1 x Premier TENS Unit 1 x Charger and USB Cable 4 x Packs of Self-Adhesive Electrodes (4 electrodes per pack) 1 x Pair of Patient Cables 1 x User Manual Med-Fit UK Ltd. Unit 8 Martel Court S. Park Business Park Hamilton Road Stockport SK1 2AF.
Contents CONTROLS AND FUNCTIONS STEP BY STEP GUIDE CHARGING INSTRUCTIONS WARNINGS AND CAUTIONS GENERAL DESCRIPTION INTRODUCTION TO EMS MANUAL ADJUSTMENTS CHOOSING THE CORRECT PARAMETERS THE PREMIER PLUS EMS PROGRAMMES P1 - P12 EMS & MASSAGE PROGRAMMES EMS ELECTRODE PLACEMENT WARRANTY SKIN PATCH TEST ELECTRODE INSTRUCTIONS TECHNICAL SPECIFICATION...
CONTROLS AND FUNCTIONS Left channel socket Right channel socket Intensity display for channel 1 and 2 Red key lock button Charging port Mode display for both Timer Indicator programmes and The time can only be adjusted manual mode in manual mode, time is set to continuous on all programmes Mode button select programme mode or...
STEP BY STEP GUIDE Before using your premier TENS for the first time, please fully charge the battery see page 7. Press the button to switch on the TENS. This is located at the bottom left hand corner of the TENS device. Please push and hold down the button for 2 seconds to turn the unit on.
Page 6
STEP BY STEP GUIDE Fig1 Connect one of the EMS lead cables supplied to the top of your EMS machine. As shown in Fig1. CH1 = Channel 1 Now connect the other end of the EMS cable to the CH2 = Channel 2 self-adhesive electrodes supplied.
CHARGING INSTRUCTIONS Important information The USB charging cable connects to the EMS charging port as shown in fig 1 (Please ensure you connect the cable the right way round). fig1 Charging Port When the battery has 10% charge, a low battery indicator will appear on the screen.
WARNINGS & PRECAUTIONS PLEASE NOTE: Patients must read and understand the warnings and precautions before using this device. Do not allow your machine or electrodes to be used by anyone else, as they are designed for single patient use only. It is recommended that proper medical advice on the use of TENS is sought from a Qualified Practitioner (Physiotherapist, Doctor or Nurse) before use, to ensure safe and effective treatment.
Page 9
WARNINGS 1. The long term effects of chronic electrical stimulation are unknown. 2. Stimulation should not be applied over the carotid sinus nerves, particularly in patients with a known sensitivity to the carotid sinus reflex. 3. Stimulation should not be applied over the neck or mouth. Severe spasm of the laryngeal and pharyngeal muscles may occur and the contractions may be strong enough to close the airway or cause difficulty in breathing.
GENERAL DESCRIPTION The EM6200A Digital EMS is a battery operated pulse generator that sends electrical impulses electrodes to the body and reach the nerves and underlying muscle group. The device is provided with four controllable output channels, each independent of each other.
Page 11
INTRODUCTION TO EMS HOW EMS WORKS 1. Relaxation of muscle spasms 2. Prevention or retardation of disuse atrophy 3. Increasing local blood circulation 4. Muscle re-education 5. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis 6. Maintaining or increasing range of motion The EMS units send comfortable impulses through the skin that stimulate the nerves in the treatment area.
ADJUSTING THE CONTROLS USING THE MANUAL ADJUSTMENT 1. Power On/Off/Pause Button The power of unit can be turned on by pressing the On/Off/Pause button. You may start to adjust the settings when the liquid crystal display is on. Press and hold for 2 seconds to switch off.
Page 13
ADJUSTING THE CONTROLS USING THE MANUAL ADJUSTMENT 3. Set Control S E T By pressing the "Set" control you select the setting you intend to adjust. The value is set by pressing the "Increment" or "Decrement" controls when the "Set" value is flashing.
Page 14
ADJUSTING THE CONTROLS FOR EMS 10. Steps to Set a EMS Program The settings can be adjusted according to the following steps. a. Turn on the Power b. Select a Mode Select a mode by pressing the "Mode" control. The mode you selected will show up on the top of liquid crystal display.
Page 15
ADJUSTING THE CONTROLS FOR EMS d. Set On Time The On Time controls the length of stimulation. By pressing the "Set" control, the contraction time can be adjusted. Both channels' stimulation is cycled on and off by the contraction and relaxation settings. The range is adjustable from 2 seconds to 90 seconds.The total "ON"...
Page 16
Pulse Width is adjustable from 50 µs to 300 µs. Press "SET" control to enter this ADJUSTING THE CONTROLS FOR EMS f. Set Pulse Width menu, then press “ ” or “ ” to adjust the setting. If no instructions 120 µs setting regarding the pulse width are given in therapy, set the control to the suggested 70- Width...
Page 17
ADJUSTING THE CONTROLS FOR EMS h. Set Timer The treatment time is adjustable between 5 - 60 minutes and Continue(C). Press "SET" control to enter this menu, then press "Increment" or "Decrement" to adjust the setting. The liquid crystal will show the balance treatment time after the stimulation is started.
Frequency Selection With all muscle stimulators, it is important that the frequency, pulse width and work rest times are set correctly, to both maximise the results and use the device in a safe and responsible manner. 15-25Hz (pulse per second) 5-15Hz (pulse per second) 25-50Hz (pulse per second) This frequency range would be selected to improve tone and muscle stability.
Page 19
Work Rest (Contraction /Relaxation) Times When setting the work / rest times it is important that the rest time is at least Ramp Time equal to the work time; for example 4 seconds on time 4 seconds off time. For weaker muscles a larger rest time will be required for example 4 seconds on This allows for muscle recovery and reduces muscle fatigue.
Page 20
MUSCLE STIMULATOR & MASSAGE PROGRAMMES Programmes P1 - P4 Muscle Training & Muscle Re-Education Programmes Choose from one of the programmes 1, 2, 3 or 4. Suitable Conditions These programmes can be used for muscle training, prevention of muscle atrophy. Please follow the electrode placement chart for individual muscles shown on pages 23 to 26.
Page 21
HELPFUL TIPS FOR SUCCESSFUL TENS TREATMENT Once you have familiarised yourself with the controls and features of your EMS device, it is important to place the electrodes in the correct place. This may take 3 or 4 attempts to find the most suitable position, for maximum effect. If you are using two electrodes, place the electrodes directly onto the muscles.
ELECTRODE PLACEMENT Arms Biceps This position is useful for muscle training and for gaining volume using the devices' muscle strengthening programmes. But it is equally very useful for diminishing the effects of lactic acid (substances manufactured by the muscles after sporting exertion and that result in pain during the following few hours).
Page 25
ELECTRODE PLACEMENT Gluteals The Trapezius Muscles and Dorsals This position is ideal for shaping the gluteal muscle. The Legs and Calves The Abdominals It is the abdominal muscle which, when electro-stimulated, will make a six-pack appear. www.tensmachineuk.com Telephone: 0161 429 7330...
Page 26
ELECTRODE PLACEMENT The Abdominals The Pectorals These muscles are very BE SURE to follow the difficult and painful to positioning. work on. So, go on, the DO NOT POSITION ON electrodes positioned THE HEART AREA ! like this will allow you to work on your muscles Telephone: 0161 429 7330 www.tensmachineuk.com...
Limited of Liabilities and Disclaimer of Warranties Med-Fit UK Ltd sole obligation in the case of any breach of its warranties set forth in the paragraph above, shall be, at Med-Fit UK Ltd option, to repair or replace the Product without charge to Purchaser or to refund the purchase price of the Product.
ELECTRODE INSTRUCTIONS Turn Stimulator OFF before applying or removing electrodes Application 1. Skin site must be very clean and dry. Dirty, flaky or oily skin will prevent electrodes from adhering to the skin. If necessary, trim excess hair with scissors. If skin is oily wipe down with an alcohol or electrode skin prep prior to application.
Page 29
SAFETY-TECHNICAL CONTROLS For safety reasons, review the following checklist before using your EM6200A Digital 1. Check the device for external damage. - deformation of the housing. - damaged or defective output sockets. 3. Check the usability of accessories. - patient cable undamaged - electrodes undamaged.
Page 30
GRAPHIC SYMBOLS Degree of Electrical Protection BF Timer Increment Decrement Consult instructions for use Manufacturer Serial number Lock Low battery Pause DC current (DC Power source) Comply with MDD 93/42/EEC requirement as amended by 2007/47/EC. Notify body det norske veritas (DNV) Power The label attached to the back of device contains important information about this device model, supply...
The Premier Plus EMS Programmes P1-P12 PROGRAMME SYN/ALT Rate Width Ramp On Time Off Time Timer (Hz) ( s) (sec) (sec) (sec) (min) ACL repair/joint protection back muscle SYNCHRONOUS Spasm small muscle SYNCHRONOUS Spasm Postoperative SYNCHRONOUS Arthroscope SYNCHRONOUS Disuse atrophy SYNCHRONOUS Shoulder Subluxation SYNCHRONOUS...
This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided, and this unit can be affected by portable and mobile RF communications equipment. 1)* Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit.
Page 33
Guidance and manufacture’s declaration – electromagnetic immunity The DEVICE is intended for use in the electromagnetic environment specified below. The customer or the user of DEVICE should assure that it is used in such an environment. Immunity test IEC 60601 Compliance level Electromagnetic environment test level...
Page 34
Guidance and manufacture’s declaration – electromagnetic immunity The DEVICE is intended for use in the electromagnetic environment specified below. The customer or the user of DEVICE should assure that it is used in such an environment. Immunity test IEC 60601 Compliance Electromagnetic environment test level level - guidance...
Page 35
Recommended separation distances between portable and mobile RF communications equipment and the DEVICE . The DEVICE is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the DEVICE can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the DEVICE as recommended below, according to the maximum output power of the communications equipment.
Need help?
Do you have a question about the Premier Series and is the answer not in the manual?
Questions and answers