Changing the Helium tank
1a
1b
2
1a
Grasp both sides of
Grasp helium tank and
helium tank panel and
slowly slide drawer out.
1b
Pull out to open.
7
8
Install fresh helium tank.
Fully tighten yoke T-handle
clockwise.
Note: Once the helium alarm sounds, there are
24 Autofills remaining in tank.
| 23 |
IAB Insertion / CARDIOSAVE Operation
3a
3b
3a
Remove helium tank
knob from Holder and
3b
Attach to left end
of helium tank, then
fully turn clockwise
to close.
9
Slowly open helium
tank knob counter-
clockwise (listen for
any escaping helium).
|
Cardiac Assist
|
4
5
Slowly loosen yoke
Remove helium tank.
T-handle counter-
clockwise (some helium
may escape).
10
11
Slide helium tank drawer
Verify full helium level
in and replace helium
via icon on Monitor Display.
tank panel.
6
Replace washer,
if available.
He
CARDIOSAVE
OPERATION