Gas Analyser Not Communicating - CryptOn 700 Series Hints & Tips

Emissions tester
Hide thumbs Also See for 700 Series:
Table of Contents

Advertisement

Emissions Tester CCP700/800 Series

Gas Analyser not communicating

Check mains supply OK and the analyser module is switched on
Check communication cable
Check platform Config for conflicts
LHC Residue Test Failure
Check sample probe in free air, blow out sample probe & pipe assembly if required
Change gas filters if heavily contaminated and wash filter bowls with warm soapy water
Low Gas Flow
Ensure that the filters have been removed and cleaned or replaced, sample probe has been
checked for damage and blockages.
High Lambda Readings
The system is drawing in air, please ensure the following – All filter bowls are undamaged and
sealed (hand tight and a little bit, do not over tighten or you will damage the o ring). Sample
probe and pipe are undamaged and connections are tight.
Remove Probe from Exhaust message
Check the O2 sensor two pin connector is still connected and the millivolt reading is no less than 7
millivolts.
(at beginning of an MOT test)
Hints & Tips
Emissions Tester CCP700/800 Series
Hash Error/reset/ Audit Log Invalid
If the machine is coming up with an error relating to the wrong hash then to fix this issue you will
need to clear the old hash code to do this:-
Get the customer to go into to service mode > Serial Number Update > enter the Serial at the top
of the screen into the entry box.
On the Gas programme, select Service Mode on the right hand side
Ask for the displayed calibration request code
Give the customer the reply code
After this, customer should have lots more options down the right side
Select "Service Mode Diagnostics"
Then "Gas Analyser Serial Number Update"
Ensure that the serial number matches the one on the unit - customer MUST re-enter
serial no. Then, click on "Audit Log Diagnostics" (right hand side)
Will come up with a load of "gobble-dee-gook"
Just click Exit/Blue Arrow/X - back to the main screen
F0 00 00 00 (aka IR Bench)
This error could mean that the gas has moisture inside. The weather plays a part in this failure as
this can be caused by condensation on cold mornings. It can also be caused by water not being
emptied from the water trap each day.
On a 680EN1
To fix the issue:-
If this fails to resolve the issue then an exchange gas for warranty or a fixed price repair are the
only options.
8
Disconnect the 3 pipes from the small bowls (2 small at the top and 1 Large at the
bottom)
Run Leak test for 2 hours with a heater (nothing more than an internal building
radiator) next to the device.
Switch the device off using the red rocker switch at the back at the bottom
Switch the device back on
Hints & Tips
9

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 700 Series and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

800 seriesCcp700 seriesCcp800 series

Table of Contents

Save PDF