Installation Instructions
(Translation of the original installation instructions)

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Dewert Okin SG 300 and is the answer not in the manual?

Questions and answers

Summary of Contents for Phoenix Mecano Dewert Okin SG 300

  • Page 1 SG(AG) 300 control unit Installation Instructions (Translation of the original installation instructions)
  • Page 3: Foreword

    SG(AG) 300 control unit Foreword Foreword Disclaimer and Exclusion of Liability DewertOkin is not responsible for damage resulting from: Failure to observe these instructions, Changes made to this product which have not been approved by DewertOkin, or The use of replacement parts which have not been approved or manufactured by DewertOkin. Contact address DewertOkin GmbH Weststraße 1...
  • Page 4: Table Of Contents

    Table of contents SG(AG) 300 control unit Table of contents Foreword Disclaimer and Exclusion of Liability Contact address Creation of a complete operating instruction manual for the entire end product Usage in medical products Table of contents General Information About these installation instructions Conventions used Safety Notices Proper and Intended Usage...
  • Page 5: General Information

    SG(AG) 300 control unit General Information General Information About these installation instructions In order to install the SG(AG) 300 control unit successfully and safely in the end product, these instal- lation instructions must be observed. These instructions are not an operating manual for the end product.
  • Page 6: Safety Notices

    Safety Notices SG(AG) 300 control unit Safety Notices Proper and intended usage The SG(AG) 300 control unit is designed for use as a control unit and power supply for the appropriate DewertOkin drive systems. it can be used for care purposes (CARE model). it can be used in a hospital (HOSP model).
  • Page 7: Selection And Qualification Of Personnel

    SG(AG) 300 control unit Safety Notices Using the drive systems in medical applications This DewertOkin product is in compliance with the safety requirements found in IEC 60601-1/ EN60601-1. We strongly recommend that the end product (including all its components) which you are manufacturing for a medical application should also be in compliance with the safety require- ments found in IEC 60601-1/EN 60601-1.
  • Page 8: Product Labelling

    Safety Notices SG(AG) 300 control unit Product labelling 2.4.1 Type label A type label on each SG(AG) 300 control unit specifies the exact name and serial number of the drive. It also states the technical specifications valid for that particular control unit. The following illustration shows where the specifications are located on the ratings plate of the SG(AG) 300 control unit.
  • Page 9: Possible Combinations

    SG(AG) 300 control unit Possible combinations Possible combinations The SG(AG) 300 control unit can be combined as follows. The following basic combinations are possible: SG 300 control unit max. 5 single drives, max. 2 handsets, one locking element or operating element: e.g. supervisor, locking device or short circuit plug SGAG 300 control unit max.
  • Page 10: Device Description

    Device description SG(AG) 300 control unit Device description The SG(AG) 300 control unit is a control unit and power supply for one or more DewertOkin drives. A pluggable power cord is used to connect the SG(AG) 300 control unit to the mains power supply.
  • Page 11 SG(AG) 300 control unit Device description 4.1.2 Variant 2: SGAG 300 control unit Figure 4 Components for the SGAG 300 control unit A Power supply via pluggable power cord B SGAG 300 control unit C Connection sockets for drives, locking- D Screw fixing point to the end product /operating element and handset with m e- chanism to protect again pulling out...
  • Page 12 Device description SG(AG) 300 control unit 4.1.3 Optional grounding cable The optional grounding cable can be used to connect the end product with the grounding conductor on the power supply cord. Attach the grounding cable to your application in com- pliance with all applicable standards and using state-of-the-art methods.
  • Page 13 SG(AG) 300 control unit Device description 4.1.4 Mains power supply WARNING Please follow these operating instructions carefully. Yo u could be injured by fire or electrical shock if you do not follow these assembly instru ctions. The appropriate power cord is included, depending on the regional version (USA, continental Europe, the UK, Australia or Japan).
  • Page 14: Technical Data

    Technical data SG(AG) 300 control unit Technical data Mains power supply 100 - 240V AC, 50/60Hz Current consumption at nominal opera- max. 4.0 A – 1.6 A (depending on input voltage, tions depending on the used power supply) Mode of operations Intermittent duty 2 min./18 min.
  • Page 15 SG(AG) 300 control unit Technical data Figure 7 Dimensions of the SG(AG) 300 control unit, top view (in mm) Figure 8 Dimensions of the SG(AG) 300 control unit, front view (in mm) SG 300 SGAG 300 Figure 9 Dimensions of the SG(AG) 300 control unit, side view (in mm) 75082(-)
  • Page 16: Installation

    Installation SG(AG) 300 control unit Installation Safety notices to observe during installation Basic safety rules must be followed in order to ensure that the end product can be continually op- erated in a safe manner. These rules must be observed while using the end product and while in- stalling the SG(AG) 300 control unit.
  • Page 17: Installation Procedure

    SG(AG) 300 control unit Installation Installation procedure Before installing the SG(AG) 300 control unit, make sure that you are observing all of the safety no- tices found in the "Safety notices to observe during installation" section. 6.2.1 Installation and dismounting for the control unit There are four mounting holes in the SG(AG) 300 control unit which can be used to attach it to the end product with the appropriate screws (for example, use screws meeting the requirements of DIN EN ISO 7049 with Ø...
  • Page 18 Installation SG(AG) 300 control unit 6.2.2 Electrical connection CAUTION Electrical components should be connected or disconnected only when the power su pply cord is unplugged. NOTICE There is a delay after the supply voltage is applied before the device actually turns on.
  • Page 19 SG(AG) 300 control unit Installation Opening the shield cover Figure 11 Opening the shield cover on the SG(AG) 300 control unit A Shield cover B Locking clip 1 Pull out the mains plug from the outlet. CAUTION You should only connect and disconnect the cables when they are completely discon- nected from any live cu rrent! 2 Use a suitable tool to loose the both locking clips, refer to figure 11.
  • Page 20 Installation SG(AG) 300 control unit 6.2.3 Connecting the pluggable power cord to the SG(AG) 300 control unit WARNING Please follow these operating instructions carefully. You could be injured by fire or electrical shock if you do not follow these assembly instru ctions. The appropriate power cord is included, depending on the regional version (USA, continental Europe, the UK, Australia or Japan).
  • Page 21 SG(AG) 300 control unit Installation The pluggable power cord should be attached to the mains connecting socket (C) of the control unit. CAUTION You should only connect and disconnect the cables when they are completely disco n- nected from any live current! Plug the power plug from the power cord (B) into the socket (C).
  • Page 22 Installation SG(AG) 300 control unit 6.2.4 Rechargeable AG 300 battery to the SGAG 300 control unit CAUTION Work on electrical components should be conducted only when the mains power connection is unplugged. Connection and installation of the rechargeable AG 300 battery Figure 1 Connecting the rechargeable AG 300 battery to the SGAG 300 control unit A Rechargeable AG 300 battery...
  • Page 23 SG(AG) 300 control unit Installation 1 Insert the rechargeable AG300 battery (A) in the rear receptacle of the controller (B). Then press the rechargeable AG300 battery into the upper recesses until the snap element (C) engages. 2 Connect the cable (E) to the battery socket. 3 Wrap the strain relief (D) around the battery plug and tighten it in position with the screw (F, 3.9 x 9.5 DIN EN ISO 7049) to the SGAG 300 control unit.
  • Page 24: Operating Notes

    Operating Notes SG(AG) 300 control unit Operating Notes The factual information contained within may be used when you are creating the end-product manual. The installation instructions do not contain all information required for the safe opera- tion of the end product. They only describe the assembly and operation of the SG(AG) 300 control unit as a partially assembled piece of machinery.
  • Page 25 SG(AG) 300 control unit Operating Notes Avoiding electrical risks WARNING Be sure that all live (current - carrying) parts of the drive system and power supply cannot be touched. In part icular, be sure that unused power and control unit co n- nections are covered ad equately.
  • Page 26: Notice For Operating With Optional Configuration

    Operating Notes SG(AG) 300 control unit Avoiding cable damage Be sure that your operating instructions inform the user about the possible cable risks. CAUTION The cables (particularly the connecting cable) should not be run over. In order to prevent injuries or damage to the drive and SG(AG) 300 control unit, no me chanical strain should be placed on the cables.
  • Page 27: Troubleshooting

    SG(AG) 300 control unit Troubleshooting Troubleshooting This chapter contains remedial actions should any malfunctions occur. If you experience an error that is not listed in this table, please contact your supplier. CAUTION Only qualif ied specialists who have received electrician training should carry out troubleshooting and r epairs.
  • Page 28 Troubleshooting SG(AG) 300 control unit Signal tone Meaning Measure / Action The battery is discharged and has Recharge the battery Warning tone when the switched off. completely. handset is used. A continuous alarm tone The control unit is broken. Please contact your sup- can be heard.
  • Page 29: Maintenance

    SG(AG) 300 control unit Maintenance Maintenance u You should only use spare parts which have been manufactured or approved by DewertOkin. Only original or approved spare parts guarantee sufficient levels of safety. Maintenance Type of check Explanation Time interval Check the function and A qualified electrician should Periodic inspections can safety of the electrical sys-...
  • Page 30: Cleaning And Care

    Maintenance SG(AG) 300 control unit Cleaning and care The SG(AG) 300 control unit was designed so that it would be easy to clean. The smooth sur- faces can be conveniently cleaned. ACHTUNG Never clean the SG(AG) 300 control unit in an automated washing system or with a high -pressure cleaner.
  • Page 31: 10. Disposal

    SG(AG) 300 control unit Disposal 10. Disposal The SG(AG) 300 control unit consists of electronic components, cables and metal and plastic parts. You should observe all corresponding national and regional environmental regulations when disposing of the SG(AG) 300 control unit. The SG(AG) 300 control unit should not be disposed of with normal household waste! The disposal of the rechargeable battery is regulated in the EU by the Battery Directive...
  • Page 32: Eu Declaration Of Conformity

    EU Declaration of Conformity EG-Konformitätserklärung In compliance with Appendix IV of the EMC-Directive Nach Anhang IV der EMV-Richtlinie 2014/30/EU 2014/30/EU In compliance with Appendix IV of the LVD-Directive Nach Anhang IV der EU-Niederspannungsrichtlinie 2014/35/EU 2014/35/EU Nach Anhang VI der RoHS-Richtlinie 2011/65/EU In compliance with Appendix VI of the EU RoHS (inkl.
  • Page 33 Additional Information The following standards and norms were used in the versions with at least IPX4, in accordance with EN 60601-1, IEC 60601-1, third Edition, Medical electrical equipment EN60601-1, Section 4 General requirements EN60601-1, Section 6 Classification EN60601-1, Section 8 Protection against electrical shock hazard EN60601-1, Section 11.1 Protection against excessive temperatures...
  • Page 34 In accordance with EN 60601-2-52:2010, IEC 60601-2-52:2009, "Particular requirements for the safety and essential performance of medical beds", the following standards are used: EN 60601-2-52, Section 201.6.2 Protection against electrical shock: Protection class II EN 60601-2-52, Section 201.7.6.3 Control panel symbols (depending on model, customer requirements) EN 60601-2-52, Section 201.8.11.3.2 Power supply lead: only >= 2.5 m length Power...
  • Page 35 DewertOkin GmbH Weststraße 1 32278 Kirchlengern, Germany Tel: +49 (0)5223/979 -0 Fax: +49 (0 )5223/75182 http://www.dewertokin.de Info@dewertokin.de ID No.: 7 5082...