Contents Cautions in Using with Safety ............... 1 Explanation ........................1 Table of Illustrated Symbols ..................2 Fundamental Matters of "WARNING!" and "CAUTION!" ..........3 Fundamental Matters of "WARNING!" and "CAUTION!" ..........4 Before Using This Unit ................5 ...
Cautions in Using with Safety Explanation MEANING OF ILLUSTRATED SYMBOLS Illustrated Symbols Various symbols are used in this safety manual in order to use the unit without danger of injury and damage of the unit. A list of problems caused by ignoring the warnings and improper handling is divided as shown below.Be sure that you understand the warnings and cautions in this manual before operating the unit.
Cautions in Using with Safety Table of Illustrated Symbols Warning Warning Warning Warning Warning Warning generally high voltage high temperature drive train explosive Caution Caution Caution Caution Caution Caution generally electrical shock scald no road heating not to drench Caution Caution water only deadly poison...
Cautions in Using with Safety Fundamental Matters of "WARNING!" and "CAUTION!" WARNING! Do not use this unit in an area where there is flammable or explosive gas Never use this unit in an area where there is flammable or explosive gas. This unit is not explosion-proof.
Cautions in Using with Safety Fundamental Matters of "WARNING!" and "CAUTION!" CAUTION! When opening the cover... Make sure that the pressure of equipment has decreased to 0(zero) MPa before opening the cover. Generally the cover does not open due to the safety lock mechanism under the high pressure condition.
Before Using This Unit Requirements for Installation 1. Choose a proper place for installation Do not install this unit in a place where: Rough or dirty surface. Flammable gas or corrosive gas is generated. Ambient temperature 35C and above or 5C and below. ...
Choose a correct power distribution board or receptacle that meets the unit’s rated electric capacity. Operating voltage range for respective equipment models are as follows. SK101C models: AC100 to 120V SK111C models: AC200 to 240V Electric capacity: SK101C: AC100V-120V 12.5A-15A, SK111C: AC200V-240V 7.5A-9A...
Interior Wiring on the electric power equipment be "OFF". Note that Black Power Supply Side SK101C and SK111C do not equip with the power plug. Select and connect the appropriate plug or terminal White Power Supply Side corresponding to the power capacity that is adjusted to...
Before Using This Unit Installation Procedure Set the attached drain board onto the bottom Set the drain board surface inside the chamber horizontally. The drain board stabilizes the sterile samples inside the chamber as well as protects the heater and sensor.
Before Using This Unit Installation Procedure Close the door before operation Place the sterilizing drum with properly-piled items into the pot with the positioning mark on the pot lid to be aligned to the center of the lower handle. Press the lid gently by hand and turn it clockwise for the upper and lower handles to close completely and the safety lock to remain in open position, and the safety lock will lock the upper and lower handles automatically when the pressure in pot rises.
Description and Function of Each Part Main Unit Pressure gauge Safety valve Steam exhaust valve Vent Upper handle Lower handle Door lock check Yamato label Operation panel drain valve Power switch Rubber foot...
Description and Function of Each Part Control Panel (Keys) Name Function SET key This key is used to select temperature/time mode. ▲ key This key is used to increase the temperature/time parameter ▼ key This key is used to decrease the temperature/time parameter Water lever lamp...
Operation Method Apparatus/Liquid Sterilization Course Follow the procedures below for the setting of apparatus sterilization and liquid sterilization courses. 1. Pour the water to the designated scale in chamber, lay the sterilization samples in place and close the cover, as shown in the right figure, the green water level indicator lamp, yellow lid lock indicator lamp and green temp.
Operation Method Apparatus/Liquid Sterilization Course 6. when the temperature rises to above 102 ℃, the pressure indicator is light, the top cover cannot be opened 7. When the pressure and temperature reach the set scope, the running light (sterilizing light) will be on and the time will start working for effective heat preservation and sterilization.
Operation Method Sterilizing Parameters Item to be Necessary heat preservation Relative steam Relative steam sterilized time temperature(℃) pressure(MPa) Rubber 0.1-0.105 Dressings 30-45 121-126 0.105-0.14 Utensils 121-126 0.105-0.14 Apparatus 121-126 0.105-0.14 Liquor 20-40 121-126 0.105-0.14 Dishware 121-126 0.105-0.14 Use calibration offset function Calibration offset is a function which corrects the difference between the temperature in furnace and that of controller (sensor temperature) if arises.
Handling Precautions WARNING WARNING! If a problem occurs If smoke or strange odor should come out of this unit for some reason, turn off the power key right away, and then turn off the circuit breaker and the main power. Immediately contact a service technician for inspection.
Handling Precautions CAUTION CAUTION! During a thunder storm During a thunderstorm, turn off the power key immediately, then turn off the circuit breaker and the main power. If this procedure is not followed, fire or electrical shock may be caused. Do not touch the hot section The temperature on the cover and top board on the chamber are very hot during operation or just after operation is completed.
Maintenance Method Daily Inspection and Maintenance For the safety use of this unit, please perform the daily inspection and maintenance without fail. Using the city water to this unit might attach dirt. Do inspect and maintain this point while performing daily inspection and maintenance.
Long storage and disposal When not using this unit for long term / When disposing CAUTION! When not using this unit for long term… Turn off the power and disconnect the power cord. WARNING! When disposing… Keep out of reach of children. ...
In the case of the error other than listed above occurred, turn off the power switch and primary power source immediately. Contact the shop of your purchase or nearest Yamato Scientific Service Office.
If the failure occurs, stop the operation, turn OFF the power switch, and unplug the power plug. Please contact the sales agency that this unit was purchased, or the Yamato Scientific’s sales office. < Check following items before contact >...
Specification Product name Sterilizer Model SK101C SK111C Temperature control 50to 126℃ (sterilization ) range Maximum Operational 0.142MPa Pressure Operational ambient 5℃~35℃ temperature Heater 115V /1500W 220V /1500W Drain valve Globe valve liquid level sensor Float Switch Temperature control PID control by microcomputer...
Wiring Diagram SK101C AC115V SMJ2 blue green SMJ1 SK111C AC220V SMJ2 blue green SMJ1 Symbol Part name Symbol Part name Power switch SMJ1 Control board FU1、2 fuse SMJ2 Display board Heater Temperature sensor Folat Switch Cover interlock switch...
Reference List of Dangerous Substances Never use explosive substances, flammable substances and substances that include explosive or flammable ingredients in this unit. EXPLOSIVE Ethylene glycol dinitrate (nitro glycol), Glycerin trinitrate (nitroglycerine), Cellulose nitrate (nitrocellulose), and other explosive nitrate esters Trinitrobenzene, Trinitrotoluene, Trinitrophenol (picric acid), and other explosive EXPLOSIVE: nitro compounds Acetyl hydroperoxide (peracetic acid), Methyl ethyl ketone peroxide, Benzyl...
Page 27
◆ Any books with missing pages or disorderly binding may be replaced. Instruction Manual for Sterilizer Model SK101C/111C First Edition Jan. 17. 2015 Revision Visit Our Web Site To View Our Inventory Of NEW Yamato Sterilizers and for Technical Help: www.LRE.com Or Call 800-574-2748 323-770-0634...