Figure 2-37. Heat Exchanger

    Theory of operation Figure 2-37. Heat exchanger Water trap Filter element Figure 2-38. 806 water trap assembly 2-45 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 134: Compressor Unit Operation

    Figure 2-40. 806 back panel 2.3.4.2 806 Compressor unit operation The 806 compressor is powered by and communicates with the Puritan Bennett 840 BDU. The BDU sends a signal to the compressor after completing POST, and the compressor enters either stand-by mode or run mode depending upon the state of the air-side pressure switch, PS2.
  • Page 135: Figure 2-41. 806 Cooling Fans

    The air dryer also lowers the dew point to a temperature below ambient, preventing condensation from entering the ventilator. Figure 2-41. 806 cooling fans 2-47 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 136: Summary Of Electrical Components

    Figure 2-84. Electrical system 2.4.1 Summary of electrical components The Puritan Bennett 840 ventilator electrical system, shown in Figure 2-42, includes the following: • AC distribution components, including power cord, AC panel (circuit breakers, AC filter PCB, power relay, and other components), and power switch •...
  • Page 137: Table 2-2. Electronic Component Descriptions

    Output signals from the BPS PCB include charging*, charged*, BPS model, IBATT, VBATT, and E-BP. NOTE: When the ventilator is running on battery power, the compressor and humidifier are non-operational. * indicates signal is active low 2-49 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 138 Breath delivery (BD) CPU PCB Contains the electronics and software that control all breath delivery functions in the Puritan Bennett 840 Ventilator System. Communicates with the GUI CPU to respond to operator inputs and display ventilation parameters. Analog interface (AI) PCB Provides the interface for all analog signals in the ventilator.
  • Page 139 Delivers a voltage proportional to the oxygen concentration. Ventilator alarms if the monitored oxygen   concentration is not within 7% of the oxygen percentage setting. 2-51 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 140 Motor compressor AC powered and logic-controlled, supplies compressed air to the ventilator when adequate wall air pressure is unavailable. Contains internal over-temperature protection device. 2-52 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 141: Overview Of Electrical System Operation

    PCB. 2.4.2 Overview of electrical system operation The Puritan Bennett 840 Ventilator System uses a dual-microprocessor architecture with an Ethernet network as the primary channel of communication between the two 68040 microprocessors. The BDU microprocessor handles the breath delivery control and monitoring functions;...
  • Page 142 Theory of operation This page intentionally blank. 2-54 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 143: Figure 2-42. Electrical System Block Diagram (10.4-Inch Gui-Trending-Enabled Cpu Pcb Shown)

    Figure 2-42. Electrical system block diagram (10.4-inch GUI—Trending-enabled CPU PCB shown) Remove pages 2-55 and 2-56. (Page 2-56 is also printed on the reverse side of the foldout.) Figure 2-42. Electrical system block diagram (10.4-inch GUI—Trending-enabled CPU PCB shown) 2-55 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 144 Theory of operation This page intentionally blank. 2-56 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 145 CPU PCB shown) Remove pages 2-57 and 2-58. (Page 2-58 is also printed on the reverse side of the foldout.) Figure 2-43. Electrical system block diagram (10.4-inch GUI—Non-Trending-enabled CPU PCB shown) 2-57 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 146 Theory of operation This page intentionally blank. 2-58 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 147 Electrical system block diagram (9.4-inch GUI—Trending-enabled CPU PCB shown) Remove pages 2-59 and 2-60. (Page 2-60 is also printed on the reverse side of the foldout.) Figure 2-44. Electrical system block diagram (9.4-inch GUI—Trending-enabled CPU PCB shown) 2-59 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 148 Theory of operation This page intentionally blank. 2-60 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 149 Electrical system block diagram (older 9.4-inch GUI shown) Remove pages 2-61 and 2-62. (Page 2-62 is also printed on the reverse side of the foldout.) Figure 2-45. Electrical system block diagram (older 9.4-inch GUI shown) 2-61 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 150: Ac Distribution Components

    2.4.3.1 Power cord The Puritan Bennett 840 Ventilator System includes a detachable power cord. The power cord has an IEC-standard, three-prong connector. The plug end varies, corresponding to different country requirements.
  • Page 151: Figure 2-46. Ac Panel

    Potential equalization (compressor/ connector (J2) humidifier) circuit breaker Compressor (CB2) receptacle (J4) AC filter Main circuit breaker (CB1) Power relay (K1) Fuse Auxiliary (compressor/ humidifier circuit breaker (CB2) Figure 2-46. AC panel 2-63 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 152: Figure 2-47. Puritan Bennett 840 Ventilator System Interconnect Diagram - Ac Panel

    Auxiliary mains circuit breaker Mains inlet To power supply Potential equalization point compartment Compressor outlet Power relay AC PANEL MODULE Figure 2-47. Puritan Bennett 840 Ventilator System interconnect diagram – AC panel 2-64 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 153: Power Switch

    A cover protects the power switch and prevents it from accidentally being turned off. An LED indicator beside the power switch lights to indicate that power is available to the ventilator. Power switch Indicator (S1) Figure 2-48. Power switch (S1) and indicator 2-65 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 154: Power Supply

    The power supply has no test points, and cannot be field-adjusted. Power supply voltages are displayed during the analog data display test in EST. Figure 2-49. Power supply assembly 2-66 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 155: Figure 2-50. Puritan Bennett 840 Ventilator System Interconnect Diagram - Power Distribution

    To AC inspiratory panel module POWER SUPPLY To backup power supply (BPS) (This side is located adjacent to inspiratory module) Figure 2-50. Puritan Bennett 840 Ventilator System interconnect diagram – Power distribution 2-67 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 156: Bps

    20 hours to recharge. The BPS contains a battery pack (Figure 2-53) (which includes two +12 V, lead-acid batteries and a 15 A fuse) and a PCB (Figure 2-54). Figure 2-51. 802 BPS 2-68 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 157: Figure 2-52. 803 Bps

    Theory of operation Figure 2-52. 803 BPS Fuse Figure 2-53. 802 BPS battery pack 2-69 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 158: Figure 2-54. Bps Pcb

    Theory of operation Figure 2-54. BPS PCB 2-70 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 159: Card Cage

    The card cage, shown in Figure 2-55 and Figure 2-56, includes the motherboard PCB, the BD CPU PCB, and the AI PCB. BDU CPU PCB AI PCB Motherboard PCB Figure 2-55. Card cage with all PCBs installed 2-71 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 160: Figure 2-56. Puritan Bennett 840 Ventilator System Interconnect Diagram - Card Cage

    UNIT (BDU) TEST switch (SW2) indicator GUI-to-BDU cable assy AI PCB PTS 2000 connection 9-pin BDU/compressor Data DC cable Figure 2-56. Puritan Bennett 840 Ventilator System interconnect diagram – Card cage 2-72 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 161: Motherboard Pcb

    The motherboard PCB consists of a multilayer PCB, connectors for plug-in boards and I/O interface, and related electrical filters and protective devices. AI PCB connectors BD CPU PCB connectors Figure 2-57. Motherboard PCB 2-73 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 162: Figure 2-58. Motherboard Pcb In Place

    Theory of operation Motherboard PCB Figure 2-58. Motherboard PCB in place 2-74 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 163: Figure 2-59. Motherboard Pcb Block Diagram

    EXH MOD PWR & SENS TO/FROM EXHALATION MODULE BD ALARM +12 V EXH AZ EXH MODULE CTRL INSP MOD PWR & SENS TO/FROM INSPIRATORY INSP MODULE CTRL MODULE Figure 2-59. Motherboard PCB block diagram 2-75 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 164: Bdu Cpu Pcb

    The BDU CPU PCB (Figure 2-60), in conjunction with the AI PCB, provides microprocessor control of all breath delivery functions for the Puritan Bennett 840 Ventilator System. It also communicates with the GUI CPU PCB for display and control information from the operator.
  • Page 165: Figure 2-60. Bd Cpu Pcb

    • A diagnostic LED array (with supporting circuitry) indicates the status of the BD CPU PCB. During POST, they indicate the current test step. A ninth LED displays the supervisory mode status of the CPU. 2-77 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 166 • The service mode switch is a push-button on the board edge used to activate this mode. The BD CPU PCB generates an NMI under any of these conditions: • Ethernet parity error detected • Power fail signal • A/D converter system error 2-78 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 167: Table 2-3. Novram Contents

    Calculated during SST BD NOVRAM expiratory resistance Oxygen and air flow Calculated during EST BD NOVRAM sensor (Q1 and Q2) offsets Absolute pressure Calculated during PA calibration BD NOVRAM transducer (PA) offset 2-79 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 168: Analog Interface (Ai) Pcb

    • The valve control and drive circuit provides drive signals for the PSOLs, EV, and the EV stabilizer device. PTS 2000 connector Data key connector Compressor connector AI LEDs Figure 2-61. AI PCB 2-80 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 169 The software reads this signal to determine whether the BPS is currently connected to the ventilator. • Other buffers provide conditioning for various ventilator signals. • The compressor interface circuit, in conjunction with the compressor PCB, controls and monitors the compressor operation. 2-81 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 170: Data Key Subsystem

    Call your Puritan Bennett representative if the data key requires replacement due to loss or failure. The Puritan Bennett 840 Ventilator System uses a data key (Figure 2-62) to record data specific to a particular ventilator unit. The data key provides a way to retain data when PCBs or the battery are removed from the ventilator.
  • Page 171: Figure 2-63. 10.4" Non-Te Gui Cpu Pcb

    Theory of operation Figure 2-63. 10.4” Non-TE GUI CPU PCB compact flash component Figure 2-64. 10.4” TE GUI CPU PCB 2-83 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 172: Figure 2-65. 9.4" Non-Te Gui Cpu Pcb And Backlight Inverter Pcb In Place

    Theory of operation GUI CPU Backlight inverter PCB Figure 2-65. 9.4” Non-TE GUI CPU PCB and backlight inverter PCB in place compact flash component Figure 2-66. 9.4” GUI TE GUI CPU PCB 2-84 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 173 RAM contents. • The compact flash component, available only on the Trending-enabled GUI CPU PCB, provides additional memory required for use with the Puritan Bennett 840 Ventilator System Trending software option. • The real-time clock tracks the ventilator’s operational time and short-duration power losses (up to 60 minutes).
  • Page 174: Figure 2-67. Puritan Bennett 840 Ventilator System Interconnect Diagram - Gui 10.4-Inch

    To BDU RS-232 Remote alarm Keyboard cable assy Rear of keyboard Touch frame cable assy GUI MODULE Figure 2-67. Puritan Bennett 840 Ventilator System interconnect diagram – GUI 10.4-inch LCD panels 2-86 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 175 The GUI CPU PCB generates an NMI under any of these conditions: • GUI +5 V is high • GUI +12 V is out of range • Ethernet parity error detected • SAAS microcontroller failure 2-87 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 176: Figure 2-68. Puritan Bennett 840 Ventilator System Interconnect Diagram - Gui 9.4-Inch Lcd Panels (Non-Te Gui Cpu Pcb P/N 4-073200-Sp Shown)

    Touch Frame Rear of keyboard cable assy GUI MODULE Figure 2-68. Puritan Bennett 840 Ventilator System interconnect diagram – GUI 9.4-inch LCD panels (Non-TE GUI CPU PCB P/N 4-073200-SP shown) 2-88 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 177: Lcd Panels (Te Gui Cpu Pcb Shown)

    Remote alarm Keyboard Touch Frame Rear of keyboard cable assy GUI MODULE Figure 2-69. Puritan Bennett 840 Ventilator System interconnect diagram – GUI 9.4-inch LCD panels (TE GUI CPU PCB shown) 2-89 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 178: Touch Frame Pcb

    During normal conditions, when all LEDs and IR detectors are functioning properly, the redundant detectors are used to provide a more accurate output by averaging the locations of the blocked beams. 2-90 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 179: Keyboard Assembly With Knob

    The multiposition knob assembly, which is part of the keyboard assembly, permits ventilator setting selections or changes. Knob encoder/decoder circuitry on the GUI CPU PCB determines the direction and position of the shaft based on encoder outputs. 2-91 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 180: Figure 2-71. Keyboard Assembly

    Theory of operation Figure 2-71. Keyboard assembly 2-92 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 181: Gui Led Pcb

    The indicators are color-coded red, yellow, or green to show status. Each indicator (except "compressor operating" and "on BPS power") includes redundant LED strings so the indicators will stay lit if an LED bar burns out. Figure 2-72. GUI LED PCB 2-93 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 182: Backlight Inverter Pcb And Lcd Lamps

    LCD panels simultaneously. The LCD lamps are part of the 10,000-hour preventive maintenance kit on the 9.4-inch GUI. Figure 2-73. 10.4” GUI LCD panels 2-94 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 183: Gui Alarm Assembly

    Figure 2-74. 9.4-inch LCD panels and backlight tubes 2.4.8.6 GUI alarm assembly The GUI alarm assembly (Figure 2-75), the ventilator’s primary alarm, emits alarm sounds under control of the GUI CPU PCB. 2-95 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 184: Figure 2-75. Gui Alarm Assembly

    Theory of operation Figure 2-75. GUI alarm assembly 2-96 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 185: Bdu Led Pcb

    • The absolute pressure transducer (PA) senses absolute pressure in the inspiratory module. It is used to determine atmospheric pressure for volume delivery. • The oxygen sensor (OS) amplifier provides an interface for the oxygen concentration sensor mounted on the PCB. 2-97 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 186: Figure 2-78. Puritan Bennett 840 Ventilator System Interconnect Diagram - Inspiratory Module

    4-071321-00 Press switch harness assy 4-071326-00 Air pressure Oxygen pressure switch (PS1) switch (PS2) 4-072206-00 4-072206-00 INSPIRATORY MODULE Figure 2-78. Puritan Bennett 840 Ventilator System interconnect diagram – Inspiratory module 2-98 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 187: Exhalation Transducer Pcb

    (PE), which is on this PCB, senses the pressure difference in the exhalation circuit relative to ambient air pressure. The PCB uses a +10 V reference. Figure 2-79. Exhalation transducer PCB 2-99 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 188: Figure 2-80. Puritan Bennett 840 Ventilator System Interconnect Diagram - Exhalation Module

    Exhalation Exhalation I/O module cable assy cable assy Expiratory Exhalation pressure transducer valve (PE) Exhalation transducer PCB EXHALATION MODULE Figure 2-80. Puritan Bennett 840 Ventilator System interconnect diagram – Exhalation module 2-100 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 189: Bd (Continuous-Tone) Alarm Assembly

    Figure 2-81. BD alarm assembly 2.4.12 806 Compressor unit The 806 compressor unit’s electrical components include: AC power distribution components, a compressor motor, and a compressor PCB. The compressor unit is shown in Figure 2-82. 2-101 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 190 Theory of operation Figure 2-82. 806 compressor 2-102 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 191: Compressor Unit Ac Power Distribution Components And Motor

    PCB’s primary function is to control the pressure of the air delivered to the ventilator when wall air is not present or is insufficient. Data cable Pressure transducer Power hose cord Solenoid cable Compressor cable cables Ground Figure 2-84. 806 compressor PCBA installed 2-103 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 192 Section 2.4.12.3 describes how this logic controls compressor component operation. This logic shuts off the compressor when AC is inadequate or the thermostat on the PCB detects overtemperature. 2-104 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 193: Figure 2-85. 806 Compressor Pcb Block Diagram

    (PC) CKTS CAPACITOR MOTOR COMP COMPRESSOR CONTROL HOUSING CKTS TO/FROM AI PANEL EEPROM CKTS COMP OVER TEMP +5 V COMP THERMOSTAT +12 V COMP Figure 2-85. 806 compressor PCB block diagram 2-105 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 194: Figure 2-86. Puritan Bennett 840 Ventilator System Interconnect Diagram - Compressor Unit

    Compressor unit Remove pages 2-106 and 2-107. (Page 2-107 is also printed on the reverse side of the foldout.) Figure 2-86. Puritan Bennett 840 Ventilator System interconnect diagram – Compressor unit 2-106 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 195: Compressor Unit Operation

    If pressure in the accumulator drops below 22.5 psig (due to small leaks in the system or cooling of compressed air), the compressor starts and recharges the accumulator to 27 psig (see Figure 2-87). 2-107 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 196: Figure 2-87. Compressor Operational Sequence

    After one second, a solid-state relay on the PCB is energized. After another 0.5 seconds, SOL3 is de-energized. 2-108 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 197: Breath Delivery

    The ventilator does not autotrigger when pressure sensitivity is greater than 2 cmH O or when flow sensitivity is greater than 1 L/min for pediatric patients or 1.5 L/min for adult patients. 2-109 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 198: Figure 2-89. Inspiration Gas Flow Diagram

    Theory of operation Figure 2-89. Inspiration gas flow diagram 2-110 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 199: Pressure Triggering

    Backup limits (time, circuit pressure, and ventilator pressure) prevent inspirations of excessive duration or pressure. If a particular breath is subject to more than one backup limit, exhalation is triggered by whichever method goes into effect first. 2-111 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 200: Time-Cycling Method

    2.5.2.6 High ventilator pressure limit The high ventilator pressure limit applies to volume-based mandatory breaths only. If the inspiratory pressure (measured at PI) equals or exceeds 100 cmH O, the ventilator transitions to exhalation. 2-112 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 201: Figure 2-90. Exhalation Gas Flow Diagram

    Theory of operation Figure 2-90. Exhalation gas flow diagram 2-113 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 202: Other Hardware Operations

    (exhaled tidal volume) to be transiently displayed as lower or higher than the actual exhaled volume. This is a result of initial spirometry calculations and does not reflect actual volume exhaled by the patient. 2-114 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 203: Pressure Transducer Autozero

    Autozeroing is performed every minute for 20 minutes; every 2 minutes after 20 minutes, up to one hour; and every 5 minutes after 1 hour. 2-115 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 204: Figure 2-91. Pressure Transducer Autozero Mode Gas Flow Diagram

    Theory of operation Figure 2-91. Pressure transducer autozero mode gas flow diagram 2-116 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 205: Power Monitoring And Power Fail Handling

    802 BPS: 60 minutes minimum (30 minutes on ventilators built prior to July 2007) 803 BPS (Four-hour battery life with new fully charged battery) Figure 2-92. Power loss sequence 2-117 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 206: Supply Voltage Monitoring

    • Displays the elapsed time since the loss of ventilatory support on the GUI • Does not display patient data, including waveforms, on the GUI • Does not detect patient circuit occlusion or disconnect conditions 2-118 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 207: Figure 2-93. Safety Valve Open Diagram

    Theory of operation Figure 2-93. Safety valve open diagram 2-119 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 208: Occlusion Handling

    During a severe occlusion, apnea detection, expiratory pause, manual inspirations, and maneuvers are suspended, and the P (high airway pressure) alarm limit is disabled. MEAN The GUI does allow you to change ventilator settings. 2-120 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 209: Figure 2-94. Pressure Release, Patient Circuit Occluded Diagram

    Theory of operation Figure 2-94. Pressure release, patient circuit occluded diagram 2-121 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 210 Theory of operation This page intentionally blank. 2-122 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 211: Self Tests

    TEST button while entering service mode. 3.3 Self tests and background checks The Puritan Bennett 840 Ventilator System has self test capabilities that include POST (power- on self test), SST (short self test), EST (extended self test), and background checks. These self test operations are described in Table 3-1, and Table 3-2 details the components tested by various self tests.
  • Page 212: Est (Extended Self Test)

    EST requires a “gold standard” test circuit, available from Puritan Bennett. All required software support to perform an EST is resident on the ventilator. EST testing, excluding tests of optional equipment, such as the compressor, takes about 15 minutes to complete.
  • Page 213: Power On Self Test (Post)

    3.4 Power on self test (POST) Table 3-2: Components tested by self tests Background Service mode Component tested POST checks calibrations Power supply/voltage checks AC monitor voltage Supply voltages (including BDU and GUI voltages) BPS supply voltage/current Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 214: Table 3-1. Self Tests

    Table 3-2: Components tested by self tests (continued) Background Service mode Component tested POST checks calibrations BPS charging and discharging BPS model Patient system Patient circuit leak Patient circuit occlusions/resistance Patient circuit compliance Expiratory filter occlusion/resistance Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 215 Oxygen and air PSOL forward leak Oxygen and air PSOLs stuck open or stuck in other position Inspiration and exhalation pressure transducer autozero solenoids Inspiration pressure transducer Exhalation pressure transducer Inspiration and exhalation pressure transducer cross-check Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 216 Audio alarm and user interface sound- producing subsystem Alarm cable (BDU) Nurse’s call relay +10 V reference (used by ADCs and DACs) ADC/DAC functionality Power fail capacitor Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 217 POST routines are ordered so that each routine requires successively more operational hardware than the last. This sequence allows POST to systematically exclude electronic components as causes of system malfunctions. Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 218: Safety

    POST’s memory test preserves all data necessary to determine ventilator settings and initializes the remaining memory to a predefined state. • Any other processors in the system initiates its own POST and reports the test results to the host processor. Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 219: Post Following Power Interruptions

    POST. NOTE: Puritan Bennett recommends that a BPS is always installed on the ventilator. If there is a loss of AC power with a charged BPS installed, the ventilator will switch to the DC source and will not run POST upon restoration of AC power.
  • Page 220: Structure Of Post

    Table 3-3 compares the BDU and GUI POST sequences, indicates the tests performed and shows which indicators are active during each phase. 3-10 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 221: Table 3-3. Post Structure

    NOTE: Prior to the • BDU continuous tone alarm turned beginning of Phase 1, on and off twice all of the LEDs on the Trending-enabled GUI turn on for 200 to 350 ms. 3-11 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 222: Table 3-4. Post Outcomes

    • POST continues to end. • Minor fault detected. • Check System Diagnostic Log for any associated error codes. • Ventilation proceeds. • Rerun POST/EST. • A DEVICE ALERT is annunciated. 3-12 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 223: Sst (Short Self Test)

    10 minutes in service mode before running SST, to ensure accurate testing. • Puritan Bennett recommends that you run SST every 15 days, between patients, after a major service or repair (refer to Table 5-2 on page 5-6), and when you change the patient circuit.
  • Page 224: Hardware Requirements

    1. If necessary, the compressor can be used as the only gas source. However, only the gas pathway, flow sensor, and PSOL of the single gas (air or O ) will be verified during SST. 3-14 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 225: Running Sst

    SST can affect the accuracy of compliance calculation and the delivered and measured exhaled tidal volumes. You must rerun SST after changing the patient circuit type, installing a new patient circuit, or after adding or changing accessories. 3-15 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 226 To begin normal ventilation (if SST has not detected an ALERT or FAILURE), touch EXIT SST, then press ACCEPT. The ventilator reruns POST, then displays the Ventilator Startup screen. 3-16 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 227: Table 3-6. Sst Tests

    Overriding an ALERT could cause inaccurate patient pressure estimation. FAILURE if test detects exhalation compartment occlusion, expiratory filter occlusion or damage, or you did not follow prompts to detach and reattach tubing correctly. 3-17 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 228: Table 3-7. Sst Individual Test Results

    Exit SST in order to service ventilator or review error codes by condition is declared. pressing EXIT SST. OVERRIDDEN An ALERT status was overridden, Touch EXIT SST. and ventilation is authorized. 3-18 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 229: Est (Extended Self Test)

    Table 3-10 lists the tests that comprise EST. For more details about these tests and associated diagnostic codes, see Chapter 6. For a theory of operation of EST, consult the Puritan Bennett 800 Series Ventilator System Operator’s and Technical Reference Manual.
  • Page 230: Hardware Requirements

    The gold standard circuit connects to an ADULT expiratory filter installed at the From patient port. The other end of the gold standard circuit connects directly to the To patient port: Do not install an inspiratory filter. 3-20 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 231: Figure 3-2. Est Setup

    You can touch EXIT EST during EST to halt testing. You can touch EXIT EST again to resume testing, or press ACCEPT to restart the ventilator (if EST has not detected an ALERT or FAILURE). 3-21 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 232 When all of the tests in EST are complete, the Extended Self Test screen displays all individual test results and EST outcome. Refer to Tables Table 3-10, Table 3-11, and Table 3-12 for additional information. 3-22 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 233: Figure 3-3. Est Screens During Testing

    Self tests 08:28 18 Jul 2003 Figure 3-3. EST screens during testing 3-23 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 234 Self tests NOTE: The screens given above are intended as examples only. Your unit may vary slightly in appearance. 3-24 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 235: Table 3-10. Est Tests

    Verifies that GUI LEDs are functioning by prompting the Observe GUI indicators and press operator to acknowledge that the LEDs are turned on. ACCEPT (if on) or CLEAR (if not on). 3-25 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 236 • If wall air is connected, disconnect wall air as directed. compressor pressure switch, and compressor operating states (run, standby, and disabled). • If AC is not connected, connect AC power as directed. 3-26 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 237: Table 3-11. Est Individual Test Results

    Exit EST in order to service ventilator or review error codes by touching EXIT EST. In Single Test EST, touching “NEXT” allows you to choose another test to run 3-27 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 238: How To Run Single Test Est

    Enter service mode and initiate single Test EST as follows: Turn on power to ventilator. (If power is already on, turn it off, then back on.) 3-28 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 239 If a test fails or completes with an ALERT outcome, REPEAT to retry the test, touch “NEXT” to choose another test, or exit Single Test EST by touching the EXIT button. 3-29 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 240 Self tests This page intentionally blank. 3-30 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 241: Service Mode

    H A PT ER Service mode The Puritan Bennett™ 840 Ventilator System operates in two modes: Patient Ventilation and Service Mode. The Service Mode is intended for use by a trained service technician to aid in ventilator testing and troubleshooting and to perform system calibrations.
  • Page 242 Service mode Figure 4-1. Service Mode screens Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 243: Service Mode Functions

    DATE/TIME Adjusts current date and time and allows the user to select a date format. EXIT Exits Service Mode. Other Screens Performs miscellaneous service, test, and calibration functions. Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 244: Figure 4-2. Service Mode Functions

    Transducer Test Calibration (10.4 GUI only) Compact Flash Test* External Nominal Pressure Control Line Voltage Unit Baud Rate *Only available if Trending option is installed. LOWER SCREEN Figure 4-2. Service Mode functions Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 245: Sst Result: Displaying Sst Results

    System Diagnostic Information and EST/SST Diagnostic logs can only be erased by using the Puritan Bennett 840 VTS Breath Labs Software. The alarm log is automatically erased each time a new patient is setup is selected and when EST is run.
  • Page 246: Figure 4-4. System Information Log

    Service mode Figure 4-4. System Information Log Figure 4-5. EST/SST Diagnostic Log Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 247: Alarm Log: Displaying The Alarm History

    The date formats shown in parentheses appear in the Trending graph and table screens if the Puritan Bennett 840 Trending software option is installed. If you make a change to the date format while viewing the SST Result Log or Test Summary, the time stamps do not update until the screen is closed and re-opened.
  • Page 248: Exit: Exiting Service Mode

    O or hPa). External Test Control Lets you test the ventilator using a computer, Breath Labs 840 VTS Software, and PTS 2000 tester, to run performance verification. Also used to download new application software to the Puritan Bennett 840 ventilator.
  • Page 249: Service Mode Setup

    Enter Service Mode and select the Other Screens button. Ensure air and oxygen are connected to the ventilator. Connect the gold standard circuit between the To Patient and From Patient ports. (Do not use a patient filter.) Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 250: Table 4-4. Exhalation Valve Calibration Errors

    Exhalation valve temperature out of 1. Verify that the ventilator has temperature out of range range. warmed up for at least 10 minutes in Service Mode. 2. Replace exhalation valve. 3. Replace AI PCB. 4-10 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 251 (5, 25, 50, 75, and 100 2. Verify adequate supply pressure. 3. Replace inspiratory pressure trans- ducer. 4. Replace expiratory pressure trans- ducer. 4-11 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 252: Vent Inop Test

    Service Mode will invoke the normal Service Mode. Troubleshoot as indicated in Table 4-6. You must repeat the Vent Inop Test and pass it with no failures before ventilation is allowed. 4-12 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 253: Flow Sensor Calibration

    4.2.10.5.1 Running Flow Sensor Calibration Enter Service Mode. Ensure both air and oxygen are connected to the ventilator. Connect the gold standard circuit between the To Patient and From Patient ports. 4-13 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 254: Table 4-7. Flow Sensor Calibration Errors

    4. Try installing a known good inspi- ratory module. If calibration passes, replace inspiratory elec- tronics PCB. If calibration fails, replace inspiratory module blind- mate cable. 5. Replace expiratory flow sensor (Q3). 6. Replace AI PCB. 4-14 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 255: Atmospheric Pressure Transducer Calibration

    Any such environmental changes will result in an incorrect adjustment being made to the ventilator. 4-15 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 256: Data Key Update

    On the 10.4-inch GUI only (with or without the Trending-enabled GUI CPU PCB), use the Serial Loopback Test to check the function of the serial ports on the back of the GUI. Required Equipment: Null modem cable (p/n 4-075361-00) or equivalent. 4-16 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 257: Compact Flash Test

    3. Replace GUI CPU PCB. 4.2.10.9 Compact flash test The compact flash test is only available if the Puritan Bennett 840 Ventilator System Trending software option and compact flash component on the TE GUI CPU PCB have been installed. This test checks the functionality of the compact flash component.
  • Page 258: Cal Info Duplication

    When troubleshooting with a GUI not originally matched to the BDU, do not press the ACCEPT ACCEPT key when prompted to match serial numbers. If you press , a software download will be required to restore the original serial numbers. 4-18 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 259: Table 4-12. Serial Number Setup Errors

    Serial number(s) doesn’t match the data key Press ACCEPT to confirm and proceed with Service Mode. NOTE: If an invalid data key is inserted, you can proceed with Service Mode but ventilation is not allowed. 4-19 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 260: Oxygen Sensor Calibration

    (DISABLED or ENABLED). 100% O /CAL 2 min INCREASE O 2 min During oxygen sensor calibration, the LED is turned OFF. Figure 4-6. 100% O /CAL In Progress indicator 4-20 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 261: Performance Verification

    Adapter, patient pressure (“T” connector) *4-011521-00 Cable, trigger *4-075360-00 Cable, null modem *4-075361-00 * These parts may be purchased individually or by ordering a Puritan Bennett 840 VTS Accessories Kit, P/N 4-076599-00. Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 262 Patient circuit, pediatric (reusable, without traps or heated wire) G-061223-00 Patient circuit, neonatal (disposable) Allegiance Healthcare PN 7441-452 * These parts may be purchased individually or by ordering a Puritan Bennett 840 VTS Accessories Kit, P/N 4-076599-00. Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 263 Tubing, silicone, 1/8 in. ID x 1/4 in. OD, 4.75 in. (12.1 cm) *4-008578-00 * These parts may be purchased individually or by ordering a Puritan Bennett 840 VTS Accessories Kit, P/N 4-076599-00. Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 264: Preliminary Ventilator Cleaning And Inspection

    Wye, patient circuit *4-000338-00 * These parts may be purchased individually or by ordering a Puritan Bennett 840 VTS Accessories Kit, P/N 4-076599-00. 5.3 Preliminary ventilator cleaning and inspection Clean and inspect the ventilator as follows: Warning To prevent disease transmission, use personal protective equipment when handling contaminated bacterial filters or other patient accessories.
  • Page 265: Preliminary Calibrations And Tests

    5.5 Preliminary calibrations and tests Before running the performance verification tests, perform the applicable calibrations and tests in Table 5-2. The numbers in the columns indicate the sequence in which the calibrations/tests must be performed. Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 266: Table 5-2. Pre-Performance Verification Testing And Calibration Requirements

    Unless the ventilator meets all of these requirements, PVT must be performed following the repair. 4. NeoMode only. 5. Perform Compact Flash Test only if Trending software option is installed or if compact flash component replaced. 6. Perform EST and electrical safety testing only. Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 267: Performance Verification Guidelines

    NOTE: To ensure that the ventilator performs within specifications, Puritan Bennett recommends that you use the Puritan Bennett PTS 2000 Performance Test System in conjunction with the BreathLab 840 Verification Test Software. If this equipment is not available, you can use an alternative test that checks some parameters but does not guarantee the integrity of the entire system.
  • Page 268: Electrical Safety Test

    Verify that ventilator power switch is off. Verify that ground resistance is < 0.1  If performing full performance verification testing, record this value for later entry when prompted by the Breathlab 840 VTS software program (Section 5.7.6). Turn on ventilator.
  • Page 269: Ground Isolation Check

    (Figure 5-1) and the ventilator chassis (at the power supply cooling fins) is > 975 PTS 2000 pin 3 Figure 5-1. Puritan Bennett 840 ventilator side view — PTS 2000 port Disconnect DMM test leads and resistor. 5.7.3 Extended self test (EST) NOTE: •...
  • Page 270: Regulator Setting Verification

    Connect a null modem cable between the two top serial ports. Follow the on-screen test prompts. If performing full performance verification testing, you will enter the result of the serial loopback test when prompted by the Breathlab 840 VTS software (Section 5.7.6).
  • Page 271: How To View The Test Results Log

    Using ventilator test program, press START TEST to begin test application. Follow prompts displayed on computer screen to complete performance verification. Access the Breathlab 840 VTS Software Help Screen for additional information. 5.7.6.1 How to view the test results log After you have finished the performance verification tests, you can view and print the test logs.
  • Page 272: Figure 2-2. Bdu

    Circuit Insert cork into high flow exhaust From Patient Port port during SST. After SST, attach lungs here. 2000 Figure 5-4. Puritan Bennett 840 Ventilator System PVT setup with PTS 2000 5-12 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 273: Figure 5-4. Puritan Bennett 840 Ventilator System Pvt Setup With Pts 2000

    RS-232 port input Serial port card COM 2, 3, 4 LAPTOP COMPUTER must be configured PTS 2000 Figure 5-5. Puritan Bennett 840 Ventilator System PVT setup with PTS 2000 and laptop computer 5-13 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 274: Manual Ventilator Check Using Equipment Other Than Pts 2000 Performance Test System

    • If you are using a pneumatic analyzer other than the PTS 2000 Performance Test System, you must also have an oxygen analyzer available. • The manual ventilator check (i.e., without PTS 2000 and Breathlab 840 VTS software) does not satisfy the manufacturer’s performance testing requirements. The manual ventilator check can be used as a general indicator of ventilator condition.
  • Page 275: Regulator Calibration Hose Repair Instructions

    Cut a 1/4” piece of silicone tubing (P/N 4-008577-00) and slide it over the new valve depressor (P/N 4-026790-00) and push the assembly into end of the brass fitting, ensuring that the seal is flush with the end of the brass fitting. 5-15 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 276: Figure 5-6. Regulator Calibration Hose Repair

    Performance verification Valve depressor 1/4” silicone tubing Brass regulator fitting Figure 5-6. Regulator calibration hose repair 5-16 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 277: Diagnostic Codes

    This chapter contains information to assist in the interpretation of diagnostics codes, indicated in the Puritan Bennett™ 840 Ventilator System’s error and status logs, as well as diagnostic codes reported by the diagnostic LED arrays on the BD CPU PCB and the GUI CPU PCB.
  • Page 278: Troubleshooting

    GUI (if possible) and to the LED array on the corresponding CPU PCB (BD or GUI). The safety valve and exhalation valve remain open to allow the patient to Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 279: Diagnostic Cpu Led Arrays

    LEDs present on the 10.4-inch GUI LED array. Ignore the top-most and the bottom- most LEDs and use the eight LEDs in between to determine the diagnostic code. Status LED diagnostic code LEDs (read from top to bottom) Figure 6-1. Location of BD LED array Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 280: Diagnostic Codes

    GUI problem that resulted in soft reset during operation GUI failure (during POST or background check) caused GUI or ventilator inoperative condition POST or background check alert or communication error in GUI Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 281: Organization Of Diagnostic Codes Table

    Other error identification information 6.7 Organization of diagnostic codes table Table 6-2 is a comprehensive summary of the diagnostic codes that the Puritan Bennett 840 software can report. The organization of the information in this table is as follows: • The first column in the table lists, in alphanumeric order, the codes.
  • Page 282 Failure to warm up the ventilator may result in false flow sensor or pressure transducer failures. When troubleshooting SST or EST, always use a known good patient circuit or gold standard circuit and filter(s). Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 283: Table 6-2. Puritan Bennett 840 Ventilator System Diagnostic Codes

    Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes Message Information / possible Code Test Response Corrective action displayed cause AE0110 Unable to EST Flow Alert Flow controller unable 1. Make sure oxygen supply is establish O flow...
  • Page 284 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause AE0306 Test circuit not EST Leak Test Alert Pressure not detected 1. Make sure test circuit is prop-...
  • Page 285 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause AE0611 GUI Screen Lock EST GUI Lamp Alert CLEAR key pressed to 1. Check/replace interconnect LED fails.
  • Page 286 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause AE1104 Insp check valve EST Safety Alert It took too little time to 1. Make sure test circuit is con-...
  • Page 287 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause AE1604 Run mode time Alert Compressor timer is not Listen for motor. If motor is on, Compressor running while replace compressor PCB.
  • Page 288 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause AE1701 Unable to Alert Compressor leak test 1. Troubleshoot accompanying perform Compressor cannot be run, because code AE1604 or AE1607.
  • Page 289 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause AE2103 Bad Backup EST Battery Test Alert While BPS was 1. Ensure the battery pack is fully...
  • Page 290 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause AS0011 Unable to SST flow sensor Alert 1. Flow controller 1. Make sure air supply is con-...
  • Page 291 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause AS0311 Low expiratory SST Expiratory Alert Pressure drop 1. Repeat test, following direc- filter P filter across expiratory filter tions more closely.
  • Page 292 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause AS0507 Compliance low Alert High compliance falls 1. Make sure correct patient tub- Compliance below 1.56 mL/cmH ing type was specified in SST.
  • Page 293 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause D18806 Assertion Background BDU reset 1. Check for wet expiratory or Checks (BD) inspiratory filter. 2. Check all cable connections: GUI Ethernet cable, internal cables.
  • Page 294 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause FE0003 Failed to reach EST Circuit Failure Unable to build 1. Make sure that test circuit is...
  • Page 295 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause FE0008 Failed to reach EST Circuit Failure Unable to build 1. Make sure air supply or com-...
  • Page 296 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause FE0103 Air flow sensor EST Flow Failure Inspiratory module air 1. Make sure air supply is con-...
  • Page 297 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause FE0107 Unable to EST Flow Failure Flow controller unable 1. Make sure air supply is con- establish air flow...
  • Page 298 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause FE0207 Air PSOL leak EST Gas Failure Air PSOL (PSOL2) 1. Check for leaks around the Q2...
  • Page 299 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause FE0215 Air zero flow EST Gas Failure During zero-flow check, 1. Make sure gas supplies are dis-...
  • Page 300 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause FE0402 Clear key fails. EST GUI Failure Wrong key pressed or 1. Repeat test. Keyboard Test key not pressed within 2.
  • Page 301 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause FE0901 Bad alarm cable EST BDU Audio Failure Alarm cable voltage is 1. Make sure BD alarm cable is Test out of range (<...
  • Page 302 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause FE1302 Exp valve temp EST Exp Valve Failure Exhalation valve 1. Ensure the unit has been Seal Test...
  • Page 303 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause FE1502 Bad exp heater EST Exp Heater Failure Exhalation heater 1. Replace exhalation heater. Test temperature did not 2.
  • Page 304 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause NOTE: For all "FS" codes (failures reported during SST), if the corrective action indicated is to run an EST to further diagnose the problem, perform the repair after the EST, verify the fix in EST, and then rerun SST in order to enter normal ventilator operation.
  • Page 305 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause FS0003 Air flow sensor SST Flow Failure Air flow sensor (Q2) 1. Correct the leak or occlusion in...
  • Page 306 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause FS0006 Unable to SST Flow Failure Flow controller unable 1. Make sure oxygen supply is establish O...
  • Page 307 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause FS0008 zero flow SST Flow Failure Inspiratory flow > 0.153 1. Verify no leaks at the Q1...
  • Page 308 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause FS0104 Cross-check SST circuit Failure Inspiratory/expiratory 1. Verify that the patient circuit failed pressure test pressure transducer...
  • Page 309 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause FS0201 Excessive leak SST Circuit leak Failure System pressure 1. Make sure patient circuit is dropped below failure connected and is not leaking.
  • Page 310 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause FS0405 Occluded SST Circuit Failure Expiratory limb Check for occluded patient tubing. exhalation limb Resistance pressure > 12.5 cmH (adult) or 7.5 cmH...
  • Page 311 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause HN0001 Non-maskable Other (BDU) BDU reset Replace BD CPU PCB. interrupt (NMI): Dynamic random access memory (DRAM) Parity...
  • Page 312 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause KB0003 Bad O PSOL Background Vent inop Oxygen PSOL (PSOL1) 1. If possible, run EST to deter-...
  • Page 313 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause KB0019 Bad BD -15 V Background Vent inop BDU -15 V supply out 1. Replace AI PCB.
  • Page 314 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause KB0030 Bad safety valve Background Vent inop Current to safety valve 1. Replace safety valve. current check (BDU) out of range 2.
  • Page 315 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause KB0065 SV current out of Background Vent inop Safety valve current out 1. Replace BD CPU PCB if accom-...
  • Page 316 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause KP0006 POST Phase 2 POST (BDU) Vent inop Replace BD CPU PCB. Initialization KP0007 Addressing POST (BDU) Vent inop Replace BD CPU PCB.
  • Page 317 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause KP0083 NMI Source POST (BDU) Vent inop Replace BD CPU PCB. Register Test KP0084 POST DRAM Test POST (BDU) Vent inop Replace BD CPU PCB.
  • Page 318 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause KP0116 Analog Devices POST (BDU) Vent inop Critical analog data out 1. Refer to Section 6.8 and...
  • Page 319 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause KP0131 Application Boot POST (BDU) Vent inop Replace BD CPU PCB. LB0006 Bad exhalation Background Alert Exhalation valve coil 1.
  • Page 320 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause LB0021 Bad GUI 5V Background Alert GUI +5 V supply out of 1. Check cabling from BD CPU...
  • Page 321 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause LB0056 Compressor Background Alert Compressor timer Replace compressor PCB. elapsed timer check (BDU) failure error LB0057 Compressor bad...
  • Page 322 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause LB0092 Battery Event Background Alert A status message 1. Connect the ventilator to a checks (BDU) regarding the state of known good AC power source.
  • Page 323 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause LB0092 6. Check/replace the battery pack fuse. (cont.) 7. Check that the battery voltage reads 26 V DC minimum. If not, replace the battery pack.
  • Page 324 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause LC0055 06 Socket error Communi- Alert The BD CPU requested If no other code is generated in...
  • Page 325 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause LC0077 Network msg Communi- Alert Network If no other code is generated in received timeout cations error...
  • Page 326 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause LP0115 DAC Test POST (BD) Alert Replace AI PCB. LP0116 Analog Devices POST (BDU) Alert Noncritical analog data 1.
  • Page 327 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause U05013 Assertion Background GUI reset Can occur if there is an 1. Check all cable connections. check (GUI) intermittent connection 2.
  • Page 328 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause UT0003 Address Error Background GUI reset Replace GUI CPU PCB if failure check (GUI) recurs. UT0004 Illegal...
  • Page 329 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause XB0068 Monitor apnea Background Vent inop Call Technical Support or your int fails checks (GUI) Puritan Bennett representative.
  • Page 330 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause XP0008 Kernel NOVRAM POST (GUI) GUI inop Replace GUI CPU PCB. Test XP0010 Interrupt POST (GUI) GUI inop Replace GUI CPU PCB.
  • Page 331 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause XP0090 DRAM Parity POST (GUI) GUI inop Replace GUI CPU PCB. Circuit Test XP0097 SAAS Self-Test POST (GUI) GUI inop Replace GUI CPU PCB.
  • Page 332 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause ZB0054 GUI EEPROM Background Alert This code may be 1. Ensure the GUI is installed onto checksum error...
  • Page 333 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause ZB0093 Compact FLASH Background Status Status message 1. Perform Service Mode com- error checks (GUI) indicates a compact pact flash test to reset device.
  • Page 334 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause ZC2003 DCI unknown Communica- Alert 1. Check host software for com- error Port 1 tions error patibility with ventilator.
  • Page 335 Diagnostic codes Table 6-2: Puritan Bennett 840 Ventilator System diagnostic codes (continued) Message Information / possible Code Test Response Corrective action displayed cause ZP0008 Kernel NOVRAM POST (GUI) Alert Noncritical NOVRAM Call Technical Support or your Test data failure Puritan Bennett representative.
  • Page 336: System Diagnostic Log And Bdu Post Analog Devices Test

    Q (exh) Temperature Q3 (exh) Flow Sensor SUBMUX 0: DACWRAP AI PCB Safety Valve Switched Side Safety Valve (SV) Assembly; check Inspiratory Module connections 12 V GUI Sentry GUI CPU PCB 6-60 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 337 BPS Battery Current BPS PCB AC Line Voltage Power Supply or AI PCB Exhalation Motor Current Exhalation Valve Assembly 10V Sentry Power Supply BPS Mode Signal 1. BPS PCB 2. BPS battery pack 6-61 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 338: Diagnostic Codes For Post Faults

    Table 6-2. Three or more 2. Replace AI PCB. resets not related KP0009 3. Replace BD CPU PCB. to AC failures 4. Replace mother- causes the test to board. fail. 6-62 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 339 3. Replace affected CPU PCB. EEPROM Verifies integrity of Major 1. Check the diagnostic Checksum flash memory codes’ corrective Test actions in Table 6-2. 2. Replace affected CPU KP0015 XP0015 XP0015 PCB. 6-63 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 340: Table 6-5. Bdu Only - Phase 2 Post Diagnostic Codes

    Tests DRAM by writing and Major 1. Check the diagnostic code’s verifying data patterns to corrective actions in Table 6-2. memory 2. Replace BD CPU PCB. KP0084 Tested in Short POST 6-64 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 341 1. Check the diagnostic code’s corrective actions in Table 6-2. 2. Turn off ventilator. Release switch. LP0121 3. Check for mechanical obstruc- tion to switch. 4. Replace BD CPU PCB. Tested in Short POST 6-65 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 342 Table 6-2. 2. Check cabling from BD CPU PCB to GUI CPU PCB. KP0112 3. Replace BD CPU PCB. 4. Call Technical Support or your Puritan Bennett representative. Tested in Short POST 6-66 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 343 Phase 2 NOVRAM Compares NOVRAM Minor 1. Check the diag- Test data to valid states nostic code’s cor- and ranges rective actions in Table 6-2. ZP0088 ZP0088 2. Replace GUI CPU PCB. 6-67 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 344 Table 6-2. will cause a VENT XP0087 or 2. Check associated XP0087 or ZP0087 ZP0087 INOP condition. errors in System Diagnostic Log. 3. Call Technical Sup- port or your Puritan Bennett representative. 6-68 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 345: Table 6-7. Bdu Or Gui-Phase 3 Post Diagnostic Codes

    Application Loads applica- Major 1. Check the diag- Initialization* tion software nostic codes’ cor- rective actions in Table 6-2. XP0131 2. Replace affected KP0131 XP0131 CPU PCB. Tested in Short POST 6-69 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 346: Post Interrupt Errors And Test Failures

    Loss of AC power during (BDU only) POST 5 VDC Interrupt Power Supply overvoltage (GUI only) 12 VDC Interrupt Power Supply overvoltage (GUI only) 12 VDC Interrupt Power Supply undervoltage (GUI only) 6-70 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 347 GUI alarm or alarm cabling Access fault/Bus Interrupt CPU PCB error Other Interrupt CPU PCB Unexpected interrupt POST Internal Interrupt CPU PCB Software Failure Test Failed Failure Specific analog device or CPU PCB 6-71 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 348: Sst And Est Test Sequences And Diagnostic Codes

    7. Switch Q1 and Q2. Rerun the flow sensor cali- bration and EST. If the failure transfers to the air side, return Q2 back to its original position and replace Q1. 6-72 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 349 2. Run EST Gas Supply Test to check for PSOL1 check failed leak. 3. Remove and then reseat PSOL1. 4. Replace PSOL1. 5. Perform a flow sensor calibration. 6. Replace Q1. 6-73 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 350 7. Switch Q1 and Q2. Rerun the flow sensor cali- bration and EST. If the failure transfers to the oxygen side, return Q1 back to its original position and replace Q2. 8. Replace Q3. 6-74 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 351 6. Switch Q1 and Q2. Rerun the flow sensor cali- bration and EST. If the failure transfers to the oxygen side, return Q1 back to its original position and replace Q2. 7. Replace Q3. 6-75 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 352 2. Verify that the air supply or compressor is available and adequate. 3. Verify that the ventilator system has no leaks or occlusions. If not sure, run the first four tests in EST to run the leak test. 6-76 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 353 3. If the leak test passes in EST, exit out of EST. Review the data for the EST Circuit Pressure Test to determine which pressure transducer is out of range: PI or PE. Replace applicable pressure transducer. 4. Replace AI PCB. 6-77 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 354 Prompts user to disconnect circuit from expiratory filter. Test: Establishes 60 L/min flow of either gas. Failure: Unable to deliver stable gas flow. FS0301 Make sure gas supply is connected. Unable to establish flow 6-78 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 355 2. Replace filter. Low expiratory filter P Failure: Pressure drop across expiratory filter FS0310 1. Repeat test, following directions more closely. < 0.1 cmH 2. Replace filter. Low expiratory filter P 6-79 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 356 Failure: Inspiratory limb pressure < 0.2 cmH FS0408 1. Make sure inspiratory filter is installed. 2. Repeat test, following directions more closely. Insp limb 3. Replace inspiratory filter. resistance low 4. Replace patient circuit 6-80 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 357 O intervals from 5 to 100 cmH O. Uses air as default gas. Failure: System cannot pressurize to one of the FS0501 Check patient circuit for leaks. Replace as test pressure points. necessary. Unable to pressurize circuit 6-81 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 358 Failure: Low compliance > 12 mL/cmH O (adult), FS0508 1. Make sure correct patient tubing type was 9 mL/cmH O (pediatric), or 3 mL/cmH specified in SST. Excessive (neonatal). 2. Replace patient circuit. compliance 6-82 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 359 Failure: Expiratory pressure transducer ADC FE0002 1. Replace SOL2. count at 0 cmH O is out of range. 2. Replace exhalation transducer Expiratory PCB. autozero out 3. Replace AI PCB. of range 6-83 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 360 2. Check for system leak at the expi- test pressure if test fails) ratory bacteria filter or O sensor. If not sure, run the first four tests in EST to run the leak test. 6-84 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 361 60, 5, and 1 L/min. nected. Unable to 2. Check regulated oxygen pressure. establish O 3. Replace PSOL1 or oxygen flow flow sensor (Q1). 4. Replace AI PCB. 6-85 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 362 2. Run EST Gas Supply Test to check zero flow for PSOL1 leak. Remove and then check failed reseat PSOL1 3. Replace PSOL1. 4. Perform a flow sensor calibration. 5. Replace Q1. 6-86 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 363 5. Run a flow sensor calibration. 6. Switch Q1 and Q2. Rerun the flow sensor calibration and EST. If the failure transfers to the oxygen side, return Q1 back to its original position and replace Q2. 6-87 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 364 2. Check PS2: Remove the wiring open. Wall air connectors from PS2 and jumper pressure not the connectors to each other. If detected the system now recognizes the air supply, replace PS2. 6-88 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 365 1. Make sure To patient port is blocked. SV pressure 2. Check for leaks at the O sensor, relief failed SOL1 or PI. 3. Replace the safety valve. 4. Replace PI. 6-89 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 366 PCB. detected 3. Replace PS2. Test: Verifies that compressor is depressurized by reading compressor pressure transducer (PC). Failure: PC indicates that compressor is FE0212 Replace compressor PCB. pressurized. Compressor pressure detected 6-90 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 367 (pressure switch PS1 closed). Failure: PS1 open instead of closed. FE0211 1. Make sure oxygen supply is con- nected. pressure 2. Replace Inspiratory Electronics not detected PCB. 6-91 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 368 FE0301 1. Make sure test circuit is con- 85 cmH nected. Excessive leak 2. Check for system leaks. 3. Check exhalation valve operation. 4. Verify secure connection of expira- tory filter. 6-92 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 369 15 s. 2. Replace keyboard. Exp. Pause key fails. Failure: Wrong key pressed or key not pressed FE0405 1. Repeat test. within 15 s. 2. Replace keyboard. Man Insp fails. 6-93 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 370 Failure: CLEAR key pressed to indicate LED not AE0601 1. Check/replace interconnect cable between the GUI LED PCB and the GUI High GUI CPU PCB. Alarm LED 2. Replace the GUI LED PCB. fails. 6-94 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 371 Failure: CLEAR key pressed to indicate LED not AE0611 1. Check/replace interconnect cable between the GUI LED PCB and the GUI Screen GUI CPU PCB. Lock LED fails. 2. Replace keyboard. 6-95 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 372 1. Make sure test equipment is con- stuck off when it should be on. nected properly to remote alarm Nurse call port. stuck off 2. Replace GUI CPU PCB. Deactivates nurse’s call port. 6-96 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 373 Failure: RC constant < 60 s. FE0903 1. Make sure BD alarm cable is con- nected. Bad power fail 2. Replace BD alarm. 3. Replace AI PCB. 6-97 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 374 If the problem transfers to the other gas side, return Q2 to its original position and replace Q1. 5. Replace the AI PCB. 6. Replace the Inspiratory Electronics PCB. 6-98 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 375 Insp check inspiratory check valve (CV3) may be damaged 2. Make sure CV3 is not installed valve test or incorrectly mounted. backward. failed 3. Replace CV3. 6-99 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 376 (Q3). occlusions. establish exp 2. Make sure air supply is still con- flow nected. 3. Run flow sensor calibration. 4. Replace Q3. 5. Replace Q2. 6. Replace the exhalation valve. 6-100 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 377 Failure: Measured system pressure at one or FE1401 1. Calibrate exhalation valve. more test points is out of range. 2. Replace exhalation valve. 3. Replace AI PCB. calibration 4. Replace the Exhalation Pressure Transducer PCB. 6-101 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 378 3 °C below heater- temperature) Failure: Heater temperature did not drop FE1503 1. Replace exhalation heater. sufficiently during an interval after heater was 2. Replace AI PCB. Bad exp turned off. heater 6-102 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 379 Use leak detector to check for leaks at the accumulator fittings and other tubing connections. b. Verify no leaks at CV2 within the ventilator by plugging the air inlet fitting. 3. Replace compressor PCB. 6-103 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 380 Failure: Compressor load test cannot be run, AE1610 Troubleshoot code AE1603 or because of inability to verify PC functionality. AE1606. Unable to perform compressor load test 6-104 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 381 Failure: PS2 is open, but it should be closed. AE1703 1. Make sure air supply is connected. Compressor control is restored. 2. Run Gas Supply/SV Test. Wall air pressure not detected 6-105 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 382 Test: Activates GUI serial port loopback mode. Transmits a character string. Verifies that received string is identical to transmitted string. Failure: Received message not identical to AE2001 Replace GUI CPU PCB. transmitted message. Bad GUI serial port 6-106 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 383 Battery not AC is good. charging 2. Replace BPS PCB. 3. Replace the battery pack. 4. Replace the power supply. 5. Replace the BD CPU. 6. Replace the AI PCB. 6-107 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 384: How To Troubleshoot Vga Lcd Controller Pcb Faults (Ut0002) (9.4-Inch Gui Only)

    Upper 0x028X XXXX 0x029X XXXX 0x02AX XXXX 0x02BX XXXX 0x02CX XXXX 0x02DX XXXX 0x02EX XXXX 0x02FX XXXX The example given indicates that the upper VGA LCD Controller PCB should be replaced. 6-108 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 385: How To Troubleshoot Gui Lcd Screen Messages

    The following examples provide scenarios and methods for troubleshooting these messages. NOTE: Puritan Bennett offers an in-depth training course on the Puritan Bennett 840 Ventilator System, that includes error troubleshooting and repair. Call Puritan Bennett Technical Support at 1.800.255.6774 for more information.
  • Page 386 BD and GUI serial numbers are stored in the data key. The appearance of the “Software Property of Puritan Bennett” message indicates that the data key information is not being read by the GUI.
  • Page 387: Example 2: No Communication Between The Gui And Bdu

    BDU and the GUI (EST, SST, and calibration data did not get uploaded to the GUI). The “GUI timeout occurred” and “Software Property of Puritan Bennett” messages are further evidence that the GUI and BDU are not communicating.
  • Page 388: Example 3: Post Or Background Check Errors

    Table 6-8 for POST diagnostic LED patterns). Replace indicated components based upon recommended diagnostic code corrective actions. NOTE: If the message “BD TUV failed” appears on the upper GUI screen, replace the BD CPU PCB. 6-112 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 389: Alarm Handling

    7.1 Alarm classifications Alarms in the Puritan Bennett™ 840 Ventilator System are classified by urgency level; this classification determines how the ventilator responds (Table 7-1). Some alarms are triggered by a ventilator setting or patient condition, and they can occur in the usual course of patient care.
  • Page 390: Responding To Alarms

    If necessary, clear the message or reset the alarm by pressing the alarm reset key. For specifics about alarm operation, see the Puritan Bennett 800 Series Ventilator System Operator’s and Technical Reference Manual.
  • Page 391: Figure 7-2. Alarm Log

    APNEA The set apnea interval has Check patient and settings. elapsed without the ventilator, (lockable) patient, or operator triggering a breath. Resets when patient initiates 2 consecutive breaths. Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 392 % setting for at least 30 s. (These percentages increase by 5% for 4 minutes following a decrease in the O % setting.) Alarm updated at 1-s intervals. Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 393 Alarm updated whenever 2. Run EST. (lockable) an exhaled minute volume is recalculated. Delivered inspiratory volume Check patient and settings. TI MAND high inspiratory volume limit (lockable) (breath type = VC+) Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 394 2. Check connections of AC power distribution compo- nents. LOW INSP Inspiratory target pressure 1. Check patient.  (PEEP + 5cmH PRESSURE O) when 2. Check target volume. mandatory type is VC+ (lockable) Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 395 Alarm updated whenever (lockable) exhaled minute volume is  set Check patient E TOT recalculated. limit for > 20s. & settings.  set Check patient E TOT limit for > 40s. & settings. Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 396 2. Replace and calibrate O failure. Press Ventilation is unaffected. User sensor. 100% O may also disable FIO monitor. 3. Disable O sensor. or INCREASE 2 min key, replace, or disable. Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 397 Last 2 mand and setting for volume. DELIVERED breaths, pressure > (This alarm PEAK max allowable applies to VC+ and VS level breaths.) Last 10 or more mand breaths, pressure > max allowable level Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 398 Alarm handling This page intentionally blank. 7-10 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 399: Service And Repair

    Lock cart in place by locking casters prior to servicing any part of the ventilator. • When servicing the Puritan Bennett™ 840 Ventilator System, be sure to familiarize yourself with, and adhere to all posted and stated safety warning and caution labels on the ventilator and its components, and on any service equipment and materials used.
  • Page 400: General Repair Guidelines

    The following cleaning guidelines pertain to parts that require cleaning while servicing the ventilator. For procedures on periodic cleaning and sterilization of the ventilator and accessories, refer to the Puritan Bennett 800 Series Ventilator System Operator’s and Technical Reference Manual. For periodic cleaning and inspection procedures that are required during a ventilator's performance verification, refer to Chapter 5 in this manual.
  • Page 401: Electrical Cables And Pneumatic Tubing

    Connect compressed gas sources to ventilator. Using small brush, apply leak detector fluid (P/N 4-004489-00) to pneumatic connections. If a connection leaks, repair and retest module. Dry all leak detector fluid from chassis. Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 402: Electrostatic Discharge Control

    8.8.1 ESD procedures and precautions Follow these procedures and precautions to prevent ESD damage to the ESD-sensitive microelectronic components and assemblies in the Puritan Bennett 840 Ventilator System. • Use a personnel-grounding system. Before opening the ventilator lid or removing its...
  • Page 403: Table 8-1. Testing And Calibration Requirements

    Unless the ventilator meets all of these requirements, PVT must be performed following the repair. 4. NeoMode only. 5. Perform Compact Flash Test if Trending software option is installed or compact flash component replaced. 6. Perform EST and electrical safety testing only. Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 404: Patient System And Accessories

    Service and repair 8.11 Patient system and accessories To service the patient system and accessories, consult the Puritan Bennett 800 Series Ventilator System Operator’s and Technical Reference Manual or applicable accessory manuals. Warning Connectors and tubes with the proprietary Bennett barbed cuff fittings are intended for use only with like fittings.
  • Page 405: Figure 8-1. Gui

    Table 8-3: GUI part replacement chart GUI replacement part (applies to all GUI configurations GUI mounted GUI detached except where noted) Bezel and gasket (9.4-inch only) Window and gasket (9.4-inch only) GUI (Ethernet) cable Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 406: Repairing The 10.4-Inch Gui

    (1/4-inch) blade slotted screwdriver (smaller screwdrivers may not be adequate to tighten the connector). Firmly press the cable connector again onto the alignment posts and D-sub connector, and re-tighten the jackscrews until snug. Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 407: Removing Or Installing The 10.4-Inch Gui When Mounted On Puritan Bennett

    Lock cart in place by locking casters prior to servicing any part of the ventilator. If GUI is mounted on optional Puritan Bennett 800 Series Ventilator Compressor Mount Cart or Puritan Bennett 800 Series Ventilator Pole Cart, follow these steps: Disconnect GUI interface cable from side of BDU (Figure 8-3) Detach cable from p-clip on underside of cart handle (Figure 8-4).
  • Page 408: Figure 8-4. P-Clip And Gui Cable Installed On Puritan Bennett 800 Series Ventilator Pole Cart

    Service and repair p-clip Figure 8-4. P-clip and GUI cable installed on Puritan Bennett 800 Series Ventilator Pole Cart Remove GUI from its mount: a. Grasp handle at back of GUI. b. Push GUI release knob to right to disengage rotor housing from mounting platform.
  • Page 409: Removing The 10.4-Inch Gui Cable Assembly

    Using a small flat blade screwdriver, unscrew the captive screws securing the GUI cable. Gently remove the GUI cable and set it aside. Top handle interface cable Bottom Diagnostic handle LED array Figure 8-5. Removing and replacing 10.4-inch GUI cable assembly 8-11 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 410: Replacing The 10.4-Inch Gui Cable Assembly

    If you use a small flat-blade screwdriver to separate the rear and front housings, take care not to damage the GUI front and rear panels. Figure 8-6. Removing 10.4-inch GUI rear housing 8-12 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 411: Replacing The 10.4-Inch Gui Rear Housing

    EMI shield is present. Ensure that this strap is fastened to the EMI shield as shown in Figure 8-8 upon reassembly. Lift the shield from the bottom and tilt up. Gently remove the shield from the GUI assembly and set aside. 8-13 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 412: Removing The 10.4-Inch Gui Backlight Inverter Pcbs

    For each of the backlight inverter PCBs, disconnect the harness that connects the LCD to the backlight inverter and the cable from the backlight inverter to the GUI CPU PCB (Figure 8-9). Remove each of the PCBs and place into a conductive bag. 8-14 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 413: Removing The 10.4-Inch Gui Cpu Pcb

    If you are replacing a 10.4-inch GUI CPU PCB with a TE GUI CPU PCB, properly dispose of the old RS-232 shield, four female screw-lock screws, and flex circuit; otherwise retain for reassembly. 8-15 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 414: Replacing The 10.4-Inch Non-Te Gui Cpu Pcb With The Te Gui Cpu Pcb

    • a new RS-232 shield with added EMI gaskets • longer female screw-lock screws and lock washers • addition of a thermally conductive pad to the CPU heatsink • compact flash memory required for the Puritan Bennett 840 Ventilator System Trending software option 8-16...
  • Page 415: Figure 8-12. Socket Savers And Spacers Installed Onto Rs-232 Connectors

    2.5 in 3 in Figure 8-13. RS-232 shield with gaskets installed Route the nurse call connector dust cap through the hole in the RS-232 shield while installing the shield onto the PCB. 8-17 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 416: Figure 8-14. Thermal Pad Adhered To Cpu Heatsink

    • GUI LED PCB cable (J3) • alarm speaker cable (J7) • touchframe PCB cable (J19) • keyboard cable (J20) Install the backlight inverter PCBs by reversing the steps in Section 8.13.8. 8-18 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 417: Replacing The 10.4-Inch Te Gui Cpu Pcb With A New 10.4-Inch Te Gui Cpu Pcb

    Slide the component out of the socket (J28) towards the edge of the PCB (Figure 8-15). Insert the new component into the socket. The socket is designed so that the component can be inserted only one way. Figure 8-15. How to replace the compact flash component 8-19 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 418: Replacing The Real-Time Clock

    The GUI support bracket secures the GUI CPU PCB on one side and the LCD panels on the other. The bracket must be removed to access the LCD panels, the touchframe PCB, and the inside front of the GUI housing. Refer to Figure 8-17 for parts identification. 8-20 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 419: Removing The 10.4-Inch Gui Support Bracket

    • Do not allow the LCD panels to be exposed to the ambient environment any longer than necessary as dirt and debris will be deposited on the LCD panels. Should this occur, use an ESD-protected air source to gently blow away any dust or debris. 8-21 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 420: Removing The 10.4-Inch Lcd Panels

    US English software revision P or later (or an equivalent version for your country’s language), you may replace a single LCD panel, as drivers for both new and old LCD panels are included in these versions. Contact Puritan Bennett Technical Support at 1.800.255.6774 to determine the correct software version for languages other than US English.
  • Page 421: Replacing The 10.4-Inch Touchframe Pcb

    Connect the GUI LED cable to the GUI LED PCB. 8.13.19 10.4-inch GUI alarm assembly The GUI alarm is located behind the GUI front panel and can only be accessed by removing the rear GUI housing. 8-23 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 422: Removing The 10.4-Inch Gui Alarm Assembly

    GUI housing to attach the grounding strap to the EMI shield. An in-line connector allows subsequent keyboard replacement without having to remove the rear housing. 8-24 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 423: Replacing The 10.4-Inch Gui Keyboard Assembly

    Connect the ribbon cable from the GUI CPU PCB to the keyboard. Verify that the ribbon cable is in the upper row of the keyboard connector and is firmly latched into place. Remove the adhesive cover strips from the keyboard. 8-25 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 424: Inch Gui Front Housing

    Using a Phillips screwdriver, remove the two screws that secure the fascia panel in place. Set the bezel and the screws aside. Using pliers, carefully pull the spring away from the rotor brake shoes. Let the spring hang from the base. 8-26 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 425: Figure 8-22. 10.4-Inch Gui Rotor Housing

    Install the two brake shoes and align the spring to fit in the brake shoe groove. Align the fascia panel over the rotor and secure with the two screws. Tighten until snug. Reinstall the GUI onto the cart. 8-27 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 426: Repairing The 9.4-Inch Gui

    The window is treated for glare resistance and touching it creates smudges which cannot be removed without damaging the window. Bezel Gasket Screwdriver placement Window Window gasket panels Figure 8-23. 9.4-inch touch screen bezel and window 8-28 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 427: Inch Gui Window

    Making sure gasket is in place and coated side of window is facing GUI, carefully set window into window frame (Figure 8-23). Carefully insert window gasket into groove around uncoated (front) side of window. Gently press edges of window until it fully seats into front housing. 8-29 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 428: Installing The 9.4-Inch Bezel

    Once the adhesive has been broken, disconnect ribbon cable from keyboard. Continue to pull keyboard assembly away from front housing. Remove entire assembly including keyboard laminate, keyboard PCB, and rotary encoder. 8-30 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 429: Figure 8-25. Replacing The 9.4-Inch Keyboard Assembly

    Using alcohol, clean any remaining adhesive from front housing, being careful not to let debris fall into the GUI housing. NOTE: Do not attempt to reuse the 9.4-inch keyboard assembly after removal. Keyboard assembly Keyboard laminate fig8-19c.tif Figure 8-25. Replacing the 9.4-inch keyboard assembly 8-31 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 430: Installing The 9.4-Inch Keyboard Assembly

    (1/4-inch) blade slotted screwdriver (smaller screwdrivers may not be adequate to tighten the connector). Firmly press the cable connector again onto the alignment posts and D-sub connector, and re-tighten the jackscrews until snug. 8-32 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 431: Inch Gui Rear Housing

    Remove the two screws that secure bottom handle to GUI rear housing and set aside. Top handle Interface cable Bottom handle Captive screw (2 PL) Interface cable connector Figure 8-27. Removing 9.4-inch GUI handle assembly and interface cable 8-33 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 432: Installing 9.4-Inch Gui Rear Housing

    Gently press rear housing into front housing until two enclosures fully seat. Install GUI rear housing (Section 8.14.6.1). Caution Ensure cables are not pinched between the front and rear housings. Otherwise, damage to the cables may result. 8-34 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 433: Inch Gui Alarm Assembly

    J7 (J9 older GUI CPU PCB) J5 (J7 older GUI CPU PCB) J4 (J6 older GUI CPU PCB) Alarm assembly Backlight inverter Figure 8-29. 9.4" GUI interior (TE GUI CPU PCB shown) 8-35 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 434: Installing 9.4-Inch Gui Alarm Assembly

    Lift backlight inverter PCB off standoffs and place into conductive bag or on ESD-protected mat. If you are installing a new backlight inverter PCB, remove backlight inverter PCB/GUI CPU PCB ribbon cable and save for reinstallation. 8-36 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 435: Installing 9.4-Inch Gui Backlight Inverter Pcb

    Manually install and tighten four standoffs; then using a 1/4-in. nutdriver, tighten until snug. Do not overtighten. Connect GUI LED PCB ribbon cable to GUI CPU PCB connector J3 (Figure 8-29). 8-37 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 436: Gui Electromagnetic Interference (Emi) Shield

    13 screws; then, using a screwdriver, tighten until snug. Do not overtighten. Connect these cables to bottom of GUI (Figure 8-28): • Touchframe PCB ribbon cable to GUI CPU PCB connector J19 • Keyboard ribbon cable to GUI CPU PCB connector J20 8-38 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 437: Inch Video Controller And Vga Lcd Controller Pcbs

    Orient pins on VGA LCD controller PCB so they align with corresponding connectors on GUI CPU PCB (Figure 8-33). Gently press VGA LCD controller PCB until fully seated. Install GUI EMI shield (Section 8.14.9.2). Install GUI rear housing (Section 8.14.6.2). 8-39 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 438: Inch Gui Cpu Pcb

    GUI CPU PCB TO “Carrol Touch” (A-B position) as shown in Figure 8-34. Jumpers Figure 8-34. GUI CPU PCB touch panel jumpers 8-40 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 439: Figure 8-35. Thermal Pad Adhered To Cpu Heatsink

    GUI PCB FRU kit. Peel the remaining adhesive backing from the thermal pad. Thermal pad GUI CPU heatsink Figure 8-35. Thermal pad adhered to CPU heatsink 8-41 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 440: Replacing The Compact Flash Component (Te Gui Cpu Pcb Only)

    Figure 8-36. How to remove the compact flash component 8.14.13 Replacing the real-time clock The real-time clock component on the GUI CPU PCB has an internal battery with an expected life of approximately ten years. 8-42 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 441: Inch Touchframe Pcb (Carroll Touch)

    Turn the CPU bracket assembly (LCD panels up) and place on an ESD-protected mat. Remove the two screws that secure touchframe PCB to CPU bracket (Figure 8-38). Gently lift off touchframe PCB and place in conductive bag or on ESD-protected mat. 8-43 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 442: Reinstalling The 9.4-Inch Touchframe Pcb

    Turn the CPU bracket assembly over (LCD panels down) and place on an ESD-protected mat. Route the backlight extender cable assemblies as shown in Figure 8-39. Ensure that the foam side of the cable assemblies are facing away from the LCD panels. 8-44 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 443: Figure 8-39. Routing The Backlight Extender Cable Assemblies

    Standoff Plastic spacer Figure 8-39. Routing the backlight extender cable assemblies Install GUI CPU PCB/bracket assembly (Section 8.14.11.2). Install GUI EMI shield (Section 8.14.9.2). Install GUI rear housing (Section 8.14.6.2). 8-45 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 444: Inch Backlight Panels And Lcd Panels

    Reattach the backlight extender cable assembly, then verify that it is properly connected. Install and hand-tighten the four screws. Then, using a screwdriver, tighten the screws until they are snug. Do not overtighten. Verify backlight extender cable assembly is firmly connected. 8-46 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 445: Figure 8-40. Lcd Panel And Backlight Panel Assembly

    Backlight interface PCB panel Reattach to backlight cable xtender assembly (not shown) Bend and straighten metal tabs and secure backlight panel LCD panel Figure 8-40. LCD panel and backlight panel assembly 8-47 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 446: Inch Gui Cooling Vent Filters

    Ensure cables are not pinched between the front and rear housings as you reassemble the GUI unit. Otherwise, damage to the cables may result. Pushnut Foam filter Foam filter Pushnut Figure 8-41. Replacing the 9.4-inch GUI cooling vent filters 8-48 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 447: Inch Rotor Housing

    Position fascia panel; then install and hand-tighten two screws. Using a screwdriver, tighten until snug. Do not overtighten. Install GUI (Section 8.14.5). Fascia panel GUI rear panel Spring Blue rotor Brake housing shoes Figure 8-42. 9.4-inch rotor housing assembly 8-49 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 448: Breath Delivery Unit (Bdu)

    Table 2-2, and Figure 2-11 in Chapter 2 of this manual. Caution To prevent damage to ESD-sensitive components, always follow ESD guidelines when servicing the BDU. 8-50 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 449: Figure 8-43. Bdu

    (PSOL1/PSOL2) Inspiratory electronics PCB Oxygen sensor Flow sensors (Q1/ Safety valve Inspiratory check valve (CV3) Regulators (REG1/ REG2) Inspiratory pressure transducer autozero solenoid (SOL1) Exhalation module replacement parts Exhalation valve (EV) 8-51 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 450 (CV5) Sample port Other replacement parts Motherboard PCB BDU alarm assembly Alarm blindmate cable Inspiratory blindmate cable DC power supply blindmate cable AC power supply blindmate cable AC panel Humidifier receptacle 8-52 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 451: Removing Bdu From Rta Cart

    Figure 8-44. BDU connections 8.15.2 Removing BDU from Puritan Bennett 800 Series Ventilator Compressor Mount Cart or Puritan Bennett 800 Series Ventilator Pole Cart Make these disconnections (Figure 8-44): a. Ventilator power cord from facility power (Section 8.15.3.)
  • Page 452: Bdu Power Cord And Retainer

    Remove the two screws securing the current bracket to AC receptacle of the BDU. Install the Sure-Lock retainer over the receptacle and secure with the two screws provided. Insert the power cord and ensure that it clicks firmly into place. 8-54 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 453: Installing Bdu Onto Rta Cart

    Warning Install only ventilator BDUs with serial numbers starting with 3512 onto the newer Puritan Bennett 800 Series Ventilator Compressor Mount Cart and Puritan Bennett 800 Series Ventilator Pole Cart. Other ventilator serial numbers are not compatible with the newer carts.
  • Page 454: Analog Interface (Ai) Pcb And Breath Delivery (Bd) Cpu Pcb

    BD CPU PCB sits in second slot from the right (behind the AI PCB). Refer to Figure 8-47. Caution The data key is matched to each ventilator. Ventilator is inoperable without the correct data key installed. Do not lose the data key. 8-56 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 455: Removing Ai Pcb Or Bd Cpu Pcb

    Simultaneously apply negative pressure to disengage clips. (When the clips snap open, the PCB is released.) Gently pull PCB out of card cage, detaching it from motherboard PCB. Place PCB in a conductive bag or on ESD-protected mat. 8-57 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 456: Replacing The Real-Time Clock

    Figure 8-50. BD CPU PCB real-time clock Insert the new component, ensuring that pin 1 is properly oriented with the socket. Set the ventilator time and date after replacing the real-time clock. 8-58 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 457: Installing Ai Pcb Or Bd Cpu Pcb

    1/8-in. (3-mm) flat- bladed screwdriver. Remove two screws that secure power supply assembly to BDU chassis. Gently pull out power supply assembly, unseating it from connectors (Figure 8-51). Set assembly on firm surface. 8-59 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 458: Installing Power Supply Assembly

    Connect BPS cable to power supply by opening spring-loaded cover that shields power supply connector, connecting cable, and using 1/8-in. (3-mm) flat-bladed screwdriver to tighten connector's two captive screws until snug. Do not overtighten. 8-60 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 459: Power Switch (S1)

    Reaching into power supply compartment from rear, disconnect harness from power switch (Figure 8-52). If you are installing a new power switch, remove actuator and save for reinstallation. Power switch Figure 8-52. Replacing power switch (S1) 8-61 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 460: Installing Power Switch (S1)

    From front of ventilator, remove two flat-head screws that secure humidifier receptacle to chassis. Feed receptacle through power supply cavity. Remove receptacle assembly from chassis. 8-62 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 461: Installing Humidifier Receptacle

    If desired, install humidifier receptacle cover with four flat-head screws. Install power supply assembly (Section 8.15.7.2). NOTE: Consult this manual or the humidifier vendor for the appropriate bracket model when mounting a humidifier to the ventilator. 8-63 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 462: Ac Panel

    • Blue wire from middle right-hand terminal of compressor AC receptacle Using a 5/16-in. nutdriver, disconnect ground wires from ground stud of AC panel. Remove AC panel. Ground stud AC filter PCB Relay Compressor AC receptacle Figure 8-55. AC panel 8-64 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 463: Inspiratory Module

    8.15.10.2 Additional noise suppression (ferrite cores) Additional noise suppression has been added to the Puritan Bennett 840 Ventilator System, in the form of three ferrite cores on the air flow transducer harnesses (refer to Section 9.4.8.2 for part number).
  • Page 464: Inspiratory Module O-Rings

    Make sure air source is disconnected from ventilator. Drain any condensate from air inlet filter bowl by unscrewing captive drain plug at bottom (Figure 8-57). Tighten plug. Unscrew bowl. Unscrew filter. 8-66 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 465: Fascia Panel

    8.15.10.5 Fascia panel 8.15.10.5.1 Removing fascia panel Remove the fascia panel that houses the inspiratory module by removing two screws (Figure 8-58). Fascia panel Figure 8-58. Removing inspiratory module fascia panel 8-67 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 466: Figure 8-59. Oxygen Filter Assembly And Pressure Switches

    Remove sintered metal oxygen impact filter (F1) and spring. Oxygen filter assembly Air pressure Oxygen switch (PS2) pressure switch (PS1) Oxygen inlet Oxygen Bowl filter (F3) impact filter (F1) Spring Figure 8-59. Oxygen filter assembly and pressure switches 8-68 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 467: Oxygen And Air Pressure Switches (Ps1 And Ps2)

    Connect harness to top of switch. Make sure connectors are tight and that they make contact with switch. Install oxygen filter assembly (Section 8.15.10.5.4). Using leak detector fluid (P/N 4-004489-00), verify that bowl does not leak. Install fascia panel (Section 8.15.10.5.2). 8-69 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 468: Psol Cartridge

    Carefully pull cartridge out of PSOL manifold and set on a clean work surface. Verify that all four o-rings are present and in good condition. Nylon Nylon line line Oxygen PSOL (PSOL1) Air PSOL (PSOL2) Figure 8-61. Replacing PSOL cartridge 8-70 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 469: Removing Inspiratory Module

    Using small brush, apply leak detector fluid (P/N 4-004489-00) to these locations. • Base of air and oxygen flow sensors • Transfer tubes • Check valve assembly • Regulator inlet/outlet ports • Oxygen/air inlets If a connection leaks, repair and retest module. 8-71 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 470: 10Installing Inspiratory Module

    The oxygen sensor should be replaced every year or as often as necessary. Newer Puritan Bennett 840 Ventilator System BDU covers have an access port located on the top right edge of the BDU. This access port facilitates convenient access to the oxygen sensor for the purpose of replacement by medical staff.
  • Page 471: Figure 8-64. Detail Of Oxygen Sensor

    Using both thumbs, seat the port cover in the port by press the two outside corners of the lower flap at the juncture of the housing edge, fitting the into the housing opening. 8-73 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 472: Figure 8-65. Replacing Oxygen Sensor

    Insert the threaded end of the oxygen sensor into the cavity on the check valve housing and finger-tighten the sensor without using excessive force. As the sensor is installed, ensure that it is not cross-threaded. 8-74 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 473: 13Right-Side Plate

    Install CV3 housing to PSOL manifold with four screws. Connect harness to oxygen sensor. Install inspiratory module, as needed (Section 8.15.10.10). 8.15.10.13 Right-side plate This is the plate with the inspiratory electronics PCB attached. 8-75 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 474: Figure 8-67. Removing Inspiratory Module Right-Side Plate

    • One screw at front of plate • Two screws at rear of plate Remove right-side plate, with inspiratory electronics PCB attached, and place on ESD- protected mat. Figure 8-67. Removing inspiratory module right-side plate 8-76 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 475: 15Left-Side Plate

    If seal is damaged, refer to Chapter 9 for replacement adhesive tape. Remove one screw from rear of plate. Remove two screws from front of plate. One screw is under seal. Carefully lift left-side plate off inspiratory module. 8-77 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 476: Figure 8-68. Replacing Inspiratory Module Left-Side Plate

    Disconnect harnesses from inspiratory electronics PCB, as required. Disconnect SOL1/PSOL manifold interconnect tube from PSOL manifold. Gently pull manifold away from inspiratory module. The flow sensors should be attached to the manifold. Place manifold on clean work surface. 8-78 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 477: Figure 8-69. Psol Manifold Ready For Removal

    (Figure 8-69). Tighten screws. Do not overtighten. If third flat-head screw on left-side plate was loosened to install PSOL manifold, tighten until snug. Connect tube attached to top of SOL1 to beneath PSOL manifold body. 8-79 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 478: Figure 8-70. Replacing Oxygen And Air Flow Sensors (Q1 And Q2) And Safety Valve

    Remove end caps from new sensor tube, and transfer to existing flow sensor tube. PSOL manifold Safety valve Air flow sensor (Q2) Oxygen flow sensor (Q1) Figure 8-70. Replacing oxygen and air flow sensors (Q1 and Q2) and safety valve 8-80 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 479: Figure 8-71. Oxygen And Air Flow Sensors (Q1 And Q2)

    8.15.10.18.1 Removing safety valve (SV) Remove oxygen and air flow sensors (Section 8.15.10.17.1). Remove two flat-head screws that secure safety valve (Figure 8-70) to PSOL manifold. Gently pull safety valve (Figure 8-72) out of manifold. 8-81 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 480: Figure 8-72. Safety Valve

    Install two flat-head screws that secure safety valve to PSOL manifold. Tighten until snug. Do not overtighten. Install oxygen and air flow sensors (Section 8.15.10.17.2). Caution To prevent component damage when installing the safety valve, do not apply leak detector fluid. 8-82 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 481: 19Check Valve Assembly, Regulator Assembly, And Flow Sensor Manifold

    Install two screws that secure check valve assembly to floor. Tighten screws until snug. Do not overtighten. Install PSOL manifold (Section 8.15.10.16.2). Install right-side plate (Section 8.15.10.13.2) and left-side plate (Section 8.15.10.15.2). 8-83 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 482: Figure 8-73. Disassembling Flow Sensor Manifold/Regulator Assembly/Check Valve Assembly

    Service and repair Regulator assembly Check valve assembly Flow sensor manifold Inspiratory plate Inspiratory floor plate Regulator assembly Check valve Flow assembly sensor manifold Figure 8-73. Disassembling flow sensor manifold/regulator assembly/check valve assembly 8-84 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 483: Figure 8-74. Removing Reg1 And Reg2

    (Figure 8-74). Using 7/16-in. socket or open-end wrench, install retaining nut and tighten until snug. If you are installing REG1, connect vent tube to REG1 output port, at bottom of inspiratory floor plate. Install regulator assembly with plate (Section 8.15.10.19.2). 8-85 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 484: 22Inspiratory Floor Assembly

    From top of gas inlet manifold, install two screws that secure floor to manifold. Tighten screws until snug. Do not overtighten. Install flow sensor manifold/regulator assembly/check valve assembly to inspiratory floor assembly (Section 8.15.10.19.2). 8-86 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 485: Exhalation Module

    Push the blue latch up to release collector vial/expiratory filter. Remove assembly, and disassemble as required. For complete details on using the collector vial and expiratory filter, consult the Puritan Bennett 800 Series Ventilator System Operator’s and Technical Reference Manual.
  • Page 486: Figure 8-75. Removing Exhalation Module Cover

    Service and repair Remove screw Figure 8-75. Removing exhalation module cover 8-88 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 487: Figure 8-76. Exhalation Module Disassembled

    Align cover screw locations to BDU chassis (three on front and one on rear); then tighten screws until snug. Do not overtighten. Install collector vial and expiratory filter (Section 8.15.11.1). 8-89 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 488: Removing Exhalation Module

    Do not overtighten. Connect exhalation module harness to exhalation module I/O cable. Be sure clips on each side of cable latch into place. Install exhalation module cover (Section 8.15.11.2). 8-90 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 489: Exhalation Valve (Ev)

    (Figure 8-78). This sample port has since been eliminated and a different style exhalation valve has been implemented. This new exhalation valve can be used on original Puritan Bennett 840 Ventilator Systems with the metabolic sample port, but a minor modification to the exhalation module is required.
  • Page 490: Figure 8-79. Replacing Exhalation Module Components

    Expiratory heater pressure assembly transducer autozero solenoid (SOL2) Exhalation flow sensor (Q3) Exhalation valve (EV) Exhaust Exhalation port and tube module harness Exhalation transducer PCB Figure 8-79. Replacing exhalation module components 8-92 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 491 Remove the exhalation module cover. (Section 8.15.11.2) Remove the exhalation exhaust port tube. Using two cotton swabs at a time, wet the swabs with a 50% mixture of isopropyl alcohol and sterile water. 8-93 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 492: Exhalation Flow Sensor (Q3)

    When disconnecting Q3, be careful not to jar or drop it. Remove exhalation valve (Section 8.15.11.5.3). On Puritan Bennett 840 Ventilator Systems with the EMI shielded flow sensor connector, remove the screws and washers holding the ground shield and EMI top and bottom shields together.
  • Page 493: Figure 8-80. Exhalation Flow Sensor (Q3)

    Ensure that no damage to the o-ring occurs during installation of Q3. Connect cable to connector on Q3. On Puritan Bennett 840 Ventilator Systems with an EMI shield and EMI exhalation harness assembly, install the shield so that it encloses the replacement EMI harness assembly flow sensor connector, making sure that the harness and the ground shield cable are routed as shown in Figure 8-81 and Figure 9-18.
  • Page 494: Exhalation Transducer Pcb

    Disconnect transducer cable from exhalation transducer PCB (Figure 8-82). Disconnect tube from PCB transducer port. Remove three screws that secure exhalation transducer PCB to exhalation module plate. Remove PCB and place in conductive bag or on ESD-protected mat. 8-96 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 495: Expiratory Pressure Transducer Autozero Solenoid (Sol2)

    Connect tube to transducer port. Connect exhalation transducer PCB cable to exhalation transducer PCB. Later model Puritan Bennett 840 Ventilator Systems include a USB type connector on the Exhalation transducer PCB and cable. Ensure that you are using the correct configuration for your ventilator.
  • Page 496: Exhalation Heater (Exh Htr) And Check Valve (Cv5)

    Make sure leaf is lying flat, then place check valve (Figure 8-83) (leaf side up) into groove of seal surrounding orifice. Make sure o-ring at top of flow sensor tube is in place. 8-98 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 497: Bdu Housing

    Slide exhalation I/O cable connector back into chassis. To accomplish this, it is helpful to slide your hand between card cage body and plastic rear housing of exhalation compartment. This allows you to maneuver plastic housing when positioning connector to slide through slot. 8-99 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 498: Figure 8-84. Removing Bdu Housing

    Service and repair Power switch Power indicator Gasket AC panel Exhalation I/O cable connector Figure 8-84. Removing BDU housing 8-100 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 499: Installing Bdu Housing

    Remove seven screws that secure motherboard PCB to back panel of card cage (Figure 8-85). Gently remove motherboard PCB through card cage opening. Place PCB in a conductive bag or on ESD-protected mat. 8-101 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 500: Bdu Alarm Assembly

    8.15.12.4.1 Removing BDU alarm assembly Remove BDU housing (Section 8.15.12.1). Remove two screws that secure harness to BDU alarm assembly (Figure 8-86). Unscrew ring at front of alarm. Remove alarm from mounting bracket. 8-102 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 501: Power Indicator

    BPS cable from the power supply assembly before servicing AC components. 8.15.13.1 Removing power indicator Remove BDU housing (Section 8.15.12.1). Remove screw that secures sleeve to indicator (Figure 8-52). Remove insert and sleeve; save for reinstallation. Push indicator through opening. 8-103 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 502: Installing Power Indicator

    Pull these two wires back through grommet and through blindmate cable opening at rear of power supply compartment cavity. Using 1/8-in. (3-mm) flat-bladed screwdriver, loosen two captive screws that secure cable to motherboard PCB connector (Figure 8-86). Disconnect cable. 8-104 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 503: Inspiratory Blindmate Cable

    Using 11/32-in. nutdriver or socket wrench, remove three nuts that secure cable to chassis (Figure 8-89). Gently feed firewall gasket into inspiratory compartment. Feed rest of cable through power supply module compartment into inspiratory module, and remove. 8-105 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 504: Dc Power Supply Blindmate Cable

    Remove inspiratory blindmate cable (Section 8.15.13.4.1). b. Remove four chassis-retaining screws from rear of AC panel. Remove two screws that secure cable (Figure 8-90) to chassis. Feed both connectors into power supply compartment, and remove cable. 8-106 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 505: Ac Power Supply Blindmate Harness

    Disconnect harness from AC filter PCB and ground stud in AC panel. Remove two flat-head screws securing harness bracket to front of chassis. Remove harness (Figure 8-91) through power supply compartment. 8-107 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 506: Figure 8-91. Replacing Ac Power Supply Blindmate Harness

    Using two small tie wraps, rebundle wires on either side of round grommet. Cut tie wrap ends flush with the tie wrap body. Install four chassis-retaining screws at rear of AC panel. Install BDU housing (Section 8.15.12.2). 8-108 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 507: Release Handle

    Replacement parts for the original compressor, model 804, are no longer available. Contact Puritan Bennett Technical Support for additional information. Be sure to read all applicable instructions completely before disassembly. 8-109 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 508: Servicing The 806 Compressor

    Gently tuck in the edges of the filter. 8.17.3 Removing compressor from cart 8.17.3.1 Disconnecting compressor from BDU Disconnect the ventilator power cord from the wall outlet. Disconnect the high pressure hoses. 8-110 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 509: Removing Compressor From Cart

    Loosen the two captive screws under the bottom shelf at the rear of the cart. (These screws secure the compressor to the cart.) If you are removing the compressor from the Puritan Bennett 800 Series Ventilator Compressor Mount Cart, remove the bezel from the front of the BPS assembly.
  • Page 510: Figure 8-93. Top Cover Removed From Compressor

    Service and repair Figure 8-93. Top cover removed from compressor 8-112 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 511: Installing Top Cover

    Facing the front of the compressor, remove the four (4) Phillips screws and star washers that secure the fan to the top of the plenum assembly. Disconnect the electrical cable from the fan. Remove the fan. Figure 8-94. Fans installed in compressor 8-113 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 512: Installing A Fan

    Replace the six (6) Phillips screws from the sides of the back panel. Reinstall the top cover (Section 8.17.5). Four (4) nuts Fitting Accumulator intake port o-ring (4) Figure 8-95. Removing the back panel of the compressor 8-114 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 513: Replacing An Accumulator Fitting And O-Ring

    Using a 3/16" flathead screwdriver, loosen the screw on the hose clamp that secures the pressure transducer to the compressor PCB. Disconnect the tube from the PCB. (See Figure 8-98.) Disconnect the main supply hose at the solenoid filter outlet port. (See Figure 8-96.) 8-115 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 514: Plenum Assembly

    The plenum assembly is accessed by removing the top cover and accumulator assembly, which is mounted to the back panel, and the electrical connections from the compressor PCB. 8-116 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 515: Removing The Plenum Assembly

    Disconnect the hose at the heat exchanger outlet port that runs to the water trap inlet. Taking care not to damage the heat exchanger fins, carefully lift the plenum assembly from the top of the compressor and set assembly aside. 8-117 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 516: Heat Exchanger (He)

    Secure the data ground cable to the plenum and tighten the nut. Attach the data cable to the PCB and slide its lock into place. Attach the cable to the plenum assembly with a tie wrap. 8-118 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 517: Compressor Pcb

    The air dryer assembly is accessed by removing the back panel assembly and disconnecting the supply tubing. The filters are normally replaced as part of the 15,000 hour PM kit. 8-119 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 518: Removing And Reinstalling The Air Dryer, Solenoid Valve Assembly, And Replacing Filters

    Reattach the barbed fitting to the outlet filter by screwing it in place. Rotate the solenoid valve assembly back to its original position (parallel with the air dryer assembly). Apply Teflon tape to the threaded end of the new air dryer filter/muffler. 8-120 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 519: Compressor Assembly

    Lift the compressor assembly out of the compressor compartment. Warning To prevent personal injury and equipment damage, have someone assist you when lifting the heavy compressor assembly. 8-121 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 520: Replacing The Coalescing Filter Element

    Remove the air dryer assembly (Section 8.17.16.1) and set aside· Remove the plenum assembly (Section 8.17.11.1) and set aside. Remove hose to outlet side of the coalescing filter. (Figure 8-101). Coalescing filter element Water trap Figure 8-102. Water trap assembly 8-122 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 521: Replacing The Compressor Panels

    If replacing a side panel, use the spare piece of foam in the foam kit to configure the panel to match the panel removed. Install the Tinnerman clips to the rear of the panel. 8-123 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 522: Reinstalling The Plenum Assembly

    Slightly tighten the four (4) Phillips screws that secure the plenum assembly to the right and left panels. Do not fully tighten the four Phillips screws in order to leave enough space to allow for clearance when reinstalling the rear panel/accumulator assembly. 8-124 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 523: Replacing The Main Inlet Filter And Reinstalling The Top

    There are 2 BPS units available — the 802 BPS and the 803 BPS. Both units consist of a +24 V battery pack and corresponding charging circuitry on its PCB assembly. The battery and 8-125 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 524: Removing The 802 Or 803 Bps

    If removing the 803 BPS, slide the 803 BPS from its mounting bracket (toward the front of the cart) and place on a firm work surface. GUI mount Rear caster (2 PL) Front caster (2 PL) Figure 8-103. 802 BPS and RTA cart 8-126 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 525: Installing 802 Bps

    Before installing the 803 Backup Power Source: • Turn off and unplug the ventilator from AC power. • Disconnect and remove the graphic user interface (GUI) and breath delivery unit (BDU) from the cart. 8-127 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 526: Figure 8-105. Removing Existing Brackets

    (Figure 8-105). Figure 8-105. Removing existing brackets Remove the three screws that attach the mounting bracket/backstop to the back of the new BPS and set aside (Figure 8-106). 8-128 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 527: Figure 8-106. Remove Screws From 803 Bps Bracket Backstop

    If the system does not have a compressor, skip tostep 8. If the system has a compressor, see step 6. Figure 8-108. Backstop position for system without a compressor 8-129 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 528: Figure 8-109. Backstop Position For System With A Compressor

    Using six of the screws that were included, attach the mounting bracket to the cart. Refer to the following figures to determine the proper holes to use for attachment: System without compressor – Figure 8-111 System with compressor – Figure 8-112 8-130 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 529: Figure 8-111. Position Of Mounting Bracket On Cart Without Compressor

    Use the three screws that you removed from the backstop earlier to reattach the BPS to the bracket. Figure 8-112. Position of mounting bracket on cart with compressor 8-131 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 530: Battery Pack

    2 PL Fascia panel 4 PL Front panel Floor assembly Battery pack Chassis (4 PL) Enclosure 8 PL Figure 8-113. Replacing battery pack and BPS PCB (for use on RTA cart) 8-132 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 531: Installing 802 Bps Battery Pack

    Connect battery pack harness connector to J2 on PCB. Install tie wrap to battery pack housing, making sure that it tightly wraps the battery pack harness and fuse. Install the 17 screws previously removed. 8-133 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 532: Figure 8-114. Replacing 803 Battery Pack And Pcb

    803 BPS assembly Tie wrap (2 pl) with mounting bracket Protective sleeve and backstop PCB assembly Cable clamp Panel end BPS harness assembly Figure 8-114. Replacing 803 Battery Pack and PCB 8-134 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 533: Bps Pcb

    Remove the nut holding the BPS harness assembly to the PCB. Remove BPS harness from cover assembly. Remove three 5/16” nylon locking nuts and washers. Remove 11/32”nut and star washer. Retain the end plate for later assembly. Remove PCB assembly. 8-135 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 534: Installing 803 Bps Pcb

    BPS, located in the BDU shelf. The Puritan Bennett 800 Series Ventilator Pole Cart cannot be used with a compressor. This cart can also be used with a one-hour battery or an optional four- hour battery, located in the cart base.
  • Page 535: Figure 8-116. Bps Assembly In Puritan Bennett 800 Series Ventilator Compressor Mount Cart

    The BPS assembly in the Puritan Bennett 800 Series Ventilator Compressor Mount Cart is shown in Figure 8-116. Refer to Figure 8-117 to connect BPS to Puritan Bennett 840 Ventilator System when mounted on Puritan Bennett 800 Series Ventilator Compressor Mount Cart.
  • Page 536: Removing And Replacing The Battery In The Puritan Bennett

    8.19.1 Removing and replacing the battery in the Puritan Bennett 800 Series Ventilator Compressor Mount Cart BPS The Puritan Bennett 800 Series Ventilator Compressor Mount Cart BPS comes with either a one-hour or four-hour battery depending upon the customer order. This section instructs you on replacing the battery.
  • Page 537: Removing The Four-Hour Battery

    Service and repair 8.19.1.1 Removing the four-hour battery Remove the front bezel from the BDU shelf on the Puritan Bennett 800 Series Ventilator Compressor Mount Cart by grasping the bezel and gently spreading the sides away from the BPS rails and pulling the bezel toward you (Figure 8-118).
  • Page 538: Replacing The Four-Hour Battery

    Caution Use care when installing the front cover to avoid damaging the battery charging status indicators. Verify or install the four-hour battery indicator label to front of BPS bezel (Figure 8-121). 8-140 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 539: Figure 8-121. Battery Charging Status And Indicator Labels

    8.19.1.3 Removing the one-hour battery Remove the front bezel from the BDU shelf on the Puritan Bennett 800 Series Ventilator Compressor Mount Cart by grasping the bezel and gently spreading the sides away from the BPS rails and pulling the bezel toward you (Figure 8-118).
  • Page 540: Figure 8-122. Location Of Four Top Cover Screws

    NOTE: If converting from a four-hour to a one-hour battery, order the one-hour battery shield to hold the battery in place. Disconnect battery harness from battery cable. Lift the battery out of the BPS. 8-142 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 541: Replacing The One-Hour Battery

    Reinstall the top cover of the BPS enclosure with the four screws previously removed. Verify or install the one-hour battery indicator label to front of BPS bezel (Figure 8-121). Verify or install the serial number label (Figure 8-124). 8-143 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 542: Figure 8-124. Serial Number Label Location Puritan Bennett 800 Series Ventilator Compressor Mount Cart

    Service and repair Figure 8-124. Serial number label location Puritan Bennett 800 Series Ventilator Compressor Mount Cart Warning Ensure the battery indicator label matches the battery installed. Unexpected power loss when operating on battery leading to significant patient injury could occur if a label is applied stating the ventilator has a four-hour battery when it actually has a one-hour battery installed.
  • Page 543: Removing And Replacing The Battery In The Puritan Bennett

    8.19.2 Removing and replacing the battery in the Puritan Bennett 800 Series Ventilator Pole Cart The battery in the Puritan Bennett 800 Series Ventilator Pole Cart is located in the cart base. Depending upon the customer order, there will either be a one-hour or four-hour battery installed.
  • Page 544: Four-Hour Battery Replacement

    Install two Phillips screws to secure in place (Figure 8-127). Figure 8-127. Securing Phillips screws on Puritan Bennett 800 Series Ventilator Pole Cart battery carrier (four-hour battery shown) Install battery cover.
  • Page 545: Figure 8-128. Battery Charging Status And Indicator Labels

    Service and repair Figure 8-128. Battery charging status and indicator labels (Puritan Bennett 800 Series Ventilator Pole Cart) 8.19.2.2.1 One-hour battery removal Remove Phillips screw at bottom front of battery cover and remove cover by lifting up and pulling forward.
  • Page 546: Figure 8-129. Orientation Of One-Hour Battery

    Install two Phillips screws to secure in place (Figure 8-127). Verify or install one-hour battery indicator label to front of cart (Figure 8-128). Verify or install the serial number label (Figure 8-130). 8-148 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 547: Charging Circuit Pcb Removal And Replacement For Puritan Bennett

    8.19.3 Charging circuit PCB removal and replacement for Puritan Bennett 800 Series Ventilator Pole Cart On the Puritan Bennett 800 Series Ventilator Pole Cart the BPS is incorporated into the cart and uses a charging circuit located on the underside of the BDU shelf cover.
  • Page 548: Figure 8-131. Pcb Cover

    Harness connector J3 Battery connector J2 Harness connector J1 Figure 8-132. Battery and harness connectors Remove four Nylock nuts fastening PCB assembly to charging circuit and nylon washers. Remove PCB assembly. 8-150 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 549: Charging Circuit Pcb Replacement

    Compressor Mount Cart or Puritan Bennett 800 Series Ventilator Pole Cart The casters on the Puritan Bennett 800 Series Ventilator Compressor Mount Cart and Puritan Bennett 800 Series Ventilator Pole Cart are press fit into the base and are not designed to be removed from the carts.
  • Page 550: Installing Casters On Puritan Bennett 800 Series Ventilator Compressor Mount Cart Or Puritan Bennett 800 Series Ventilator Pole Cart

    Compressor Mount Cart or Puritan Bennett 800 Series Ventilator Pole Cart The casters on the Puritan Bennett 800 Series Ventilator Compressor Mount Cart or Puritan Bennett 800 Series Ventilator Pole Cart, cannot be installed without a special tool and are sold already installed into their respective bases.
  • Page 551: Removing/Installing Flex Arm Inserts On Rta Cart

    Series Ventilator Compressor Mount Cart and Puritan Bennett 800 Series Ventilator Pole Cart On Puritan Bennett 800 Series Ventilator Compressor Mount Cart and Puritan Bennett 800 Series Ventilator Pole Cart, remove the flex arm inserts as follows: Using snap ring pliers, remove circlip from cart handle (Figure 8-136).
  • Page 552: Figure 8-136. Flex Arm Insert On Puritan Bennett 800 Series Ventilator Compressor Mount Cart And Puritan

    Service and repair Use snap ring pliers to remove circlip Figure 8-136. Flex arm insert on Puritan Bennett 800 Series Ventilator Compressor Mount Cart and Puritan Bennett 800 Series Ventilator Pole Cart 8-154 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 553: Parts List

    H A P T ER Parts list 9.1 How to use this parts list This chapter begins by showing the entire Puritan Bennett™ 840 Ventilator System, including accessories. Subsequent figures show ventilator subassemblies and their component parts. This chapter uses the following conventions: •...
  • Page 554: Table 9-1. Abbreviations Used In Parts List

    Internal Lock NIST Non-interchangeable Screw Thread Next Higher Assembly NOVRAM Nonvolatile Random-access Memory Outside Diameter Panhead Printed Circuit Board PEEP Positive End Expiratory Pressure Part Number Ready-to-Assemble Sleeved Index System Socket Split-ring Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 555: Puritan Bennett 840 Ventilator System Patient System And Accessories

    Parts list 9.2 Puritan Bennett 840 Ventilator System patient system and accessories Patient system and accessories parts list Item no. Part no. Quantity Description (Figure 9-1) Puritan Bennett 840 Ventilator System 4-032006-00 • Flex arm assembly (See Section 9.3.1 for parts breakdown.) •...
  • Page 556 T-T53204-00 • Screw (secures mounting receptacle) (not shown) 4-078105-00 • Mounting, receptacle, clip-on style 4-018506-00 • Flex tube, 21 in. (53 cm) ("gold standard" test hose, for use with EST) (not shown) Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 557: Figure 9-1. Puritan Bennett 840 Ventilator System Patient System And Accessories

    Puritan Bennett 840 Ventilator System patient system and accessories Remove pages 9-5 and 9-6. Page 9-6 is printed on the reverse side of foldout. Figure 9-1. Puritan Bennett 840 Ventilator System patient system and accessories Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 558: Puritan Bennett 840 Ventilator System Neomode Patient System And Accessories

    Parts list 9.2.1 Puritan Bennett 840 Ventilator System NeoMode patient system and accessories NeoMode patient system and accessories parts list Item no. Part no. Quantity Description (Figure 9-2) 4-076440-00 • Upgrade Kit, NeoMode, English, Non-US 4-076441-00 • Upgrade Kit, NeoMode, English, US only 4-076442-00 •...
  • Page 559: Figure 9-2. Neomode Patient Circuit And Accessories

    Parts list Slide humidifier onto bracket Figure 9-2. NeoMode patient circuit and accessories (shown on Puritan Bennett 800 Series Ventilator Pole Cart) Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 560: Flex Arm Assembly, Oxygen And Air Hose Assemblies, Power Cords

    • Shaft with pivot, 0.34 m (13.55 in.) long • Shaft with pivot, 0.38 m (15.0 in.) long • Shaft, rotating 4-032025-00 • Bolt, carriage 4-032021-00 • Clamp, left 4-032022-00 • Clamp, right 4-032023-00 • Handle, button-head, small Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 561: Figure 9-3. Flex Arm Assembly

    Parts list 6 (3 PL) 5 (3 PL) 3 (3 PL) 2 (6 PL) 1 (3 PL) Figure 9-3. Flex arm assembly Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 562: Oxygen Hose Assemblies

    Egypt, India, Italy, Kuwait, Poland, Portugal, South Africa) (See Section 9.2 for NHA.) 4-074708-00 Hose assembly, oxygen, DISS female x DISS female, blue, 3-m (10-ft) (for Switzerland) (See Section 9.2 for NHA.) 9-10 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 563: Figure 9-4. Oxygen Hose Assemblies

    Parts list 1 (clear) 8 (green) 9 (white) 10 (blue) Figure 9-4. Oxygen hose assemblies 9-11 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 564: Air Hose Assemblies

    Hose assembly, air, DISS female x DISS female, yellow, 3-m (10-ft) (for Switzerland) (See Section 9.2 for NHA.) Common module for the following: DISS male, DISS female, Air Liquide, NIST, CIG (See Section 9.4.9.2 for details.) 9-12 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 565: Figure 9-5. Air Hose Assemblies

    Parts list 1 (clear) 8 (black) 9 (white/black) 10 (yellow) 11 (white) Figure 9-5. Air hose assemblies 9-13 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 566: Power Cords

    Power cord, United Kingdom (See Section 9.2 for NHA.) 4-071421-00 Power cord, Denmark (See Section 9.2 for NHA.) 4-078218-00 Mounting receptacle kit (retainer and mounting screws) 4-070034-00 Screw, mounting (secures power cord receptacle clamp for latching power cord) 9-14 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 567: Figure 9-6. Power Cords

    Parts list Figure 9-6. Power cords 9-15 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 568: Ventilator Major Assemblies

    Parts list 9.4 Ventilator major assemblies Figure 9-7. Ventilator major assemblies on RTA cart 9-16 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 569 • Charcoal paint (spray) 4-017168-00 • Light gray paint (liquid) 4-017169-00 • Blue paint (liquid) 4-017170-00 • Charcoal paint (liquid) 1. Contact your Puritan Bennett representative to obtain 806 compressor unit ordering information. 9-17 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 570: Major Ventilator Assemblies (Puritan Bennett 800 Series Ventilator Compressor Mount Cart)

    Parts list 9.4.1 Major ventilator assemblies (Puritan Bennett 800 Series Ventilator Compressor Mount Cart) Ventilator major assemblies on Puritan Bennett 800 Series Ventilator Compressor Mount Cart 9-18 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 571 Parts list Major assemblies parts list (ventilators mounted on Puritan Bennett 800 Series Ventilator Compressor Mount Cart) Item no. Part no. Quantity Description (Figure ) Puritan Bennett 840 Ventilator System (See Section 9.2 for NHA.) • Graphic user interface (GUI): 10.4 inch (See pages 9-30 to 9-41 for parts breakdown.)
  • Page 572: Major Ventilator Assemblies (Shown On Puritan Bennett 800 Series Ventilator Pole Cart)

    Parts list 9.4.2 Major ventilator assemblies (shown on Puritan Bennett 800 Series Ventilator Pole Cart) Figure 9-8. Major ventilator assemblies (shown on Puritan Bennett 800 Series Ventilator Pole Cart 9-20 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 573 Parts list Major ventilator assemblies parts list (ventilators mounted on Puritan Bennett 800 Series Ventilator Pole Cart) Item no. Part no. Quantity Description (Figure 9-8) Puritan Bennett 840 Ventilator System (See Section 9.2 for NHA • Graphic user interface (GUI): 10.4 inch (See pages 9-30 to 9-41 for parts breakdown.)
  • Page 574 • Label, exhaust information, BDU, English/French • Label, exhaust information, BDU, German • Label, exhaust information, BDU, Japanese • Label, exhaust information, BDU, Portuguese • Label, exhaust information, BDU, Spanish/Italian • Label, exhaust information, BDU, Spanish/Portuguese 9-22 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 575 • Label, gas inlet, BDU, English/French • Label, gas inlet, BDU, German • Label, gas inlet, BDU, Japanese • Label, gas inlet, BDU, Portuguese • Label, gas inlet, BDU, Spanish/Italian • Label, gas inlet, BDU, Spanish/Portuguese 9-23 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 576 • Label, cooling vent, BDU, German • Label, cooling vent, BDU, Japanese • Label, cooling vent, BDU, Portuguese • Label, cooling vent, BDU, Spanish/Italian • Label, cooling vent, BDU, Spanish/Portuguese • Label, BPS connection 9-24 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 577 • Label, compressor, 220-240 V, 50 Hz, English/French • Label, compressor, 220-240 V, 50 Hz, German • Label, compressor, 220-240 V, 50 Hz, Portuguese • Label, compressor, 220-240 V, 50 Hz, Spanish/Italian 9-25 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 578 Label, alarm graphics, BDU, US 4-073028-00 Label, alarm graphics, GUI, non-US 4-075429-00 Label, alarm graphics, GUI, US 4-072505-00 Label, display, BPS, non-US (not shown) 4-075430-00 Label, display, BPS, US (not shown) 9-26 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 579: Figure 9-9. Label Kits (Sheet 1 Of 2)

    Parts list 8-000 Figure 9-9. Label kits (Sheet 1 of 2) 9-27 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 580: Figure 9-9. Label Kits (Sheet 2 Of 2)

    Parts list Graphic user interface (GUI) Figure 9-9. Label kits (Sheet 2 of 2) 9-28 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 581 Parts list This page intentionally blank. 9-29 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 582: Inch Gui (Graphic User Interface) Assembly With Trending-Enabled (Te) Gui Cpu Pcb

    • Gasket, EMI, 0.157” H x 0.315” W x 3.6” L 10011943 • Gasket, EMI, 0.157” H x 0.315” W x 3.0” L 4-076686-00 • Gasket, EMC, Chrometrics, RS-232, 5" 10015223 • Pad, thermal 9-30 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 583 Parts list 10.4-inch GUI assembly (TE GUI CPU PCB) parts list 4-075865-00 • Shield, dielectric, GUI 10021121 • FRU, IC, mem, compact flash 10021120 • FRU, IC, real-time clock (not shown) 9-31 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 584 Parts list This page intentionally blank. 9-32 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 585: Figure 9-10. 10.4-Inch Gui Assembly With Te Gui Cpu Pcb

    10.4-inch GUI assembly with TE GUI CPU PCB Remove pages 9-33 and 9-34. Page 9-34 is printed on the reverse side of foldout Figure 9-10. 10.4-inch GUI assembly with TE GUI CPU PCB 9-33 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 586: Inch Gui (Graphic User Interface) Assembly

    4-071944-SP • Cable, flex circuit assembly, RS-232 4-076856-SP • Cable, flex circuit assembly, video 4-076688-00 • Washer, nylon 4-075865-00 • Shield, dielectric, GUI 10021120 • FRU, IC, real-time clock (not shown) 9-34 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 587: Figure 9-11. 10.4-Inch Gui Assembly

    Replace this page and next with 11 x 17 foldout page: Figure 9-11. 10.4-inch GUI assembly Remove pages 9-35 and 9-36. Page 9-36 is printed on the reverse side of foldout Figure 9-11. 10.4-inch GUI assembly 9-35 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 588: Inch Gui (Graphic User Interface) Handle

    • Screw, PAN, 6-32 x 0.75, sems 4-073007-00 • Handle, GUI bottom • Kit, Label, Product information (Refer to Section 9.4.3, for part number.) • Label, serial number (Contact your Puritan Bennett representative for ordering information.) 4-076663-00 • Window, LED status, GUI 9-36...
  • Page 589: Figure 9-12. 10.4-Inch Gui Handle Assembly

    Parts list 4 (2 PL) Figure 9-12. 10.4-inch GUI handle assembly 9-37 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 590: Inch Graphic User Interface (Gui) Rotor Assembly

    • Housing assembly, GUI rear (See Section 9.4.5 for parts breakdown.) 4-073019-00 • Rotor, GUI 4-075232-00 • Spring, rotation, GUI 4-073018-00 • Shoe, rotor brake, GUI 4-070034-00 • Screw, 6-32 x 0.38, sems SQCN 9-38 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 591: Figure 9-13. 10.4-Inch Gui Rotor Assembly

    Parts list Figure 9-13. 10.4-inch GUI rotor assembly 9-39 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 592: Inch Gui Keyboards And Speaker Assembly

    • Screw, 6-32 X 0.38, sems, SQCN 4-079013-SP • Speaker assembly, GUI, 16 OHM 4-076652-00 • Cable assembly, keyboard 10020648 • FRU, ground strap (used only on keyboards with ground strap) 9-40 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 593: Figure 9-14. 10.4-Inch Gui Keyboard And Speaker Assembly

    Parts list Fasten ground strap to GUI EMI shield Figure 9-14. 10.4-inch GUI keyboard and speaker assembly 9-41 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 594 Parts list This page intentionally blank. 9-42 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 595: Inch Graphic User Interface (Gui)

    • Label, product information, GUI, Spanish/Portuguese (Available only as part of ventilator label kit.) • Label, serial number (Contact your Puritan Bennett representative for ordering information.) 4-070036-00 • Screw, PAN, 6-32 x 0.75, sems (Attaches chain to rear housing) 4-075186-00 •...
  • Page 596 • Cable extender, short 10015223 • Pad, thermal (Trending-enabled only) 4-075865-00 • Shield, dielectric, GUI 10021121 • FRU, IC, mem, compact flash (Trending-enabled only) 10021120 • FRU, IC, real-time clock (not shown) 9-44 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 597: Figure 9-15. 9.4-Inch Gui Assembly

    Replace this page and next with 11 x 17 foldout page: Figure 9-15. 9.4-inch GUI assembly Remove pages 9-45 and 9-46. Page 9-46 is printed on the reverse side of foldout Figure 9-15. 9.4-inch GUI assembly 9-45 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 598: Inch Gui Front Housing Assembly

    • Keyboard assembly, Japanese (Includes PCB and knob) 4-075833-SP • Keyboard assembly, Portuguese (Includes PCB and knob) 4-075831-SP • Keyboard assembly, Spanish (Includes PCB and knob) 4-071313-00 • Cable assembly, keyboard 9-46 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 599: Figure 9-16. 9.4-Inch Gui Front Housing Assembly

    Parts list Figure 9-16. 9.4-inch GUI front housing assembly 9-47 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 600: Inch Gui Rear Housing Assembly

    • Rotor, GUI 4-075232-00 • Spring, rotation 4-073018-00 • Shoe, rotor-brake, GUI 4-073032-00 • Foam kit (not shown) 4-073042-00 • Nut, push (Attach to foam kit to rear housing) (not shown) 9-48 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 601: Figure 9-17. 9.4-Inch Gui Rear Housing Assembly

    Parts list 3 (2 PL) Tilt/swivel assembly 6 (2 PL) Figure 9-17. 9.4-inch GUI rear housing assembly 9-49 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 602: Breath Delivery Unit (Bdu)

    BDU chassis) • Inspiratory module (See Section 9.4.9.2 for parts breakdown.) • Label, product information, BDU (Refer to Section 9.4.3 for labeling part numbers.) • Label, serial number (Contact your Puritan Bennett representative for ordering information.) 4-071594-00 • Gasket, BPS cover 4-071593-00 •...
  • Page 603 • Screw, PAN, 4-40 x 0.312, sems (Attach card cage I/O panel to PCBs and BDU chassis/cover) 4-073041-00 • Standoff, #4, M/F (Attach card cage I/O panel to AI PCB PTS 2000 connector) 9-51 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 604 4-078219-00 • Power cord mounting receptacle, retainer only (not shown) 4-078218-00 • Power cord mounting receptacle kit, retainer and mounting screws (not shown) 10021120 • FRU, IC, real-time clock (not shown) 9-52 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 605: Figure 9-18. Breath Delivery Unit (Bdu)

    Replace this page and next with 11 x 17 foldout page: Figure 9-18. Breath delivery unit (BDU) Remove pages 9-53 and 9-54. Page 9-54 is printed on the reverse side of foldout Figure 9-18. Breath delivery unit (BDU) 9-53 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 606: Exhalation Module

    4-072212-SP • Sensor with o-rings, flow, exhalation (Q3) • • Sensor, flow, exhalation (Q3) • • o-ring 4-070034-00 • Screw, PAN, 6-32 x 0.38, sems (Attach exhalation valve to exhalation chassis) 9-54 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 607 • Screw, PAN, 4-40 x 0.312, sems 10050832 • tape, 1” aluminum foil, conductive, 12” L 10050841 • copper finger stock, adhesive, .75” L 10048240 • copper finger stock, adhesive, 3” L 9-55 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 608 Parts list This page intentionally blank. 9-56 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 609: Figure 9-19. Exhalation Module

    NOTE: Lubricate o-rings with Krytox grease (P/N 4-732130-00). (2 PL) (2 PL) (2 PL) (2 PL) 10 (3 PL) (4 PL) (3 PL) exhalation I/O cable assembly Figure 9-19. Exhalation module 9-57 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 610: Inspiratory Module

    • Tape, adhesive transfer (Available in multiples of 1-ft lengths. Minimum order is 10 ft.) (20.3 cm) • Bracket, regulator 4-070047-00 • Screw, PAN, 6-32 x 5/16, sems (Attach SOL1 to regulator bracket) 9-58 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 611 PCB to pneumatic chassis) 4-079073-00 • Ferrite (use when upgrading older ventilators with NeoMode option) 4-078159-00 • Epoxy, conductive (Use only with EMI upgrades to ventilator) 10050840 • Copper finger stock, 2.5” L 9-59 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 612 Parts list This page intentionally blank. 9-60 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 613: Figure 9-20. Inspiratory Module

    NOTE: Lubricate o-rings with Krytox grease (P/N 4-732130-00). (3 PL) (3 PL) (3 PL) (2 PL) 5 (2 PL) Oxygen (3 PL) 16 (2 PL) (2 PL) Figure 9-20. Inspiratory module 9-61 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 614 4-071860-00 • Screw, FH, 8-32 x 0.5 (Attach SV to PSOL/SV manifold) 4-005817-00 • Screw, PAN, 6-32 x 1-1/8 (Attach outlet check valve manifold to PSOL/SV manifold) 4-071878-00 • Washer, silicone 9-62 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 615: Figure 9-21. Proportional Solenoid (Psol) Valve Assembly

    Parts list (4 PL) 8 (2 PL) NOTE: Lubricate o-rings with Krytox grease (P/N 4-732130-00). (2 PL) Figure 9-21. Proportional solenoid (PSOL) valve assembly 9-63 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 616 4-071931-00 • Filter, bronze, 1/2 x 1/16 (oxygen impact) (F1) 4-070036-00 • Screw, PAN, 6-32 x 0.75, sems (Attach inlet manifold to floor) 4-071861-00 • Spring, compression, 0.480 diameter x 0.375 9-64 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 617 • • Inlets, Air Liquide 4-076793-00 • • Inlets, NIST 4-076792-00 • • Inlets, CIG 4-076794-00 • Tag, serial number (Contact your Puritan Bennett representative for ordering information.) 4-071920-00 • o-ring 4-071918-00 • Tube, transfer, oxygen • Head, air filter 4-071922-00 •...
  • Page 618 4-071921-00 • o-ring 4-075567-00 • Washer, BDU interconnect 4-071932-00 • Spring, compression, 0.845 x 0.750 • Seat, check valve, compressor (CV4) • Poppet, check valve, compressor (CV4) • o-ring • o-ring 9-66 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 619: Figure 9-22. Inspiratory Module Floor Assembly

    (2 PL) (2 PL) (2 PL) (2 PL) (6 PL) (2 PL) (2 PL) (2 PL) (2 PL) NOTE: Lubricate o-rings with Krytox grease (P/N 4-732130-00). Figure 9-22. Inspiratory module floor assembly 9-67 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 620: Bdu Cover

    • Clamp, LED (Attaches LED indicator assembly to front panel) 4-070035-00 • Screw, PAN, 8-32 x 1/2, sems (Secures LED indicator assembly to front panel) 4-076897-00 • Cover, O2 access port • Nut 9-68 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 621: Figure 9-23. Bdu Cover Assembly

    Replace this page and next with 11 x 17 foldout page: Figure 9-23. BDU cover assembly Remove pages 9-69 and 9-70. Page 9-70 is printed on the reverse side of foldout Figure 9-23. BDU cover assembly 9-69 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 622: Bdu Chassis Assembly

    • Standoff, #4, M/F (Attach motherboard PCB to enclosure) 4-007566-00 • Washer, SR, #4 (Attach motherboard PCB to enclosure) 4-070900-SP • PCB, motherboard 4-070034-00 • Screw, PAN, 6-32 x 0.38, sems (Attach motherboard PCB to enclosure) 9-70 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 623: Figure 9-24. Bdu Chassis Assembly

    Replace this page and next with 11 x 17 foldout page: Figure 9-24. BDU chassis assembly Remove pages 9-71 and 9-72. Page 9-72 is printed on the reverse side of foldout Figure 9-24. BDU chassis assembly 9-71 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 624: Bdu Chassis Kit

    2 4.875 in. (12.4 cm) pieces 4-070056-00 • Grommet 4-071583-00 • Gasket, chassis, enclosure 4-071535-00 • Foot 4-070041-00 • Screw, PAN, 8-32 x 1/2, sems (Attach feet to chassis) 9-72 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 625: Figure 9-25. Bdu Chassis Kit

    Parts list (3 PL) (4 PL) (4 PL) (10 PL) (2 PL) 4.875 in. (12.4 cm) (2 PL) Figure 9-25. BDU chassis kit 9-73 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 626: Compressor Unit

    • Assembly, solenoid valve, 806 • Assembly, compressor motor (Available only as part of 10,000 hour preventative maintenance) 4-076263-00 • Clamp, air dryer 4-076723-00 • Shipping container, 806 Compressor (not shown) 9-74 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 627: Figure 9-26. 806 Compressor Unit

    Replace this page and next with 11 x 17 foldout page: Figure 9-26. 806 compressor unit Remove pages 9-75 and 9-76. Page 9-76 is printed on the reverse side of foldout. Figure 9-26. 806 compressor unit 9-75 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 628: Compressor Base Assembly

    • Clamp, capacitor, motor, 1.379" (220V) 4-075215-00 • Upper stopper 4-076318-00 • Pop rivet (not shown) 4-075247-00 • Sleeve, shock mount 4-076290-00 • Vibration mount 4-076282-00 • Pressure relief valve, 36 PSIG (not shown) 9-76 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 629: Figure 9-27. 806 Compressor Base Assembly

    Parts list Figure 9-27. 806 compressor base assembly 9-77 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 630: Compressor Unit Enclosure Assembly

    • Foam kit, sound, front panel (foam not shown) 4-076206-00 • Support, foam filter 4-074374-00 • Filter, foam 4-076218-00 • Bracket, accumulator 4-076477-00 • Accumulator 4-005483-00 • Nut 4-076242-00 • Foam kit, sound, back panel 9-78 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 631: Figure 9-28. Compressor Unit Enclosure Assembly

    Replace this page and next with 11 x 17 foldout page: Figure 9-28. Compressor unit enclosure assembly Remove pages 9-79 and 9-80. Page 9-80 is printed on the reverse side of foldout. Figure 9-28. Compressor unit enclosure assembly 9-79 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 632: Compressor Unit Plenum Assembly

    • Screw, 6-32 x 0.38 4-076297-00 • Cable, data, DC 4-076287-00 • Power cord, AC, 806 4-019135-00 • Cooling fan, 115V (Included in 4-020303-00 • Cooling fan, 100V PM kit) 4-020305-00 • Cooling fan, 220-240V 9-80 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 633: Figure 9-29. Compressor Plenum Assembly

    Replace this page and next with 11 x 17 foldout page: Figure 9-29. Compressor plenum assembly Remove pages 9-81 and 9-82. Page 9-82 is printed on the reverse side of foldout. Figure 9-29. Compressor plenum assembly 9-81 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 634: Backup Power Source (802 Bps)

    Label, product information, BPS, Japanese (not shown) 4-075412-00 Label, product information, BPS, Portuguese (not shown) 4-075085-00 Label, product information, BPS, Spanish/Italian (not shown) 4-075098-00 Label, product information, BPS, Spanish/Portuguese (not shown) 9-82 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 635: Figure 9-30. 802 Backup Power Source (Bps)

    Parts list 2 (2 PL) 4 (4 PL) (6 PL)  to J2 (4 PL) (2 PL) Figure 9-30. 802 Backup power source (BPS) 9-83 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 636: Backup Power Source

    Screws, Mounting Cart (not shown) 10029082 Protective sleeve C-492044-00 Cable tie, small 4-078166-00 6-32 Nylock nut (not shown) 10025456 8-32 nut with star washer (not shown) 10026139 Insulating washer (not shown) 9-84 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 637: Figure 9-31. 803 Bps Assembly

    Parts list Figure 9-31. 803 BPS assembly 9-85 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 638: Rta (Ready-To-Assemble) Cart Assembly

    • Screw, FH, 8-32 x 3/8 (Attach BPS slide brackets to cart) (not shown) 4-074652-00 • Drawer 4-070520-SP • Backup power supply (BPS) 4-075315-00 • Wall air water trap kit (not shown) 4-070770-00 • E-cylinder holder 9-86 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 639: Figure 9-32. Rta Cart Assembly

    Parts list 4 (3 PL) 11 (2 PL) 12 (2 PL) 6 (3 PL) 10 (4 PL) 8 (2 PL) 9 (2 PL) Figure 9-32. RTA Cart assembly 9-87 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 640: Puritan Bennett 800 Series Ventilator Compressor Mount Cart Assembly

    Parts list 9.4.14 Puritan Bennett 800 Series Ventilator Compressor Mount Cart assembly Item No. Part No. Quantity Description Figure 9-33 10046822 Puritan Bennett 800 Series Ventilator Compressor Mount Cart with one-hour battery 10046823 Puritan Bennett 800 Series Ventilator Compressor Mount Cart with...
  • Page 641 Screw, 1/4-20 x 1/2 PPHMS not shown 10043481 Screw, 1/4-20 x 1/2 SFHCS (humidifier mount screws) not shown 10053615 Screw, Slotted flat head (humidifier mount screws) not shown 10040930 Hook, tubing and cable, gray 9-89 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 642: Figure 9-33. Puritan Bennett 800 Series Ventilator Compressor Mount Cart Assembly

    Parts list Figure 9-33. Puritan Bennett 800 Series Ventilator Compressor Mount Cart assembly 9-90 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 643: Puritan Bennett 800 Series Ventilator Pole Cart Assembly

    10041445 Kit, grip handle,Puritan Bennett 800 Series Ventilator Pole Cart 10045975 Flex arm insert, with lockring 10046826 Cart with 1-hour battery, Puritan Bennett 800 Series Ventilator Pole Cart 10046827 Cart with 4-hour battery, Puritan Bennett 800 Series Ventilator Pole Cart...
  • Page 644 Screw, 1/4-20 x 7/8 SFHCS (humidifier mount screws) not shown 10053615 Screw, Slotted flat head (humidifier mount screws) not shown 10043479 Screw, 10-32 x 5/8 BSHCS, Nylock not shown 10035462 Hook, tubing and cable, blue 9-92 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 645: Figure 9-34. Puritan Bennett 800 Series Ventilator Pole Cart

    Parts list 6, 7 Figure 9-34. Puritan Bennett 800 Series Ventilator Pole Cart 9-93 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 646 Parts list 9-94 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 647 Adhesives, use of 8-3 in EST 3-27 AI PCB. See PCB, analog interface in SST 3-18 Air check valve (CV2). See Check valve, air Altitude requirements 1-4 Air dryer, description 2-44 Puritan Bennett 800 Series Ventilator System Service Manual Index-1...
  • Page 648 Breath type (displayed), function and range 2-114 illustrated parts breakdown 9-53 Breathing circuit, ventilator. See Patient circuit panel identification 1-48–1-55 BRIGHT key, description 1-40 part replacement Brightness (display) key, description 1-40 AC panel 8-64–8-65 Index-2 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 649: Figure 2-4. Compressor Unit

    CV5. See Check valve, exhalation replacement 8-83 Circuit breaker auxiliary (CB2), description 2-62 Data key main (CB1), description 2-62 caution 8-56 power supply, trip point 1-5 contents 2-82 Circuit, patient. See Patient circuit Index-3 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 650 Expiratory filter (F9). See Filter, expiratory 2-14 Environmental requirements 1-4 Expiratory pressure transducer autozero solenoid Error messages. See Diagnostic codes or Alarm (SOL2). See Solenoid, expiratory pressure messages transducer autozero EST service function 4-7 Index-4 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 651 (F3) real-time clock 8-20 description 2-19 rotor housing 8-26 replacement 8-69 touchframe PCB 8-22 oxygen pneumatic noise (F7), description 2-19 rear housing Flex arm assembly, illustrated parts breakdown 9-8– removing 8-12 Index-5 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 652 Heat exchanger (HE), description 2-44 (SOL1) 8-86 Heater assembly PSOL cartridge 8-70–8-71 exhalation (EXH HTR) PSOL manifold 8-78–8-80 description 2-39 right-side plate 8-75–8-77 replacement 8-98, 8-99 safety valve (SV) 8-81–8-82 High circuit pressure limit 2-112 Index-6 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 653 Oxygen inlet filter (F3). See Filter, oxygen inlet maintenance or Service and repair Oxygen pneumatic noise filter (F7). See Filter, Manifold, outlet, description 2-28 oxygen pneumatic noise MANUAL INSP key, description 1-42 Index-7 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 654: Figure 2-82. 806 Compressor

    2-6 replacement 8-22 illustrated parts breakdown 9-3–9-5 GUI, 9.4-inch pneumatic diagram 2-37 backlight inverter See also 840 Ventilator System Operator’s and description 8-36 Technical Reference Manual replacement 8-36 theory of operation 2-38 PC. See Transducer, pressure, compressor replacement 8-40...
  • Page 655 Power monitors, on GUI CPU PCB, description 2-85 REG1. See Regulator assembly, oxygen Power on self test (POST) 3-3–3-9 REG2. See Regulator assembly, air structure 3-10–3-12 Regulator troubleshooting. See Section 6 Index-9 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 656 See Exhalation module, part Screen replacement EST/SST Diagnostic Log 4-6 GUI (graphic user interface), 10.4-inch. See GUI SERIVCE MODE 4-2 (graphic user interface), 10.4-inch, part System Information Log 4-6 replacement Index-10 Puritan Bennett 800 Series Ventilator System Service Manual...
  • Page 657 8-86 leak testing, pneumatic connections 8-3 Sound generator interface, description 2-87 part ordering 8-4 Speaker assembly. See Alarm assembly patient system and accessories. See 840 Ventilator Specifications 1-4–1-29 System Operator’s and Technical Reference air/oxygen regulator bleed 1-5 Manual alarm volume 1-6...
  • Page 658 VGA controller. See PCB, VGA LCD controller VIM breath 2-111 Voltage line, how to set low alarm threshold 4-9 sentry circuit (on BD CPU PCB), description 2-81 supply, monitoring, description 2-118 Index-12 Puritan Bennett 800 Series Ventilator System Service Manual...