16. TROUBLESHOOTING
16.1.
Site~Rite® 8 Ultrasound System Troubleshooting
SYMPTOM
Error Code
EC:201-203
EC:205-206
EC:211
Error Code
EC:204
Error Code
EC:101
System plays a sound
every minute.
Error Code
EC:110
Power Problem
A/C adapter indicator
does not change states
when A/C power is
connected
System Does Not Charge
Error Code
EC:102
Error Code
EC:103
®
Site~Rite
8 Ultrasound System
Troubleshooting – Site~Rite
DESCRIPTION
Ultrasound Hardware Failure
Probe Disconnect Error
System temperature too high
Low Battery
Battery Error
System will not turn on or turns
on but immediately turns off
Connection Error
Battery Charging Failure
USB Storage Device Error
Hard Drive Error
®
8 Ultrasound System
Reboot the system.
Note: If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (Section 11.5).
If applicable, ensure the probe is securely connected to the system. See
Section 3.3.
Note: If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (Section 11.5).
Shut down the system and allow to cool.
Note: If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (Section 11.5).
Connect system to power for continued operation and to recharge battery.
Connect AC adapter for operation. Ensure the battery switch is in the ""
position
(Section 4).
Note: If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (Section 11.5).
1.
Press and hold the power switch for at least 2 seconds to turn the
system on.
2.
Connect the system to AC power for operation and to recharge the
battery.
Note: If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (Section 11.5).
1.
Unplug the AC adapter and plug it back into the system.
2.
Restart the system if necessary.
Note: If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (Section 11.5).
1.
Unplug the AC adapter and plug it back into the system.
2.
To charge the battery in temperatures exceeding 32°C, the system may
need to be turned off.
Note: If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (see Section 11.5).
The USB storage device may be full or may not be a compatible USB storage
device. Replace the USB storage device.
Note: If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (see Section 11.5).
Restart the system.
Note: If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (see Section 11.5).
SOLUTIONS