[5] Troubleshooting

5-1. Alarms

This incubator is provided with the following alarms. If an alarm condition should occur, check for a
possible cause of the alarm and take the proper measures. If the incubator seems to be defective,
it is in need of repairs. Indicate on the incubator that it is out of order and contact your local Atom
representative.
Cat-
Alarm name
egory
MC high
Incubator air tem-
temperature
perature is rising.
alarm
MC (override
Incubator air tem-
mode ) or SC
perature is rising.
high tempera-
ture alarm
Incubator air tem-
Set temperature
perature deviates
alarm
from the set tem-
(MC 3˚C)
perature.
Skin temperature
Set temperature
deviates from the
alarm (SC 1˚C)
set temperature.
Skin tempera-
Abnor mality in
yellow skin tem-
ture probe
perature probe.
alarm (MC)
Check the con-
Skin tempera-
nection of the yel-
ture probe
low skin tempera-
alarm (SC)
ture probe.
❋1 If any disruption should occur to the skin temperature 2 probe, warning information will be given to the
user.
❋2 The alarm will not occur for 30 minutes when the setting is changed.
WARNING
Condition causing
Message
the alarm
This alarm will occur
before the incubator
air temperature ex-
ceeds 38˚C.
This alarm will occur
before the incubator
air temperature ex-
ceeds 40˚C.
This alarm will occur
if the incubator air
temperature deviates
from the set tempera-
ture by 3˚C or more
This alarm will occur
if the skin tempera-
ture1 deviates from
the set temperature by
❋2
1˚C or more
.
This alarm will occur
if a wire of the skin
temperature 1 probe
has snapped or short-
❋1
circuited.
This alarm will occur
if the skin tempera-
ture 1 probe is not
connected in the SC
❋1
mode.
The unit during the
Resetting the
alarm condition
alarm
The alarm will be
The incubator
reset automatically
heater will be
when the incuba-
turned off.
tor air temperature
falls below 37˚C.
The alarm will be
The incubator
reset automatically
heater will be
when the incuba-
turned off.
tor air temperature
falls below 38˚C.
The alarm will be
T e m p e r a t u r e
reset automatically
control will be
when the deviation
continued.
i s m a i n t a i n e d
❋2
.
within 3˚C.
The alarm will be
T e m p e r a t u r e
reset automatically
control will be
when the deviation
continued.
i s m a i n t a i n e d
within 1˚C.
The skin tem-
perature probe
will not function.
(The skin tem-
Remove the skin
perature is not
temperature probe
displayed.)The
from the infant.
heater will be
turned off (in the
SC mode).
The alarm will be re-
The heater will
s e t a u t o m a t i c a l l y
be tur ned of f.
when the skin tem-
(SC will not func-
perature probe is con-
tion.)
nected.
Troubleshooting
Alarm
Priority
silence
10
Medium
min.
10
Medium
min.
15
Low
min.
15
Low
min.
15
Low
min.
10
Medium
min.
61