Section 13
13.
AK Protocol Format
The master computer and the Model 650 analyzer communicates via the RS232 serial link. The
Model 650 analyzer acts as a "slave" and only responds to commands.
Serial Interface Parameters:
1)
Baud from 300 to 9600 bps, can be selected via the display.
2)
7 or 8 data bits, 1 or 2 stop bits, and the parity (yes/no).
3)
The data transmission is full duplex (no echo) with XON/XOFF protocol.
4)
The "don't-care" byte" (byte 2) is adjustable (factory setting 20H).
Command Format:
<STX>
don't care
function code
space 20H
channel N° always "K0" for the analyzer
space
data
<ETX>
Answer Format:
<STX>
don't care
function code
space
status
space
data
<ETX>
13.1. Serial Interface and AK-Commands
The serial interface enables remote control of the Model 650 analyzer by
It is implemented as an RS232 V24 interface and meets all requirements of the AK protocol.
A 9-pin male connector at the back of the unit is used to connect a master computer with the
following pin assignment:
Pin 3 = Txd (transmit)
Pin 2 = Rxd (receive)
Pin 5 = Gnd (ground)
13.2.
Interface Parameters
Baud rate:
Data bits:
Stop bit:
Don't care:
Parity:
XON/XOFF:
California Analytical Model 650 Noxygen
650 NOXO2_v9.0.doc
02H Example: ASTZ K0
any byte (default 20H)
code 4 byte long (e.g., ASTZ)
20H
20H (onIy if followed by data, otherwise <ETX> )
data bytes (depending on the command)
03H
02H Example: STZ 0 SREM STBY
adjustable, factory setting 20H
same code as command package (e.g., ASTZ)
20H
0 without error or 1 to 9 when error (see also ASTF
command)
20H (onIy if followed by data, otherwise <ETX> )
parameter (depending on the command)
03H
9600, 4800, 2400, 1200, 600, 300 baud
7 or 8
1 or 2
1 byte, adjustable (e.g. 32)
Even, odd, none
Active or not active
Model 650 Noxygen Analyzer
a master computer.
Operators Manual
Page 68 of 86