15.Error Message & Trouble Shooting
Error
Err 01
Initial value of A/D is fail
Load cell connection failure,
Err 02
Error in A/D conversion part
Err 05
Input value is over the range
Err 07
Product ID is larger than 19
Hi limit weight is larger than
Err 08
Maximum weight
Low limit weight is larger
Err 09
than hi limit weight
Err 11
The unit of weight is different
Zero value exceeds the initial
Err 13
zero range
Err 21
The resolution exceeds 1/20,000
he weight for span calibration
Err 22
is set to under 10% of maximum
capacity
The weight for span calibration
Err 23
is set to over 100% of maximum
capacity
Err 24
The value of Span is high or low
The password before and the cur-
Err 31
rent password are not same
Err 32
The number of sample is too low
Err 50
Internal memory data is erased
Reason
Check the L/C connector & test
A/D conversion in Test Mode 2
Check the L/C connector & test
A/D conversion in Test Mode 2
Input the value from 32 to 255
The range of Product ID is 0 to 19
Set the Hi limit weight to under
the maximum weight
Set the Low limit weight to under
the Hi limit weight
Can't process cause the unit of
weight calibration is not same
Remove a weight from the platter
and turn on power
Lower the resolution
The weight of Span calibration
Should be set to over 10% of max.
Capacity (Check it in CAL 1)
The weight of Span calibration
Should be set to under 100% of
max. capacity (Check it in CAL 1)
Adjust resolution
Input the new password again
Put more samples or
Decrease the number of samples
The product is in trouble
Contact the A/S center