Set-Up Commands - GE Druck DPI 740 User Manual

Precision pressure indicator
Table of Contents

Advertisement

50

Set-up commands

Command:
Format:
Description:
Example:
Query:
Command:
Format:
Description:
Example:
Query:
SU1=0
K0200 Issue No. 1
SA- Set instrument address
SA = <nn>
Set instrument address to <nn>. Changes the
network address of the instrument in the
range 0 to 98. Address 99 is used for global
commands.
SA=10 set instrument address to 10
SA?
query instrument address
Reply:
SA=10
SU - Set-up "regular" units
SUn = <units index>
Selects the pressure units that appear as
default selection F2 in measurement mode.
Both required parameters must be specified
i.e. the push-button sequence n (1, 2 or 3) and
the <units index>.
SU1=0 sets first units (1) to mbar
SU2= 18 sets second units (2) to inHg
The corresponding query command returns
the index to the pressure units selected for
the specified option.
SU1?
Reply: SUn=<units index>
DPI 740 User Manual