Hanna representative near you at www.hannainst.com. All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA Table of Included ...............................3...
Page 3
Remove the instrument & accessories from the packaging and verify damage Included has not occurred during shipping. Remove protective lm from meter. Notify your nearest Hanna Customer Service Center if damage is observed. Each instrument is supplied with: edge ® (HI 2002) Bench cradle Wall cradle Electrode holder...
Description edge ® enables the user to make fast, accurate measurements of commonly measured laboratory parameters using the Hanna edge ® digital sensors for pH. Each digital sensors has a unique serial numbe. Once connected to the meter, the sensor(s) are ready to measure their parameter along with temperature. The user interface permits you to adapt edge ®...
Product Diagram Liquid Crystal Display (LCD) Micro USB device connection for Capacitive Touch Keypad power or PC interface 3 mm jack input for edge ® Standard USB host connection digital probes for data transfer to a USB Top mounted ON/OFF button thumb-drive Diagram...
Probe Diagram Probe pH Electrode • Probes process signal directly for noise free measurements • Auto sensor recognition • Store calibration speci c data from the last calibration • Are built with materials suitable for use in chemical analysis • Have integrated temperature measurement •...
Keypad Function CAL/MODIFY - Used to enter - Used to scroll through exit calibration mode. SETUP menu. Used change SETUP, used to initiate changes of a selection when modifying con guration setting. parameter in SETUP. GLP/CFM - Used to display GLP RCL (Recall) - Used to view calibration information.
Guide to Indicators Mode tags Third LCD line, message area Con rm tag 10. Labels USB connection status 11. Second LCD line, temperature pH electrode diagnostics measurement Probes symbol 12. Temperature units Battery symbol 13. Temp. Compensation status Arrow tags, displayed when they 14.
The main operating modes of edge ® are setup, calibration, measurement, Setup/ data logging, and data export. Follow this general outline of steps to get you Installation started. The following topics are expanded upon in the sections that follow in this manual.
Setting Up Connect the power adapter cable to the edge ® bottom socket of the wall cradle. Connect the other end to the power adapter and plug into line power. Connect the 3mm probe jack to the socket located at the bottom of edge ®.
General The following General Setup options are displayed regardless of the sensor Setup being used. These settings remain when switching to another probe type or when no probe is attached. Options are tabulated in the table below with choices and default values. Options are accessed by pressing SETUP. Loop through the options by using the pq arrows.
General Option Description Choices Default Basic mode Setup Used to save battery life by automatically turning o when 5, 10, 30, 60 no key press is Set Auto O 10 MIN Available detected for or O time set and meter is not in active logging or calibration mode.
Page 13
edge ® o ers a basic operation mode that streamlines measurement Basic Mode con guration for pH measurements and is useful for many routine applications. Basic pH SETUP reduces parameter selection to the basic set. The meter limits calibration to 5 standard pH bu ers; 6.86, 7.01, 4.01, 9.18 and 10.01 bu ers. All pH measurements will display, log and export with 0.01 pH resolution.
Logging In Setup mode, choose log parameter, press MODIFY then use the u arrow to select between Interval, Manual, or Stability. When Interval is displayed, use the Function p and q arrows to select the setting for the timed interval. When Stability is displayed, use the p and q to select the measurement stability setting.
Stability Logging Logging Select Stability and choose measurement stability Function criteria in the SETUP menu. Only Stability Medium is available in Basic mode. To initiate the Stability log, press the LOG key while the instrument is in measurement. The “PLEASE WAIT” screen will be displayed brie y followed by a screen showing the stability tag, “LOG”...
Page 16
Viewing Press CFM to enter inside lot information to view recorded data. Use the ARROW keys to toggle between di erent records. Logged Use u to display GLP data including calibration information, date, time, etc. Data Press CLR then CFM when deleting records or lots. Press RCL to exit the logging type.
Page 17
Select the Manual (Stability) lot and press CLR. Viewing The ”CLEAR” message will be displayed along with Logged ”MANUAL” or “STABILITY” and CFM tag blinking on Data the LCD. Press the CFM key to con rm the deleting of the selected lot (MANUAL or STABILITY) or all records. Press CLR to exit without deleting.
PC & Storage Logged data on edge ® can be transferred from the meter to a USB ash drive Interface by using the log recall function. The minimum requirement for the drive is USB 2.0. Select the pH record you wish to export and follow the simple steps below. Connect USB ash drive to the USB port, located on the top of the meter.
Page 19
PC & Storage Logged data on the edge ® can be transferred from the meter to a PC by Interface following these simple directions. Suitable operating systems include Windows (Xp minimum), OS X or Linux. Connect edge ® to the PC using the supplied micro USB cable. Turn on edge ®.
Operational For optimum pH measurements, follow these steps: Guide Understand the bene ts and features of Standard and Basic Operation. Set up edge ® meter by con guring preferences. Calibration Measurement The “Standard” pH operation includes up to a 5-point bu er calibration, use Basic vs of custom bu ers, choice of displaying 0.001 pH resolution, use of the full Standard...
edge’s pH meter operation is con gured by using the SETUP key with a pH probe pH Meter connected to the meter. The parameter-speci c options will be seen inserted Con gurations into the menu. If Basic mode is “On”, the pH parameter list will not be displayed. See Basic mode for a description of this operation before choosing how to SETUP the meter.
Page 22
Preparation Pour small quantities of the bu er solutions into clean beakers. If possible, Calibration use plastic to minimize any EMC interferences. For accurate calibration and to minimize cross-contamination, use two beakers for each bu er solution; one for rinsing the electrode and one for calibration. If you are measuring in the acidic range, use pH 7.01 or 6.86 as the rst bu er and pH 4.01 (or 3.00*) as the second bu er.
Rinse and submerse the pH electrode approximately 3 cm (1¼”) into the second bu er solution and stir gently. If necessary, press the ARROW keys to select a Calibration di erent bu er value. The “ ” along with “STIR” tag will be displayed and “WAIT” will blink on the LCD until the reading is stable.
Page 24
Note: Custom bu er values can be adjusted ±1.00 pH around the set value during calibration. When a custom bu er is displayed, the “C1” or “C2” Calibration tag is displayed. First Calibration Point When performing a new calibration, or adding to an existing one, the user has a choice of how the rst new calibration point will be treated in reference to the existing calibration point.
Page 25
Calibration If the instrument is not calibrated the calibration history has been deleted, or it has been calibrated only at one point, the electrode condition and the electrode response gauges will be empty. For a continuous display of the electrode’s condition and response, daily calibration is necessary.
Page 26
When the reading is stable and close to the selected bu er, the “CFM” tag will blink. Press CFM to con rm Calibration calibration. After con rming the rst calibration point, the calibrated value will be displayed on the rst LCD line and the second expected bu er value on the third LCD line.
• If the bu er temperature exceeds the temperature limits of the bu er, “WRONG BUFFER TEMPERATURE” will be displayed. Calibration • Press CLR after entering calibration to clear all calibration options. “CLEAR ALL” message will be displayed and the instrument will return to measurement mode displaying “CAL DUE”...
Page 28
Calibration Wrong Old Points Inconsistency “WRONG OLD POINTS INCONSISTENCY” is displayed if Messages the new calibration di ers signi cantly from the last value of that sensor in that bu er. In this case it may be best to clear the previous calibration and attempt a new calibration with fresh bu ers.
Calibration Contaminated Bu er Messages “CONTAMINATED BUFFER” appears in order to alert that the bu er could be contaminated. Refresh your bu er and continue the calibration procedure. Broken Temperature Sensor If the temperature sensor should malfunction or break at any time, a temperature of “25.0º C” will blink on the second LCD line and the message “BROKEN TEMPERATURE SENSOR”...
Page 30
Note: If a custom bu er was used in calibration, the “C1” and “C2” tags will be pH GLP displayed. If only the one custom bu er is used in calibration, the tag will be “C1” Information and the value will be displayed. If disabled, “EXPIRATION WARNING DISABLED”...
Page 31
When a pH electrode is connected, the instrument will recognize it and the “REMOVE PROTECTIVE Measurement AND FILLING CAPS” message will be displayed along with “PH”. Press any key to skip the message. The instrument will enter measurement mode. Make sure the electrode has been calibrated before taking measurements.
Range Speci c Messages Displayed On Third LCD Line During Measurement Measurement All the messages described in General Setup (page 11) are displayed in pH range. Temperature Sensor Problem (if there is one) Cal Due or O set and Slope Value Time Date Battery or Charge Status...
If the calculated o set is higher than ± 2000 mV the “MAX LIMIT REACHED“ message will scroll on the third LCD line. Until the error is eliminated the value cannot be con rmed. Press CFM to con rm the relative mV value. The display will show “SAVING“...
Page 34
Relative mV If disabled, “EXPIRATION WARNING DISABLED” is displayed. Information Or if enabled the number of days until the calibration alarm “CAL DUE” will be displayed. (i.e. “CAL EXPIRES IN 2 DAYS”) The number of days since the calibration expired. (i.e. “CAL EXPIRED 2 DAYS AGO”).
When an ORP electrode is connected, the instrument Relative mV will recognize it and the probe code will be displayed Measurement along with “ORP”. Press any key to skip the message. The instrument will enter measurement mode. Rinse the sensor with water and a sample if possible. Submerse the electrode tip approximately 3 cm (1¼”) into the sample to be tested and stir sample gently.
Maintenance pH Probe Maintenance Remove the protective cap of the pH electrode. DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal with electrodes. They will disappear when rinsed with water. During transport, tiny bubbles of air may form inside the glass bulb a ecting proper functioning of the electrode.
Page 37
pH Probe Replace the solution in the protective cap with a few drops of HI 70300 or Maintenance HI 80300 Storage Solution or, in its absence, Filling Solution (HI 7082 or HI 8082 for double junction electrodes). Follow the preparation procedure before taking measurements. Note: NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER.
Page 38
pH Probe Alkali Error with 1.0 mol/L Na Maintenance 11.5 0.11 0.01 0.46 12.0 0.21 0.06 0.62 12.5 0.32 0.11 0.79 13.0 0.43 0.15 13.5 0.45 0.21 14.0 0.65 0.27 ORP Probe Remove the protective cap o the ORP electrode. Wash o any SALT DEPOSITS.
After performing any of the cleaning procedures, rinse the electrode thoroughly ORP Probe with distilled water, and re ll the reference chamber with fresh electrolyte (not Maintenance necessary for gel-filled electrodes) and soak the electrode in HI 70300 or HI 80300 Storage Solution for at least 1 hour before taking measurements. Temperature dependence for ORP sensors The oxidation-reduction potential (ORP) displayed in mV, is the voltage that results from the di erence in potential between the platinum metal surface and...
Speci cations pH, mV in pH, mV in ORP Temperature -2.00 to 16.00 pH -2.000 to 16.000 pH* Range -20.0 to 120.0 ºC ; -4.0 to 248.0 °F** ±1000.0 mV in pH ±2000.0 mV in ORP 0.01 pH 0.001 pH* Resolution 0.1 ºC;...
Accessories Electrodes/Probes Triple ceramic, single junction, low temperature glass, re llable pH HI 10530 electrode with conical tip and temperature sensor Single ceramic,double junction, high temperature glass, re llable pH HI 10430 electrode with temperature sensor HI 11310 Glass body, double junction, re llable pH/temperature electrode Glass body, double junction, re llable pH/temperature electrode with HI 11311 enhanced diagnostics...
Page 42
Accessories HI 8061L General Cleaning Solution in FDA approved bottle, 500 mL HI 8073L Protein Cleaning Solution in FDA approved bottle, 500 mL HI 8077L Oil & Fat Cleaning Solution in FDA approved bottle, 500 mL ELECTRODE REFILL ELECTROLYTE SOLUTIONS HI 7082 3.5M KCl Electrolyte, 4x30 mL, for double junction electrodes 3.5M KCl Electrolyte in FDA approved bottle, 4x30 mL, for double junction...
If the repair is not covered by the warranty, you will be noti ed of the charges incurred. If the instrument is to be returned to Hanna Instruments, rst obtain a Returned Goods Authorization number from the Technical Service department and then send it with shipping costs prepaid.
Page 44
Hanna Instruments reserves the right to modify the design, construction or appearance of its products without advance notice.
Need help?
Do you have a question about the edge pH and is the answer not in the manual?
Questions and answers