CONNECTION AND SETTINGS: Setting Parameters
When the parameter is set to Necessary, the data is shown by a numeric button on a yellow
background except when JOB data is imported from built-in memory. Unless the EP value
on the yellow background is modified or confirmed, the intelligent blocker prohibits the oper-
ator from blocking or saving the JOB data.
To set EP values for each individual JOB data, set the parameter to Necessary. This pre-
vents an error in position of a blocked lens cup caused by not modifying or confirming the EP
value.
20) Diopter step: 0.01, 0.12, 0.25
Factory setting: 0.25
This parameter sets the increments of SPH and CYL among 0.01, 0.12, and 0.25 D.
The setting of 0.01 is only for adjustment. Do not select it for normal use.
21) Polish/S.B setting: LEX, SE/ME
Factory setting: LEX
This parameter sets whether the lens edge is polished and/or safety-beveled with the lens
edger. Select LEX or SE/ME according to the lens edger to be used.
The selectable options for polishing and safety beveling differ depending on this parameter
setting:
22) Manual Mode Scale: 0.0, 1.0, 2.0, 3.0
Factory setting: 2.0
Select the spacing between horizontal lines of the alignment scale from among 0, 1.0 mm,
2.0 mm, and 3.0 mm that is displayed on the layout screen with the MANUAL designated.
When 0 mm is selected, the scale is not displayed.
23) Alignment mark V size: 0.0, 1.0, 2.0
Factory setting: 2.0
Select the height of the alignment scale (horizontally long ellipse), displayed on the layout
screen when MANUAL is selected, from 0 mm, 1.0 mm and 2.0 mm. When 0 mm is
selected, the horizontally long ellipse becomes a straight line.
24) Auto JOB# Increment: EXEC, NONE
Factory setting: EXEC
This parameter is effective only on the ICE-1000NT (without the tracking unit equipped).
25) Lens Holding Time: 0.0 to 5.0 sec.
This parameter is for setting the time period for the holder arm to remain at rest at the lowest
position in the range of 0.0 to 5.0 seconds.
94
Setting
LEX
When safety beveling mode is designated:
Non-Pol, Polish, Pol (EDG), Pol (S.B)
SE/ME
When safety beveling mode is not designated:
Polishing
Non-Pol, Polish
Non-Pol, Polish
Safety beveling
Non-S.B, S.B
Non-S.B, Small, Middle,
Large, Special