5. INOGEN ONE G2 ERROR CODE TABLE

If more than one alarm is triggered simultaneously, the error code number will be the sum of the
error codes given below.
Error Code
Explanation
Number
001
Low voltage on power supply
002
Software error
004 B/C
Pressure Error
CONFIDENTIAL and
96-02735-00-01 revC
PROPRIETARY
AMSR
Rechargeable Battery:
Approximately 2 to 3.5 hours for 12 cell battery
Approximately 4 to 7 hours for 24 cell battery
Approximately 2.25 to 4.25 hours for 12 cell battery
Approximately 4.25 to 8.00 hours for 24 cell battery
Temperature: 40 – 104
Humidity: 0% to 95%, non-condensing
oxygenconcentratorstore.com
Voltage:
12.0 to 16.8 VDC,
o
o
F (4 – 40
C)
Continuous Duty
Class II
Type BF
Not intended for cardiac application
Concentrator: IP 20
Carry bag:
IP 22
Not suitable for such applications
Possible Causes & Troubleshooting
Instructions
Replace the power supply or remove
the battery to see if the concentrator
was running on battery power
Unplug the external power and remove
the battery to clear the error and then
restart the concentrator
The concentrator was shut down due
to a pressure error. Ensure that the
INOGEN ONE G2 OXYGEN CONCENTRATOR
TECHNICAL MANUAL
Page 18 of 20