Cleaning, Maintenance and Storage
Cleaning Instructions
1.
Disconnect the warming unit from the power source before cleaning.
2.
Cleaning should be performed in accordance with hospital practices
for cleaning OR equipment. After every use; wipe the warming unit,
the outside of the warming unit hose, and any other surfaces that may
have been touched. Use a damp, soft cloth and a hospital approved
mild detergent, germicidal disposable wipes, disinfecting towelettes,
or antimicrobial spray.
3.
Let air dry or dry with a separate clean soft cloth.
Notice:
Do not use cleaning solutions with greater than 80% alcohol or solvents,
including acetone and thinner, to clean the warming unit or hose. Solvents
may damage the labels and other plastic parts.
Do not immerse the Bair Hugger warming unit, warming unit parts, or
accessories in any liquid or subject them to any sterilization process.
FAULT CODE
FC-001
FC-002
FC-003
FC-004
FC-005
FC-006
FC-007
FC-008
FC-009
FC-010
FC-050
FC-053, 054
FC-100
FC-101
GB / 34-8718-8817-7
POSSIBLE CAUSE
Faulty hose end sensor at
patient end.
Faulty elbow sensor.
Warming Unit Hose
disconnected from warming
unit (FC-007 only).
Faulty heater element or
faulty control board.
Defective user interface.
EEPROM error.
Defective control board.
SERVICE FREQUENCY
As needed.
STORAGE
Store all components in a cool, dry place when not in use.
Troubleshooting
The warming unit software routinely checks unusual operating conditions. If
such a condition occurs, the system may trigger a fault condition and alarm.
The warming unit can be reset by disconnecting the warming unit from the
main power source and allowing 30-60 seconds for the system to reset. If
the fault code condition reoccurs, disconnect again and wait 5 minutes for
the system to reset. If the fault condition reoccurs upon connection to the
main power source, the warming unit may have a hard-fault or permanent
error. Contact 3M technical support. If the fault does not reoccur upon
connection to the main power source, the warming unit may have a
cautionary fault.
Replace warming unit hose. Refer to service manual or IFU included with
replacement part for instructions.
Replace elbow sensor. Refer to IFU included with replacement part
for instructions.
Reattach warming unit hose.
Calibrate and return to service.
Verify fan runs in ambient.
Check resistance of J4 connector on control board. The resistance may be up to
400 Ohms.
White to black
White to blue
White to yellow
If "open", replace heater. Otherwise, replace control board.
Calibrate and return to service.
Replace user interface.
Calibrate and return to service.
Reset microprocessor.
(Press and hold Alternative Mode button and Standby buttons at same time then
connect to power source. Release buttons when Standby light remains on).
Calibrate and return to service.
Replace control board.
Calibrate and return to service.
SOLUTION
11