Clemco CMS-4 Quick Reference Manual

Clemco CMS-4 Quick Reference Manual

Carbon monoxide monitor-alarm
Hide thumbs Also See for CMS-4:

Advertisement

Quick Links

CMS-4 Carbon Monoxide Monitor-Alarm
Quick-Reference Guide for Common Functions
NOTE: This quick-reference guide is a supplement to the operations manual, which should be read in full.
TURNING ON THE CMS-4 MONITOR-ALARM
1. Press and hold the POWER/MODE button until the monitor
beeps once.
THE WARM-UP SEQUENCE AUTOMATICALLY RUNS
a. All sections of the monitor's screen turn on, the backlight turns on,
and the monitor beeps and vibrates.
b. The monitor cycles through calibration status, date/time, battery
voltage, target gas, and alarm setpoint screens.
c. After the warm-up sequence is complete, the monitor beeps 2 times
and the screen displays CO at the top left, the battery icon at the top
right, and a ppm reading at the lower right. The monitor now is in
Measuring Mode.
NOTE: If the monitor indicates a low battery (only the last bar in the
battery icon shows) or if it sounds the dead-battery alarm, replace the
battery. Refer to the operations manual for instructions.
TURNING OFF THE CMS-4 MONITOR-ALARM
1. Press and hold the POWER/MODE button. The monitor turns off after
4 beeps.
2. Release the POWER/MODE button after the monitor's screen goes blank.
SETTING THE DATE AND TIME
1. Turn OFF the monitor. (See TURNING OFF THE CMS-4 MONITOR-ALARM
above.)
2. Press and hold the AIR and POWER/MODE buttons. Release the buttons
after the monitor beeps 1 time. The monitor beeps again and BUMP
displays in the bottom left of the screen.
3. Press the AIR button 11 times (11 beeps), and DATE displays in the
bottom left of the screen.
4. Press and release the POWER/MODE button once (1 beep).
5. Press the AIR button as often as necessary to adjust the year. Press the
POWER/MODE button to save the year. The month now flashes.
6. Repeat these steps with the AIR and POWER/MODE buttons to adjust
the month, day, hour, and minute settings.
7. After adjusting or confirming all date/time settings, END displays in the
bottom left of the screen for a few seconds, and then DATE displays.
8. Press the AIR button 3 times (3 beeps), and START displays in the
bottom left of the screen.
9. Press and release the POWER/MODE button (1 beep) to begin the
warm-up sequence. After the warm-up sequence ends, the monitor
beeps 2 times. The monitor now is in Measuring Mode.
PERFORMING A FRESH-AIR ADJUSTMENT
1. Be sure the monitor is in a fresh-air environment (free of toxic gases
and with normal oxygen content: 20.9%).
Q U I C K - R E F E R E N C E G U I D E
2. Turn ON the monitor by pressing and holding the POWER/MODE button
until the monitor beeps once.
3. The warm-up sequence automatically runs for about 15 seconds.
4. After the warm-up sequence is complete, the monitor beeps twice and
the screen displays CO at the top left, the battery icon at the top right,
and a ppm reading at the lower right. The monitor now is in Measuring
Mode.
5. Press and hold the AIR button. Air displays at the top of the screen and
HOLD displays at the bottom, and then this changes to AdJ displaying at
the top and RELEASE at the bottom. Release the AIR button.
6. The monitor beeps once, and the display reads as 0 ppm near the
lower-right of the screen.
ASSEMBLING THE CALIBRATION CONNECTOR
ASSEMBLY
(See FIGURES 1 and 2.)
1. Required Material:
a. 0.5 LPM fixed-flow calibration connector with slide valve
b. 3/16" OD calibration tubing
c. Calibration cup
2. Slide one end of the 3/16" tubing onto the barbed end of the calibration
connector with slide valve.
3. Slide the other end of the 3/16" tubing onto the inlet barb of the
calibration cup.
NOTE: The calibration cup has directional flow. The flow-direction arrow
shows the direction of air flow. The tubing connects to the barb on the
backend of the flow arrow. The outlet barb (on the front end of the flow
arrow) remains open.
CALIBRATING THE CMS-4 MONITOR-ALARM
1. Required Material:
a. 25 ppm CO test-gas cylinder.
b. Calibration connector assembly. (See ASSEMBLING THE CALIBRATION
CONNECTOR ASSEMBLY.)
2. Perform a fresh-air adjustment. (See PERFORMING A FRESH-AIR
ADJUSTMENT.)
3. Confirm that the calibration connector's slide valve is CLOSED
by sliding it DOWN, toward the test-gas cylinder connection.
(See FIGURE 1.)
4. Screw the calibration connector onto the test gas cylinder.
5. Turn OFF the monitor by pressing and holding the POWER/MODE button.
The monitor turns off after 4 beeps. Release the POWER/MODE button
after the monitor's screen goes blank.
6. Press and hold the AIR and POWER/MODE buttons. Release the buttons
after the monitor beeps 1 time. The monitor beeps again and BUMP
displays in the bottom left of the screen.
7. Press the AIR button once (1 beep), and GAS CAL displays at the bottom
of the screen.

Advertisement

loading
Need help?

Need help?

Do you have a question about the CMS-4 and is the answer not in the manual?

Questions and answers

Summary of Contents for Clemco CMS-4

  • Page 1 SETTING THE DATE AND TIME NOTE: The calibration cup has directional flow. The flow-direction arrow 1. Turn OFF the monitor. (See TURNING OFF THE CMS-4 MONITOR-ALARM shows the direction of air flow. The tubing connects to the barb on the above.)
  • Page 2 Calibration Cup 16. Press the POWER/MODE button once (1 beep) to start the warm-up 3/16” ID Tubing sequence and to return to Measuring Mode. Figure 2 Open End of Calibration Cup CMS-4 Monitor Upper Retaining Clip OPEN Slide Valve !"#$!%...