JUPITER Table of contents: PRECAUTIONS AND SAFETY MEASURES ............... 2 1.1. Preliminary instructions ..................... 2 1.2. During use ......................... 3 1.3. After use ..........................3 1.4. Definition of Measurement (Overvoltage) category ............3 GENERAL DESCRIPTION ................... 4 ...
Only the leads supplied with the instrument guarantee compliance with the safety standards. They must be in good conditions and replaced, if necessary, exclusively with HT original accessories. Do not test circuits exceeding the specified voltage limits. Do not perform any test under environmental conditions exceeding the limits indicated in §...
JUPITER 1.2. DURING USE Please carefully read the following recommendations and instructions: CAUTION Failure to comply with the caution notes and/or instructions may damage the instrument and/or its components or be a source of danger for the operator. Before activating the rotary switch, disconnect the test leads from the circuit being measured.
JUPITER 2. GENERAL DESCRIPTION The instrument carries out the following measurements: DC / AC, AC+DC TRMS voltage DC / AC / AC+DC TRMS voltage with low impedance (LoZ) DC / AC / AC+DC TRMS current with standard clamp transducers ...
JUPITER 3. PREPARATION FOR USE 3.1. INITIAL CHECKS Before shipping, the instrument has been checked from an electric as well as mechanical point of view. All possible precautions have been taken so that the instrument is delivered undamaged. However, we recommend generally checking the instrument in order to detect possible damage suffered during transport.
JUPITER 4.2. DESCRIPTION OF FUNCTION KEYS 4.2.1. Key GO/HOLD , and Pressing the key GO/HOLD (for functions V , LoZV ) makes the instrument hold the value of the quantity shown on the display. The message HOLD appears on the “...
JUPITER 4.2.3. Key MODE/MXMNPK A simple pressing of key MODE/MXMNPK allows for the following operations: Selection of measuring modes AC+DC in positions V , LoZV “ ” “ ” “ ” Selection of measuring modes AC+DC (see § 4.2.8) in “...
JUPITER 4.2.6. LoZ function This mode allows carrying out AC/DC voltage measurement with a low input impedance, in order to eliminate wrong readings due to parasite voltages for capacitive couplings. CAUTION By plugging the instrument between the phase and earth conductor, because of the low impedance of the instrument during measurement, RCD protections may trip while carrying out the test.
JUPITER 4.2.11. Setting the rated voltage in Loop/Ra measurement To set the value of rated voltage necessary to instrument in order to calculate the assumed short-circuit current in position Ra Loop, proceed as follows: 1. Select the position Ra Loop 2.
JUPITER 5. OPERATING INSTRUCTIONS 5.1. DC VOLTAGE MEASUREMENT CAUTION The maximum input DC voltage is 690V. Do not measure voltages exceeding the limits given in this manual. Exceeding voltage limits could result in electrical shocks to the user and damage to the instrument.
JUPITER 5.2. AC, AC+DC VOLTAGE MEASUREMENT CAUTION The maximum input AC voltage is 690V to earth. Do not measure voltages exceeding the limits given in this manual. Exceeding voltage limits could result in electrical shocks to the user and damage to the instrument.
JUPITER 5.3. AC, DC, AC+DC VOLTAGE WITH LOW IMPEDANCE (LOZ) CAUTION The maximum input AC/DC voltage is 690V to earth. Do not measure voltages exceeding the limits given in this manual. Exceeding voltage limits could result in electrical shocks to the user and damage to the instrument.
JUPITER 5.4. RESISTANCE MEASUREMENT AND CONTINUITY TEST CAUTION Before attempting any resistance measurement, cut off power supply from the circuit to be measured and make sure that all capacitors are discharged, if present. Fig. 11: Use of the instrument for resistance measurement and continuity test 1.
JUPITER 5.5. PHASE SEQUENCE AND PHASE CONCORDANCE WITH 1 TERMINAL CAUTION Input AC voltage to carry out this test must be in range 130V ÷ 690V with a frequency in range 42.5Hz ÷ 69Hz. This test can only be performed by touching the metal parts of the conductors.
Page 18
JUPITER 4. In conditions of correct voltage and frequency, the instrument shows message HOLD “ ” symbols and the buzzer sounds continuously, waiting for a stable voltage “ ” value on phase L1 to be detected (see Fig. 14 left side).
JUPITER 5.6. OVERALL EARTH RESISTANCE WITHOUT RCD TRIPPING This function is performed in compliance with standards CEI 64-8 612.6.3, IEC/EN61557-6 and allows measuring the impedance of the fault loop, comparable to the overall earth resistance in TT systems (see § 9.2).
Page 20
JUPITER 1. set the value of rated voltage Phase-Earth (see § 4.2.11) 2. Set the limit value of contact voltage (see § 4.2.10) 3. Select position Ra Loop. 4. Press key MODE/MXMNPK and select option RCD Ra “ ” 5. In case the cable with Shuko plug is used, insert the red conductor into input terminal V ...
Page 21
JUPITER 8. In case no anomalous conditions are present, the instrument carries out the test and symbol flashes on the display. At the end of the test, the following screens appear on the display. Fig. 22: Results of measurement of overall earth resistance 9.
JUPITER 5.7. MEASUREMENT OF LINE/LOOP IMPEDANCE This function is performed in compliance with standards CEI 64-8 612.6.3, IEC/EN61557-3 and allows measuring line impedance, fault loop impedance and the assumed short-circuit current (see § 9.3). The following operating modes are available: ...
Page 23
JUPITER Fig. 24: Use of the instrument for measuring Loop L-PE impedance with leads Fig. 25: Use of the instrument for measuring Loop L-N impedance with cable provided with Schuko plug IT - 21...
Page 24
JUPITER Fig. 26: Use of the instrument for measuring Loop L-N impedance with leads Fig. 27: Use of the instrument for measuring Loop L-L impedance with leads 1. Set the value of rated phase-to-earth, phase-to-neutral or phase-to-phase voltage (see § 4.2.11) 2.
Page 25
JUPITER 6. For Loop L-PE measurements, in case measuring leads are used, insert the red conductor into input terminal V L and the black conductor into input terminal COM/E/N and connect the instrument to the system to be tested (see Fig. 24). The value of voltage Phase-Earth is shown on the display 7.
JUPITER 5.8. TEST ON A AND AC RCDS This function is performed in compliance with standard IEC/EN61557-6 and allows measuring the tripping time and the current of the instant general RCDs of the system (see § 9.1). The following operating modes are available: ...
Page 27
JUPITER Fig. 30: Use of the instrument for RCD test on single-phase system with leads Fig. 31: Use of the instrument for RCD test on three-phase system with leads IT - 25...
Page 28
JUPITER 1. Set the limit value of contact voltage (see § 4.2.10) 2. Select position RCD 3. Press key MODE/MXMNPK and select one of the following options: Mode AUTO RCD T Measurement of the RCD s tripping time in automatic ’...
Page 29
JUPITER 10. The first test carried out is the one with test current In and polarity 0°. The result of partial measurement is shown on the display with indication NOT OK (see Fig. “ ” “ ” left side). Symbol flashing indicates that the RCD must be reset.
Page 30
JUPITER Tripping time in manual mode (test current In) Fig. 35: Tripping time in manual mode at In 10. The correct measurement result is shown on the display (see Fig. 35 left side) with – indication “ ” 11. The correct measurement result is shown on the display (see Fig. 35 right side) with –...
The instrument carries out the measurement both with flexible clamp transducer (optional accessories) and with other standard clamp transducers in the HT family (optional accessories). With transducers having an Hypertac output connector, the optional adapter NOCANBA is necessary to obtain the connection.
Page 32
JUPITER 6. For standard clamp transducers, press key MODE/MXMNPK to select “ ” “ ” AC+DC measurement. Anyway, the instrument automatically recognizes AC or DC “ ” quantities. 7. Insert the red cable into input terminal V L and the black cable into input terminal COM/E/N.
Page 33
The instrument carries out the measurement both with flexible clamp transducers (optional accessories) and with other standard clamp transducers in the HT family (optional accessories). For inrush current with high DC component is recommended the use of AC/DC clamps.
Page 34
JUPITER 9. Upon recognition of the event, measurement stops automatically and the instrument shows, in its main display, the Max RMS value calculated according to the evaluation time of 100ms (default) indicated on the secondary display (see Fig. 43 right side) –...
JUPITER 6. MAINTENANCE CAUTION Only expert and trained technicians should perform maintenance operations. Before carrying out maintenance operations, disconnect all cables from the input terminals. Do not use the instrument in environments with high humidity levels or high temperatures. Do not expose to direct sunlight.
JUPITER 7. TECHNICAL SPECIFICATIONS 7.1. TECHNICAL CHARACTERISTICS Accuracy calculated as [%reading + (no. digits*resolution)] at 23°C 5°, <80%HR DC Voltage (Autorange) Range Resolution Input Accuracy Overload protection impedance 0.0 690.0 (0.5%rdg + 2digits) 1M 690VDC/ACrms AC, AC+DC, LoZ TRMS Voltage (Autorange)
Page 37
JUPITER Harmonic voltage and current (Autorange) Fundamental Accuracy (*) Harmonic order Resolution frequency (non-zeroed values) (5.0rdg+20digits) 0.1V / 0.1A /0.1% 1 25 42.5Hz 69Hz (5.0rdg+10digits) (10.0rdg+10digits) THD% 0.1% ’ Accuracy of harmonic amplitudes expressed in % is evaluated considering the accuracy of parameters...
JUPITER 8. ASSISTANCE 8.1. WARRANTY CONDITIONS This instrument is warranted against any material or manufacturing defect, in compliance with the general sales conditions. During the warranty period, defective parts may be replaced. However, the manufacturer reserves the right to repair or replace the product.
JUPITER 9. THEORETICAL APPENDIXES 9.1. TEST ON DIFFERENTIAL SWITCHES (RCD) Purpose of the test Checking standards that the general differential protection devices (G) have been correctly installed and adjusted and that they maintain their characteristics over time. This check must make sure that the RCD trips at a current non higher than its rated operating current IdN and that the tripping time does not exceed the maximum time indicated in the standard for general RCDs (according to the description in Table 2).
JUPITER 9.2. OVERALL EARTH RESISTANCE IN TT SYSTEMS Purpose of the test Checking that the protection device is coordinated with the value of earth resistance. We cannot assume a priori a reference limit value for earth resistance as a reference when checking measurement results.
JUPITER 9.3. LOOP AND CALCULATION OF ASSUMED SHORT-CIRCUIT CURRENT Purpose of the test Fault ring" (Loop) indicates the circuit run through by the current generated by an “ insulation fault to earth (dead short). The fault ring includes: The phase winding of the transformer.
JUPITER 9.4. VOLTAGE AND CURRENT HARMONICS Any periodic non-sinusoidal wave may be represented by a sum of sinusoidal waves, each with a frequency which is a whole multiple of the fundamental, according to the relationship: ...
Page 44
JUPITER Odd Harmonics Even Harmonics Not multiple of 3 Multiple of 3 Relative Voltage Order h Relative Voltage Relative Voltage Order h Order h %Max %Max %Max 6..24 Table 3: Limits for the harmonic voltages the supplier may introduce into the network.
Need help?
Do you have a question about the JUPITER and is the answer not in the manual?
Questions and answers