Page 1
® ® MEDIREG II, MEDISELECT MEDICAL HIGH PRESSURE REGULATORS Document Nr: 735 800 000 092 Date of issue: 12/02/2010 Revision Nr: Regulators IFU GB,FR,DE,NL Rev 03 Page 1/44 12/02/2010...
Page 2
• Use only the product and its associated equipments in well ventilated area. Before initial use the product should be kept in its original packaging. GCE recommends use of the original packaging (including internal sealing bag and caps) if the product is withdraw from operation (for transport, storage).
Page 3
-20/+60 ° C -30/+60 ° C 10/100 % 10/100 % 600/1200 mbar 600/1200 mbar • In case of storage at a temperature below -20 ° C, d o not operate the regulator until it has been allowed to increase its temperature to a minimum of -20 ° C •...
Page 4
$ % & ' " GCE regulators can be supplied with a flow-metering device - flow control head “C” or flowmeter “D”. This function is used to supply a gas flow (l/min) at atmospheric pressure directly to the patient through the flow outlet “E”, e.g. through a cannula or a facemask.
Page 5
• Check if the product service is due or that the total life time of the product and the gas cylinder has not been exceeded, (refer to GCE or owner’s date coding system). If service or life time has been exceeded, remove the product (or the gas cylinder) from service &...
Page 6
Do not use excessive force. • If any leakage is detected, use the procedure in chapter 6.3 and return the product to GCE for service. • Ensure the flow control knob is on the “ZERO” position. • Ensure the cylinder valve is open – in the “ON” position.
Page 7
Pressure outlet II • Ensure the counterpart is compatible with the pressure outlet features. • Screw the counterpart. • Check the counterpart is fully screwed. • When is pressure outlet used by medical product with high flow consumption (for example lung ventilator with request of source flow 100 l/min at minimal pressure 2,8 bar) check the required capacity of source device with regulator pressure outlet performance listed in appendix 1.
Page 8
• Ensure that the flow control knob is on the ZERO position (if any). • Ensure the accessory IS NOT connected to the pressure outlet. • Slowly open the cylinder valve by turning the hand wheel in anticlockwise direction approx 1 to 1½ turns. •...
Page 9
• Quick coupler. The repairs shall be carried out by a GCE authorised person only. Any product sent back to a GCE authorised person for maintenance shall be properly packaged. The purpose of the maintenance has to be clearly specified (repair, overall maintenance).
Page 10
& workmanship. GCE shall not be liable for loss of production, loss of profit or any other consequential damage or indirect loss. In the event of any fault in the goods due to defective design,...
Page 11
The liability of the regulator is irrevocably transferred to the owner or operator to the extent that it is modified, serviced or repaired by personnel not employed or authorised by GCE or if the apparatus is used in a manner not conforming to its intended use.