Fresenius Medical Care 5008 Brief Operating Manual

Fresenius Medical Care 5008 Brief Operating Manual

Hide thumbs Also See for 5008:

Advertisement

The 5008 Therapy System
Brief Operating Guide

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 5008 and is the answer not in the manual?

Questions and answers

Summary of Contents for Fresenius Medical Care 5008

  • Page 1 The 5008 Therapy System Brief Operating Guide...
  • Page 2 Starting system and set-up ‘Alpha’/blood pump clip Post-dilution connector SafeLine/‘alpha clip’ Pre-dilution connector ONLINE HDF pump Blood/air optical detector ONLINE Substituate port Blood Temperature Monitor (BTM) Online Rinse port Caution The colour coding of the dialyser couplings may vary from that shown above –...
  • Page 3 Switch on the system Select ‘treatment’ (or ‘disinfection’). Connect concentrate and bibag , ensure correct concentrate is ® selected in dialysate menu. T1 test starts automatically. Open the EBM doors and follow on screen instructions for lining – arterial line first, ensu e the sound is generated when inserting the red ‘alpha clip’...
  • Page 4 Caution Transducer Protector (Hydrophobic filter 1. In order to avoid a possible ingress/loss of air and wetting of the transducer protector in case of pressure fluctuations, it is important to ensure that the transducer protector is correctly and tightly connected. 2.
  • Page 5 Venous bubble catcher and level detector. 1. It is important to ensure that the level detector is clean and dry and that the venous bubble catcher is correctly positioned so that it makes good contact with the level detector. 2. If there is a leakage of fluid f om the Safeline/ arterial patient line connection during preparation it is essential to clean and dry both the venous bubble catcher and the...
  • Page 6 Connecting a Patient ‘Alpha’/blood pump clip Post-dilution connector SafeLine/‘alpha clip’ Pre-dilution connector ONLINE HDF pump Blood/air optical detector ONLINE Substituate port Blood Temperature Monitor (BTM) Online Rinse port Caution The colour coding of the dialyser couplings may vary from that shown above –...
  • Page 7 When the patient access is prepared and the full prime volume has been reached, stop the prime by pressing the blood pump button on the screen. Message ‘choices’ – ‘continue’ (rinse) or ‘exit’ (rinse) are displayed, press ‘exit’ (rinse). Follow on screen instructions – connect arterial and venous lines to patient access.
  • Page 8 Reinfusion/Disconnection ‘Alpha’/blood pump clip Post-dilution connector SafeLine/‘alpha clip’ Pre-dilution connector ONLINE HDF pump Blood/air optical detector ONLINE Substituate port Blood Temperature Monitor (BTM) Online Rinse port Caution The colour coding of the dialyser couplings may vary from that shown above – please check your local policy/colour coding.
  • Page 9 Press ‘reinfusion’ and follow on screen instructions. Disconnect the SafeLine from the post or pre-dilution connector and attach to the recirculation adapter. Disconnect arterial line from patient. Once connected press ‘OK’ (only press ‘NaCl’ if you have run out of acid/bicarb and need to use saline to wash back). The blood pump will stop automatically when clear fluid is recognised by the optical detector .
  • Page 10 5008 Alarm Handling Quick Reference Guide Caution This guide does not replace any information in the manufacturer’s operating manual, which should be read before operating the machine. Additional information and help regarding alarms is available by pressing the question mark (help key) in the alarm warning...
  • Page 11 1. ONLINE Connection error (i) Blood pump section has not fed in properly, or (ii) the clamp has not been released on the arterial line, or (iii) part of the circuit is disconnected. Check all connections and that the clamps have been released (use the info’/? button to provide hints on possible causes).
  • Page 12 4. Extracorporeal Blood Module (EBM) Sensor alarm There is a leakage of 20ml of blood or priming fluid in the extracorporeal blood module. Open the doors and clean the drip tray/sensor. Check all connections to identify the leak. plus leakage alarm 5.
  • Page 13 6. Hydraulic leakage (treatment not possible) 70ml of fluid has inte nally leaked in the hydraulics. Discontinue treatment and contact Technicians for assistance. 7. Micro bubble alarm Follow on screen instructions, use operating manual if necessary. 8. Air handling procedure Disconnect patient and re-circulate blood lines following the on-screen instructions.
  • Page 14 10. Venous pressure measurement incorrect If during prime – remove venous transducer protector and reconnect. Press ‘repeat check’. Most common cause is a flooded venous transducer p otector. Press blood system screen and using down arrow reduce level of fluid in the venous bubble catche . Press ‘re-check’ on the screen. If necessary change venous transducer protector.
  • Page 15 12. Optical detector does not sense blood The optical sensor is below the venous bubble catcher – either the bloodline is incorrectly inserted in the line guide, or treatment has been discontinued without using the re-infusion programme. Follow on screen instructions. 13.
  • Page 16 Notes...
  • Page 17 Notes...
  • Page 18 Notes...
  • Page 19 Notes...
  • Page 20 Notes...
  • Page 21 Notes...
  • Page 22 Fresenius Medical Care UK Ltd. · Nunn Brook Road · Huthwaite · Sutton in Ashfield · Notts · NG17 2HU · England · Telephone:+44 (0) 1623 445100 · Fax: +44 (0) 1623 550807 Head Office: Else-K öner-Straße 1 · 61352 Bad Homburg v. d. H. · Germany www.fmc-ag.com...