Identification and revision of the EasyOne Air spirometer This revision V03 of the EasyOne Air Operator’s Manual applies to EasyOne Air with a serial number greater than 1,000,000 and with a firmware version greater than V1.13.0. If you are in doubt whether this revision of the Operator’s Manual applies to your particular EasyOne Air sensor, please contact the ndd Servicing Department.
1.4. Intended use of the EasyOne Air spirometer The EasyOne Air spirometer is intended for prescription use only to conduct diagnostic spirometry testing of adults and pediatric patients over 4 years old. The EasyOne Air spirometer is used by general practitioners, specialists, and health care professionals, in hospitals and clinics, in pharmacies, and in clinical settings in occupational medicine.
3.3. Setting up the physical components ......20 3.4. Setting up the EasyOne Air spirometer......23 3.5.
Page 7
4.1. Introduction to the EasyOne Air spirometer ..... 25 4.1.1. Overview of EasyOne Air ......25 4.1.2.
Page 8
10. Configuration 10.1. Configuration of EasyOne Air ....... . 72 10.1.1.
Page 9
13. Specifications and bibliography 13.1. List of specifications for EasyOne Air ......89 13.2. List of specifications for the lithium-ion battery of EasyOne Air ..91 13.3.
IEC 60950-1 standard for the electrical safety of IT equipment. Patient health Inadequate qualification of personnel who operate EasyOne Air can put the hazard patient’s health at risk. The patient may suffer an asthma attack or a bronchospasm while performing spirometry.
Page 11
For preventive measures, make sure that the device is not used near flamma- ble gases (for example, anesthetic agents). If not in daily use, remove the battery from EasyOne Air and store the battery in the original packaging. False diagnosis Environment conditions outside the range described in the specifications can cause measurement errors.
Page 12
Install anti-virus software on your PC. Malfunction Damaged components can lead to false results. Make sure that no liquid penetrates into EasyOne Air, for example through the battery compartment or the USB port. Leakage of acids Do not connect the battery improperly.
EasyOne Air. Keep a distance of at least 30 cm (12 in) to any part of EasyOne Air. Increased emissions Accessories, transducers, and cables other than those specified by the manu-...
2.4. About requirements for connections to external devices EasyOne Connect is installed on a PC or laptop, to which EasyOne Air is connected via the cradle or wirelessly. The PC or laptop is not part of the equipment supplied by the manufacturer and is considered a requisite accessory that is necessary to use EasyOne Air with the PC software EasyOne Connect.
EasyOne Connect can use any printer that is installed on your PC. To print using EasyOne Connect, connect the printer to your PC and not to the cradle. Description of print settings of EasyOne Air 7 8 Setting up the physical components 2 0 Application Note Printer Compatibility List Application Notes for further information ...
Page 22
Plug the ndd power supply into a power outlet, which is easily accessible for connection and disconnection. b) Connect the cable of the ndd power supply to the right USB port of the cradle (marked with the plug icon).
CAUTION FAILED UPDATE An update may fail unnoticed. After performing an update, restart EasyOne Air and EasyOne Connect and per- form a calibration check. Cybersecurity 3 7 Checking for correct operation of EasyOne Air 8 6 1 Connect the USB flash drive from the box contents to your PC.
Page 26
Introduction When EasyOne Air is turned on and not in use, the screensaver and automatic power-off function are activated to save battery power. You can manually switch off the device by pressing the power button. The EasyOne FlowTube respiratory tube for single-patient use ...
In the patients menu, you have access to patient data and also to the reports, which are sorted by patient here. In the tools menu, you can configure EasyOne Air and perform calibration checks and have access to settings and database functionalities.
Page 29
In the print settings menu, you can configure the printer that is connected to the cradle. In the factory reset menu, you can reset the configuration of EasyOne Air to its delivery status. However, the patient data will not be deleted. Configuration of EasyOne Air 7 2 Operator’s Manual V03 •...
Page 30
The charging level is at about 50%. Indicates that EasyOne Air is connected to a power supply, but cannot charge. Check whether EasyOne Air is connected to the ndd power supply or to a different USB power source. Only use the ndd power supply.
EasyOne Connect is especially helpful to work with patient data and to evaluate results because you can use a full-size PC monitor and keyboard. If you want to perform spirometry with EasyOne Air stand-alone, for example off-site, EasyOne Connect enables you to enter all relevant patient data beforehand and synchronize that data to EasyOne Air.
Introduction 4.6. About sources for predicted normal values The available predicted normal values for EasyOne Air are updated as required through software updates. You can find the currently used sources for predicted normal values in the document below. Application Note Reference Predicted Normal Values Application Notes for further information ...
Cybersecurity 5.6. Dealing with a lost or stolen EasyOne Air or PC In case EasyOne Air or the PC that EasyOne Connect is installed on has been lost or stolen, do the following: Prerequisite The manufacturer strongly recommends that you activate user handling ...
Page 40
Optionally, fill out the other fields. f) Choose OK. User handling is activated. Now, users must log in to use EasyOne Air. 3 To prevent unauthorized access when EasyOne Air is unattended, specify a time period for automatic power-off: a) Choose >...
Performing spirometry Animation program for children If you use EasyOne Connect in combination with EasyOne Air, an animation program for children is available. The animation is available for the FVC test, the FVL test and the SVC test. You can choose between three animations: balloon, cake, and monkey.
Choosing a test and selecting an existing patient If you test a patient that you have tested before or if you have transferred patient data onto EasyOne Air from EasyOne Connect on your PC, start by selecting an existing patient.
Page 47
Visually check the EasyOne FlowTube for defects before use and between trials. Before you use EasyOne Air with the patient, make sure that there are no obvious or wet residues on EasyOne Air from prior cleaning. To prepare the test 1 Partly unwrap the EasyOne FlowTube.
Page 51
EasyOne FlowTube holder or the inside of EasyOne Air while cleaning. To clean EasyOne Air, do not fully dip EasyOne Air into any fluid. To clean EasyOne Air and to disinfect your hands after each patient, do the...
Bluetooth built-in. List of box contents 1 7 If EasyOne Air is directly connected to a PC over USB, performing tests is not possible. Leakage currents could harm the patient. Connect EasyOne Air to your PC wirelessly, only.
On the printed report, parameters that are below the lower limit of normal (LLN) are marked with an asterisk (*). Unacceptable trials are struck through. In addition to the marks, EasyOne Air and EasyOne Connect offer an automatic interpretation aid.
To clean EasyOne Air, do not immerse EasyOne Air in any fluid. To clean EasyOne Air from environmental dirt, you can use a damp cloth. After each patient, clean EasyOne Air. You can use a soft cloth with a cleaning solution according to the following list.
5 If you want to print the report, do the following: a) Dock EasyOne Air to the cradle that is connected to a printer. b) Choose Print. A progress bar is displayed, which indicates the progress of the print- out.
EasyOne Connect to your EMR system. EASYONE AIR MAY FREEZE If EasyOne Air is connected to your PC via USB and if there are also electromag- netic fields from other electrical devices nearby, EasyOne Air may freeze.
EasyOne Connect displays a message whether the sending has been successful. You can remove EasyOne Air from the cradle, or you can repeat steps 2 to 7 to create and send additional patient records. About connectivity and data exchange 3 6 9.2.3.
Configurations in the settings menu of EasyOne Air > 10.1.1.1. Description of system settings of EasyOne Air > > On this screen, you can choose basic settings for EasyOne Air. Setting Purpose Speaker volume To set the speaker volume Key click...
Therefore, the predicted normal values are subject to change. These changes are delivered to you via software updates. Updated information is published in an Application Note on the ndd website. Application Note Reference Predicted Normal Values Application Notes for further information 3 List of bibliographic references ...
Configuration 10.1.1.4. Description of print settings of EasyOne Air > > You can determine basic settings for printouts when using EasyOne Air and the cradle. There are more comprehensive configuration options available for printouts when using EasyOne Connect. Setting Purpose...
To archive a copy of the complete database (including patient data and results) and of the configuration settings, perform backups regularly. You can also use the backup to set up another EasyOne Air with the same database and configuration settings.
Security policies, like password strength and password expiration, must be handled by the operating institution, for example the hospital or practice. EasyOne Air provides the option to enforce strong password rules. See Description of security settings of EasyOne Air 7 If user accounts are password-protected, users are only prompted for their password once for login.
Calibration of EasyOne Air is not possible. You can merely check EasyOne Air for correct calibration. Do not attempt to repair EasyOne Air yourself. A third-party service must not at- tempt to repair EasyOne Air either. Do not open or remove the casing.
2 Perform a spirometry test on yourself or on another known subject. 3 Check the results of steps 1 and 2 for plausibility. a) If the results of steps 1 and 2 are plausible, EasyOne Air is operating correctly. b) If the results of steps 1 and 2 are not plausible, troubleshoot.
Contact your EasyOne distribution partner or the ndd Servicing Department. To reactivate EasyOne Air, you must setup EasyOne Air like you did for the first-time setup. List of specifications for EasyOne Air 8 9 About charging the battery of EasyOne Air ...
Page 90
Specifications and bibliography Instrument Type BF applied part (the surface of the EasyOne Air enclosure) classification Powered internally with a lithium-ion battery Short usage time, less than 10 minutes Not intended to be used in the presence of oxygen-rich environments...
Do not use the battery in combination with any other batteries. Keep out of the reach of children. Only charge the battery when inside of EasyOne Air and in conjunction with the cradle and the power supply provided by the manufacturer.
Page 92
Specifications and bibliography The following standard conditions are identical to the conditions of EasyOne Air. Storage, transport, and operation under these conditions is safe. Standard storage and transport Temperature -20°C to 50°C (-4°F to 122°F) conditions Relative humidity 5% to 90%...
From the test menu in EasyOne Connect, choose the OSHA test. On EasyOne Air, choose the OSHA test from the more menu. The patient performs the same breathing maneuver as for the FVC test. Reports meet the requirements of the following institutions and standards: •...
IEC 60601-1-2:2014 4 edition (according clause 7 and 8.9, tables 4 to 9). EasyOne Air is a medical device that requires special safety precautions and must be installed and placed in operation in accordance with the information given in tables below.
Electromagnetic emission Guidance and manufacturer’s declaration – electromagnetic emissions EasyOne Air is intended for use in the electromagnetic environment specified below. The user of EasyOne Air should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment – guidance...
IEC 60601-1-2. Guidance and manufacturer’s declaration – electromagnetic Immunity EasyOne Air is intended for use in the electromagnetic environment specified below. The user of EasyOne Air should assure that it is used in such an environment. Immunity test standard...
Page 104
If the measured field strength in the location in which the EasyOne Air is used exceeds the applicable RF compliance level above, the EasyOne Air should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the EasyOne Air.