Functions And Features; Mains Supply - Thermo Scientific SL40 F Instruction Manual

Table of Contents

Advertisement

Functions and Features

The following table gives an overview of the important functional and performance characteristics of
the Thermo Scientific SL40 / SL40 FR.
Table 1-3. Functions and Features
Component / Function
Structure / Housing
Chamber
Drive
Keys and display
Controls
Internal memory
Functions
Acceleration / braking profiles
Rotor recognition
Imbalance recognition
Lid lock

Mains Supply

The following table contains an overview of the electrical connection data for the Thermo Scientific
SL40 / SL40 FR. This data is to be taken into consideration when selecting the mains connection
socket.
Table 1-4. Electrical connection data of the Thermo Scientific SL40 / SL40 FR
Cat.
7500 4543
7500 4542
Thermo Scientific
Mains
Frequency
voltage
refrigerated
230 V
50 / 60 Hz
ventilated
230 V
50 / 60 Hz
Description / Features
Galvanized steel chassis with armoured plating
Stainless steel / varnished steel
Induction drive without carbon brushes
Easy-to-clean keypad and display surface
Microprocessor-controlled
The most recent data is saved
RCF-selection, temperature control, pretemp with cooling equipped
devices
9 acceleration and 10 braking curves
Automatic
Electronic, contingent on rotor and speed
Automatic lid closing and locking starting from an initial hold position
Rated
Power
current
consumption
10,5 A
1950 W
11 A
1700 W
1
Introduction and Description
Functions and Features
Equipment
Building
fuse
fuse
15 AT
16 AT
15 AT
16 AT
SL40 F / SL40 FR
1-5

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sl40 fr

Table of Contents