Lng liquid ultrasonic meter for cryogenic liquefied natural gas applications (98 pages)
Summary of Contents for Emerson Rosemount Analytical RDO
Page 1
Manual 51-RDO Model RDO April 2010 ® Optical Dissolved Oxygen Analyzer and Sensor RDO is a registered trademark of In-Situ Inc. All rights reserved.
MODEL RDO TABLE OF CONTENTS TABLE OF CONTENTS No. Title Page Section 1: Installation ........................4 Box Contents..........................4 Optional Mounting Kit........................4 Mounting Options ...........................4 1.3.1 Analyzer Dimensions ........................4 1.3.2 Pole Mounting ..........................5 1.3.3 Wall Mounting Using Mounting Tabs .....................6 1.3.4 Wall Mounting With User-supplied Screws and Hardware ............6 Installing Dome Connectors and Plugs ..................7 Section 2: Wiring..........................8 Electro-static Discharge (ESD) Recommendations ..............8...
Page 4
Emerson Process Management makes no warranty of any kind with regard to this material, including, but not limited to, its fitness for a particular application.
MODEL RDO INSTALLATION SECTION 1.0 INSTALLATION 1.1 BOX CONTENTS 1.3 MOUNTING OPTIONS Remove the analyzer from the shipping box. The ship- 1.3.1 Analyzer Dimensions ping box should contain: • Figure 1-1 calls out the analyzer dimensions with • Analyzer (1) the lid closed.
Page 6
MODEL RDO INSTALLATION 1.3.2 Pipe Mounting 1. Remove the four nuts and four screws from the mounting kit. 2. Open the enclosure and drop one nut into each drilled corner of the box (Figure 1-3). 3. Use a screwdriver to push the nut down and set it in place (Figure 1-3).
MODEL RDO INSTALLATION 1.3.3 Wall Mounting Using Mounting Tabs The optional mounting kit contains a set of wall mount- ing hardware that includes four screws, nuts, and tabs for mounting the analyzer to a wall or panel. Follow the instructions included with the mounting tab hardware to attach the tabs (Figure 1-6).
MODEL RDO INSTALLATION 1.4 INSTALLING DOME CONNECTORS 3. Place the lock nut concave side down on the AND PLUGS threads and using a screwdriver blade and mallet, tighten the lock nut until there is approximately one Install dome connectors and plugs as shown in Figure thread showing.
MODEL RDO WIRING SECTION 2.0 WIRING 2.3 ENSURING GOOD ELECTRICAL DANGER CONNECTIONS To ensure that all sensors and power sources function Only properly trained and qualified personnel properly, make sure that: should install the RDO analyzer described in this • Each individual wire is stripped and tinned to ¼...
Page 10
MODEL RDO WIRING The AC power board is located under the metal shield on the right side of the enclosure. The terminal strip on the board accommodates the AC power connections as well as two high voltage relay connections. See Figure 2-1. FIGURE 2-2.
Page 11
MODEL RDO WIRING 2.5 ANALOG AND DIGITAL OUTPUTS AND LOW VOLTAGE RELAY CONNECTIONS DANGER Requires 24-12 AWG cable. Requires stripped and tinned ® RuggedCable system Requires stripped and tinned ® RuggedCable system Requires 24-12 AWG cable CONNEXIONS POUR AUTOMATE PROGRAMMABLE INDUSTRIEL DANGER Requires 24-12 AWG cable Débrancher toute alimentation à...
MODEL RDO WIRING 2.6 RDO PROBE WIRING DANGER Make sure that power to the instrument is discon- nected before making any wiring connections. CONNEXIONS POUR LE RACORD DES SONDES FIGURE 2-4. Earth ground screws for DANGER analog outputs and sensors. Débrancher toute alimentation à...
MODEL RDO PLACING PROBE IN SERVICE SECTION 3.0 PLACING THE RDO PROBE IN SERVICE 3.1 UNPACKING 3. Align the arrow on the cap with the index mark on the probe and firmly press the cap onto the probe, Remove the probe and sensing cap from the box. The without twisting, until it seals.
MODEL RDO DISPLAY AND OPERATION SECTION 4.0 DISPLAY AND OPERATION 4.1 DISPLAY AND KEYPAD Table 4.1 shows the icons used in the display. Figure 4-1 shows the keypad and display window. Table 4-1 Display icons and their descriptions Error with parameter Calibration stabilizing Calibration nominally stable View...
MODEL RDO PROGRAMMING THE ANALYZER SECTION 5.0 PROGRAMMING THE ANALYZER 5.1 INITIAL CONFIGURATION b. Use the Up/Left Arrow key to enter a positive or negative value. Enter the following information upon first use or any- c. Press the Down/Right arrow key to move the time after you have restored the default settings.
MODEL RDO PROGRAMMING THE ANALYZER FIGURE 5-1. Menu Tree 1. Select Menu, then Settings, then Probes. Select • Enable or disable the measurement. If the meas- RDO Probe A or RDO Probe B, if it is present, then urement is enabled, the probe will make the meas- Select Options.
MODEL RDO PROGRAMMING THE ANALYZER as soon as you press any key to leave. 5.3.4 Deleting a Probe d. Scroll to Sentinel and press Select. Use the If you plan to permanently remove a probe, follow the arrow keys to change the setting. steps below.
MODEL RDO PROGRAMMING THE ANALYZER 3. Select Temperature 8. Scroll to 20mA Trim: Use the arrow keys to adjust the loop current to 20 mA. a. Press the Select key to toggle between enable and disable. If the box is checked, temperature 5.5.2 Alarm Relays has been selected.
MODEL RDO PROGRAMMING THE ANALYZER 4. Scroll to Mode and select it. Choose high alarm or 1. Select Menu, then Settings, then Power Save. low alarm. 2. Select External. Do not choose Battery. 5. Scroll to Parameter and select it. Scroll through the 3.
MODEL RDO PROGRAMMING THE ANALYZER You may set up the analyzer with passwords for an 5.14 DIAGNOSTICS administrator and a user. The user can access the The following diagnostic information is available for the Calibrate, Hold Options, and Diagnostics menus. The RDO probe: administrator can access all functions, most notably, •...
MODEL RDO CALIBRATION SECTION 6.0 CALIBRATION 6.1. CALIBRATION OPTIONS There are three ways to calibrate the RDO probe: • Saturation: Both a two-point (100% and 0% satura- tion) and one-point (100% saturation) calibration are available. 100% saturation refers to water com- pletely saturated with atmospheric oxygen.
MODEL RDO CALIBRATION 5. Allow at least 5 minutes for the temperature to stabi- 4. The 0% saturation step starts automatically. lize before starting the calibration. Keeping the sen- Remove the probe and fill the calibration chamber to sor in the shade will help reduce drift caused by the the upper fill line with approximately 60 mL of fresh sun’s heat.
MODEL RDO CALIBRATION 7. The calibration report screen appears. The slope will be between 0.8 and 1.2 and the offset will be between -0.2 an +0.2. Press OK. 8. Remove the probe from the calibration chamber. Rinse it and return it to the process liquid. 6.4 CONCENTRATION CALIBRATION 1.
MODEL RDO MAINTENANCE SECTION 7.0 MAINTENANCE 7.1 ANALYZER 2. Gently scrub the probe body with a soft-bristled brush or a nylon dish scrubber. Use a mild detergent 7.1.1 Cleaning the analyzer enclosure to remove oil or grease. Soak in vinegar followed by Clean the analyzer case and front panel by wiping with deionized water to remove mineral deposits or ex- a clean, soft cloth dampened with water ONLY.
Page 28
MODEL RDO MAINTENANCE 7.3 REPLACEMENT PARTS Part number Description R0094030 RDO analyzer, AC power, no data logging R0086460 RDO sensor with 32 ft (10 m) of integral cable R0082490 RDO sensor with twist lock connector R0087560 RDO analyzer pipe and wall mounting kit R00CBL10 Twist lock connector cable, 32 ft (10 m) R00CBL20...
Page 29
MODEL RDO SPECIFICATIONS SECTION 8.0 SPECIFICATIONS 8.1 RDO PROBE 8.2 RDO ANALYZER Wetted Materials: Delrin , ABS, Viton , titanium, polycar- Enclosure: Polycarbonate, rated NEMA 4X, IP67 bonate/poly(methyl methacrylate) blend Dimensions (W x H x D): 6.3 x 6.3 x 3.6 in (16 x 16 x Dimensions: length 8.0 in (203 mm) diameter 1.9 in 9.0 cm) (47 mm)
Page 32
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE right now. http://www.raihome.com Specifications subject to change without notice. Credit Cards for U.S. Purchases Only. Sira MC070110/00 Emerson Process Management 2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250 http://www.raihome.com...
Need help?
Do you have a question about the Rosemount Analytical RDO and is the answer not in the manual?
Questions and answers