A A p p p p l l y y i i n n g g l l i i n n e e n n s s
To apply linens:
1. Part the D-rings (C) (Figure 2).
2. Thread the four linen corners through the D-rings (C) attached to the bottom cover to secure the linens (A) to the support
surface (B).
N N o o t t e e - - Do not pull the linens tight. Keep the linens loose and as smooth as possible on top of the support surface for
effective use of the T T r r u u T T u u r r n n .
T T r r a a n n s s f f e e r r r r i i n n g g a a p p a a t t i i e e n n t t f f r r o o m m o o n n e e p p a a t t i i e e n n t t s s u u p p p p o o r r t t p p l l a a t t f f o o r r m m t t o o a a n n o o t t h h e e r r
W W A A R R N N I I N N G G
• Never transfer or transport a patient while T T r r u u T T u u r r n n is in use.
• Do not use the support surface as a transfer device.
• Do not use the support surface handles to lift or move the support surface with a patient on board.
• Do not stick needles into a support surface through the support surface cover. Holes may allow body fluids to enter the
inside (inner core) of the support surface.
• Do not exceed the safe working load of the hospital bed frame when supporting both the patient and the support surface.
Excess weight could cause unpredictable safety and performance of this product.
• Always make sure that the opposite siderail is raised when placing a patient on the support surface to reduce the risk of
patient fall.
To transfer the patient from one patient support surface to another:
P P r r e e r r e e q q u u i i s s i i t t e e : : Follow hospital protocols required to transfer a patient from one surface to another.
1. Position one patient support platform alongside the other patient support platform while minimizing the gap between the
two platforms.
2. Set the brakes to O O N N for both patient support platforms.
3. Adjust the patient support platform heights so that they are level with one another.
4. Transfer the patient following all applicable safety rules and institution protocols for patient and operator safety.
EN
10
A
C
B
F F i i g g u u r r e e 2 2 – – A A p p p p l l y y l l i i n n e e n n s s
2872-009-005 Rev B.1