... the device is stored and operated according to LogTag Recorders ’ recommendations. Disclaimer The UTRIX-16(M) monitors temperature exposure and not the quality of the goods it accompanies. Its purpose is to signal if product quality evaluation/testing is required. Typographical Conventions Text refers to buttons on the UTRIX-16(M).
Follow the instruction to install and start the software. permanently plugged in USB extension cable to protect your If you only plan on evaluating data, plug the UTRIX-16(M) into a computer’s USB socket. computer‘s USB socket. Depending on the settings made during A computer running Windows XP or later, Mac OSX or Linux.
Analyzer will scan all USB ports and display the configuration options for connected UTRIX-16(M)‘s. Although you can configure many devices at the same time, it is practical to limit the number of devices to about 10, using a powered USB hub. It is not possible to configure different models at the same time.
LogTag Analyzer. Any additional upper alarms must have higher threshold values If you configure the UTRIX-16(M) to start after a delay period, the than any previously entered alarm. recorder will not immediately record temperature readings after Any additional lower alarms must have lower threshold values...
ALERT Microsoft Windows indicator blinks every 4 seconds. Depending on the alarm Up to four drivers will now be installed, depending on the UTRIX-16 configuration the red indicator may remain on for the rest of the trip, (M)'s configuration. or the green...
Unplugging You can unplug a UTRIX-16(M) at any time; there is no need to stop the device via the shortcut menu in the notification area on the taskbar. If you plug a unit into a USB port during a start delay, the delay period will re-start when the unit is unplugged.
2 ALERT for the recorder to stop. seconds The UTRIX-16(M) is connected to a USB port and waiting for USB Two quick flashes of the green indicator Ready for USB communication. every second The UTRIX-16(M) is generating files. No USB activity will be...
LogTag_[serial_number]_[trip number]_[file creation date]_file creation time]_[OK or ALM]. USB Paused Readings A gap is shown in the graph where the UTRIX-16(M) was plugged into Recorder Information and Configuration a USB port at the time it would otherwise have taken a reading.
Alarm Status Box Recorder Information and Configuration This shows at a glance if the UTRIX-16(M) recorded alarm conditions On recorders with multiple alarms these sections are combined and ✖ during the trip (showing a red...
Marked readings In the data list readings are marked after the temperature value..with a + symbol where the UTRIX-16(M) has been downloaded ® with LogTag Analyzer.
If a default configuration does not suit, customers can compile their own profile by specifying the parameters in the tables below when ordering. ® Please note that minimum order quantities apply for UTRIX-16(M) recorders with profiles not stocked by LogTag Recorders .
Technical Specifications Model Numbers Environmental UTRIX-16 - Multi-Trip PDF Temperature Recorder with dual alarms IEC 60529: IP64 with USB cap fitted UTRIX-16M - Multi Trip PDF Temperature Recorder with 6 alarms Fits into IP67 Protective Enclosure 200-000020 Temperature Sensor Measurement Range Case Material -25°C to +70°C (-13°F to +158°F)
Need help?
Do you have a question about the UTRIX-16 and is the answer not in the manual?
Questions and answers