  • Page 1 ENGLISH Instruction Manual READ INSTRUCTIONS IN FULL BEFORE USE PP2MGE001...
  • Page 2 The Pico is also not suitable for co-sleeping or bed sharing. SNUZA is a registered trademark of Snuza International (Pty) Ltd Unit 8-9 Roeland Square Roeland Street...
  • Page 3: Table Of Contents

    CONTENTS 1. Important: Please read first 2. Introduction 3. Precautions and Contraindications 4. Parts Identification 5. Quick Start Guide (QSG) 6. Diaper Fitment and Placement 7. Operating your Pico 7.1 Fitting your Pico 7.2 Switching on/off 7.3 Pause 7.4 Charging 7.5 Indicator Lights 7.6 Movement Indicator Light 7.7 Bluetooth Low Energy (BLE)
  • Page 4: Important: Please Read First

    English 1. IMPORTANT - PLEASE READ FIRST! The Pico is designed to attach to your baby's diaper or child's pajama pants while they sleep and will continously monitor breathing motion, skin temperature, sleep position and sleep state. If connected, the information is streamed to your smartphone via BLE (Bluetooth low energy) technology.
  • Page 5 English In order for the Pico to record your baby's position accurately, go to the settings page on the app and set where the Pico has been attached to your baby diaper or child's pajama pants. Depending on your baby's sleep position, you can also place the Pico to the left or right-hand side of the naval.
  • Page 6: Introduction

    English 2. INTRODUCTION Intended Use: The Pico is a portable baby breathing motion monitor which safely and conveniently monitors your baby or child's breathing motions while sleeping. Intended User: This equipment is intended to be operated by the baby or child’s caregiver. The caregiver must have a good understanding of the instructions and safety measures within the user manual.
  • Page 7 English If the vibration succeeds in rousing the baby and breathing is detected after such an event the Pico will continue to monitor breathing movement. After three vibration/rouse events, the Rouse Warning will alert you to the fact that your baby’s breathing movements have stopped for 15 seconds on three occasions.
  • Page 8: Precautions And Contraindications

    In case your baby develops a skin sensitivity where the device makes contact with his/her skin, refer to section 7.11 in the manual. If the skin sensitivity persists, please contact us at support@snuza.com. • Keep out of reach of pets, pests and other children as damage to the device might cause malfunction.
  • Page 9: Parts Identification

    English 4. PARTS IDENTIFICATION FRONT BACK Page 6...
  • Page 10 English Page 7...
  • Page 11: Quick Start Guide (Qsg)

    A confirmation email will be sent to your selected email address. Once confirmed, you will be taken to the Snuza Portal. Navigate to the ‘Device Registration’ tab in the menu. Enter your device serial number and complete the required detail fields.
  • Page 12: Diaper Fitment And Placement

    Basic App Functions The Snuza Connect App is designed to bring you information on your baby’s breathing rate, body position, skin temperature, fall detection, battery levels and alert you if the Pico’s clip is open. Customise your warning settings to your preference in the App.
  • Page 13: Operating Your Pico

    English . 7 OPERATION 7.1 Fitting your Pico When attaching the Pico you should put your fingers in between the baby’s nappy and the baby’s skin. Make sure that the monitor is clipped securely onto the diaper and away from the baby’s skin when clipping on in order not to harm the baby.
  • Page 14: Switching On/Off

    English 7.2 Switching on / off The Pico can be turned on by briefly pressing the power button (6) and can be switched off by pressing and holding the power button (6) for 2 seconds. The corresponding startup and shutdown tones will be heard (refer to point 7.5) Alternatively, the Pico can also be switched on and off directly from the app (if it has been enabled in the app settings).
  • Page 15: Movement Indicator Light

    Flashes pink 6 times immediately after startup to indicate that there is an important notification for you on the server that needs your attention. Pink Flashing View this important message by logging into your account on to the Snuza Connect Portal (connect. snuza.com) Page 12...
  • Page 16: Battery Indications

    English 7.8 Battery Indications At startup: As explained above at “Startup and shutdown sequences” section When battery level is low (while the Pico is on): Battery Indicator Light will flash red every 5 seconds and will beep twice every 60 seconds When battery level is critical (while the Pico is on)"...
  • Page 17: Charger Indicator Lights

    English Skin Temperature: The app can be configured to alarm when skin temperature falls below or rises above set thresholds. This alert only presents on the smartphone and will sound only if connected to the Pico via Bluetooth. Body Position: This alarm can be set in the app to notify you when your baby moves into a specific body position such as turning onto his/her tummy which is not recommended as part of safe sleeping techniques.
  • Page 18: Clothing

    English 7.11 Clothing The Pico can be clipped onto the folded waistband of the diaper or the waistband of elasticated pants. It is not recommended to have clothing between the Pico and the skin. The Pico can be used under clothing, but not over. Too much clothing may attenuate or muffle the alarm.
  • Page 19: Rouse Warning

    English . 8 ROUSE WARNING Note: If the smartphone is connected, all warnings and alarms, as well as all real-time data are visible on the app. During monitoring, if no breathing motion is detected for 15 seconds, the Pico vibrates gently in an attempt to rouse your baby.
  • Page 20: Battery

    English . 9 BATTERY 9.1 Battery Indicator If the battery becomes depleted during monitoring, the Battery Indicator Light (1) will flash as follows: Battery Indicator Light (1) will flash red every 5 seconds and will beep twice every 60 seconds. When the battery becomes critical the Battery Indicator Light will flash red every 5 seconds and beep twice every 10 seconds.
  • Page 21: Maintenance

    English 10 MAINTENANCE 10.1Disposal: The device, and packaging material must be disposed of in accordance with your country’s laws governing the disposal and recycling of batteries: 10.2Cleaning and Storage: • Gently wipe with a damp cloth to remove any dirt. A mild soap may be used for stubborn dirt.
  • Page 22: Troubleshooting

    The Battery Indicator Light (1) will show red Please visit www.snuza.com for further information. 12 DISCLAIMER Customers use this monitor at their own risk. Snuza International (Pty) Ltd, its members, employees, agents, contractors, representatives, manufacturers, distributors and associative (SNUZA) shall not be held...
  • Page 23: Guarantee

    English 13 GUARANTEE Snuza International (Pty) Ltd guarantees this product to be free from defects in materials and workmanship for a period of 12 months from the original date of purchase, verified by a sales receipt. Our guarantee liability shall be limited to replacing the unit free of charge and does not include transport costs.
  • Page 24: Technical Description

    Device Serial Number (the first two numbers indicate the year of manufacture See the Disposal section for proper disposal of the monitor Snuza The Pico has an expected Typical Service Life: service life of 5 years from the date of manufacture. Page 21...
  • Page 25 English Permissible environmental operating conditions: · Temperature: 5 ºC TO 40 ºC. Environmental · Humidity: %15 TO %93, Non Conditions: Condensing, · Air pressure: 700HPA TO 1060HPA Permissible environmental conditions of transport · 25 °C without relative humidity and storage after having control;...
  • Page 26 Contact Us: For assistance with using or maintaining Snuza The Pico or to report unexpected operation or events, contact us via: Website: www.snuza.com/contact E-mail: support@snuza.com Telephone: +27 (0)21 461 8530 Physical address: Unit 8 - 9 Roeland Square, Roeland Street,...