Summary of Contents for Shenchen LabV Series

  • Page 1 Shenchen Precision Pump Manual of LabV Series...
  • Page 2 Important Information Please read the manual carefully before operating the product. Warning Tubing may have crack due to wear. It results in the overflow of fluid from tubing. In that time human body and instruments may be damaged. So user must check usually and change tubing in time. Connect directly the power line to the wall socket, and avoid using the extense electric line.
  • Page 3: Table Of Contents

    LabV Series Product Appearance ..........- 1 - LabV Series Keyboard Instruction ..........- 2 - LabV Series Operation Interface Structure: ......... - 3 - Main Functions Operation Process ..........- 14 - LabV Series Technical Specification .......... - 18 - LabV Series Function and Features ..........
  • Page 4: Labv Series Product Introduction

    Shenchen Pump LabV Series LabV Series Product Introduction: 4.3 inch color touch screen control; dynamic display working status; flow data, setted parameters and system settings displays in the same screen. Intelligent calibration and online micro adjusting function; three measurement mode; fixed volume measurement, fixed time and volume, timer start and stop.
  • Page 5: Labv Series Keyboard Instruction

    Shenchen Pump LabV Series LabV Series Keyboard Instruction Stop Button Press stop button, stop working. Forbidden buttons can be used on the main interface. Full Button When stop state or transferring state, press this button, the pump will running with highest speed. This button can be used for wash tube or fast filling...
  • Page 6: Labv Series Operation Interface Structure

    Start Button: Press this button, the motor start running. When fixed volume measurement function or fixed time and volume function turn on, press this button, the pump will start work with the function. LabV Series Operation Interface Structure: LabV Series Operation Interface Instruction Boot Interface: After power on the system, enter the welcome interface, choose system language, Chinese/English.
  • Page 7 Shenchen Pump LabV Series Main Interface: Main Interface Composition as below: Speed/Flow Rate Display: In the flow rate mode, display the current flow rate, the motor speed is displayed at the C frame. In the rotating speed mode, display the current set up speed, flow rate is displayed at the C frame.
  • Page 8 Shenchen Pump LabV Series start and stop function is turn on. System Setting Button Click this button, set up other parameters. Flow Calibration Button Click this button enter the flow rate calibration interface. Date & Time Button Click this button enter set up current date and time interface.
  • Page 9 Shenchen Pump LabV Series Unit/Clr Button When input flow rate or volume, this button is unit switch, you can choose different unit. When it is Clr, you can clear the current input data. Backspace Button: Delete an input digital. ESC Button: Cancel the current input, back to prev interface.
  • Page 10 Shenchen Pump LabV Series Click the Setting button in the main interface, then click Suck-Back button to enter the back suction angle setting interface. Click angle button, enter suck-back angle then click ENT. This can set all suck-back angle when motor stop running, except dispensing is ON.
  • Page 11 Shenchen Pump LabV Series external control interface . B. Various independent external control mode setting switch work only when the corresponding external control function is turned on. 7. External Speed Control Setting Interface: External Speed Control Setting Interface as below:...
  • Page 12 Shenchen Pump LabV Series 8. Communication Setting Interface: Communication Setting Interface as below: Set baud rate Communication protocol ShenChen RS232/RS485 ON/OFF Slave No. Click Setting button in the main interface, then click Communication button to enter Communications Settings interface. This pump support MODBUS--RTU Mode.
  • Page 13 Shenchen Pump LabV Series 9. Flow Rate Calibration Interface: Flow Rate Calibration Interface as below: Actual Vol. Volume adjust FixedVolume 0.0000ml +0.0000ml Volume Test 10.00 Run Time 1.00 1.00 Reset The top left corner display the function, when fixed volume measurement turn on, display fixed volume;...
  • Page 14 Shenchen Pump LabV Series CAL button, the calibration is finished. Now the request flow rate or volume is close to theoretical data. E. If request higher accuracy, you can click Add and Dec button to micro adjust the flow rate, to reach high accuracy transferring and dispensing.
  • Page 15 Shenchen Pump LabV Series 10. Date & Time Interface: Setting Date & Time Interface as below: 12-hour 2012-3-15 24-hour 8:45:35 AM Set Date Thursday Set Time Back Click the System Setting button from the main interface, click Date and Time button, enter date and time setting interface.
  • Page 16 Shenchen Pump LabV Series Fixed volume Volume 600.00 After turn on this function, the peristaltic pump will measure the volume automatically, when the volume reach set up volume, the pump will stop working automatically. The flow rate can be changed during the pump working.
  • Page 17: Main Functions Operation Process

    Shenchen Pump LabV Series Peristaltic pump transfer fixed volume in fixed time, transfer number of times is the run times, click suspend time button then input suspend time, prompt box display current diameter, after click the OK button, click the start button, the pump begin dispensing according to the diameter.
  • Page 18 Shenchen Pump LabV Series Boot into Click the Calibra- the main tion button, cali- interface brate the flow rate Click here input flow rate If fixed volume Turn off fixed or dispensing volume or disp- Back to the main inter-...
  • Page 19 Shenchen Pump LabV Series 3. Dispensing Boot into Back to the main inter- the main face after finish interface calibration, click start key start working Click the setting button, choose pump head and Display the The pump display the current...
  • Page 20 Shenchen Pump LabV Series The current time 01:30:00 PM 01:30:00 PM StartTime OverTime Over time Start time 08:30:00 AM 05:30:00 PM Switch for the timer start Switch for the timer stop Timer start repeat date Timer stop repeat date Once...
  • Page 21: Labv Series Technical Specification

    Protection rating IP31 control LabV Series Function and Features 4.3 inch color touch screen control, dynamic display working status, the flow volume and motor speed display in the same screen. Intelligent calibration function, it can calibrate the flow rate and dispensing volume, ensure the flow accuracy, suitable for high accuracy transferring liquid.
  • Page 22: Product Dimension (Mm)

    Shenchen Pump LabV Series automatically after the volume reaches the set value. During this process, the volume can be changed. It is suitable for laboratory liquid dosing and chemical reaction process. Fixed time and volume function: After turn on this function, the peristaltic pump will transfer fixed volume within set time.
  • Page 23: External Control Interface Instruction

    Shenchen Pump LabV Series External Control Interface Instruction External control interface as below: 0-10V 0-5V 4-20mA I-/V- CW/CCW R/S 2 R/S 1 OUT1 OUT2 . Analog signal input terminal: Analog Signal Ext. Speed to 600rpm through analog signal. 0-10V: 0V to 10V voltage signal input terminal.
  • Page 24 Shenchen Pump LabV Series 5V level input 24V level input 12V level input GD2: External control signal common input terminal. NC: Null. CW/CCW: External control direction signal input rising edge Pulse mode: the direction change once when receive one pulse signal effectively Level mode: when high level the pump clockwise running,when low level;...
  • Page 25 Shenchen Pump LabV Series . The motor working status output terminal: Output motor working status as below: If connect with relays, when the motor running, K1 connect; when the motor stop running, the K1 disconnect. . RS232 Communication : Choose RS232 in the Communication setting interface, this terminal is active.
  • Page 26 Shenchen Pump LabV Series . RS485 Communication Interface: Choose RS485 in the communication setting interface, this terminal is active. GD1: RS485 signal interface A+: Connect RS485 A+ terminal B-: Connect RS485 B- terminal Instruction: No matter choose RS232 or RS485, the communication protocol is standard MODBUS protocol.
  • Page 27: Maintenance

    Shenchen Pump LabV Series Maintenance When pump is not working, please loose the cartridges of pressing the tubing for avoiding changing the shape of tubing because of longtime extrusion. Keep the rollers of pump head clean and dry, otherwise it can quicken the tubing wearing, reduce the useful life of tubing and lead the rollers to damage in earlier.