Digital torquetester/wrenches with continuous rs-232 output (4 pages)
Summary of Contents for Imada MX-500N Series
Page 1
Uploaded by TheGioiCongNghiep.com INSTRUCTION MANUAL TEST STAND MX-500N Series MV-500NII Series Before use, read through this instruction manual to use this unit correctly. Keep this instruction manual as this includes a “Certificate of warranty”...
Page 3
Instruction Manual MX-500N / MV-500NII Series INTRODUCTION Thank you very much for adopting Imada product. MX/MV-500N series is the force measuring instrument which combines with Imada brand digital force gauge or mechanical force gauge. It measures tensile and compressive force by motorized stable speed.
Page 4
Never use this machine for any purposes other than measurement. “Over load stop” function for MX-500N series does not avoid overload completely. In addition, please deal with it according to cautions of an instruction manual attached with a force gauge.
Page 5
Instruction Manual MX-500N / MV-500NII Series OVERVIEW H(mm) Model MX/MV-500N(-S) MX/MV-500N(-S)-L 773 MX/MV-500N-E Table MX-MV-500N-E-L Crosshead Limit knobs Force gauge mount Operation panel Grip mount GF-1 Manual No. MXV50801E...
Page 6
Instruction Manual MX-500N / MV-500NII Series Dimensions of force gauge mount 4 holes 4.5 mm dia. For digital force gauge. 4 holes 3.5 mm dia. For mechanical force gauge Rear view Inlet MV Series Fuse Case MX Series IO Connector EN-00 Connector MV-E Series EN-00 Connector...
Page 7
Instruction Manual MX-500N / MV-500NII Series Operation panel FS LED Mode LED Run LED Up/Down buttons Emergency button FS volume Power switch FS button MX series Speed volume Mode button Run LED Up/Down Buttons Emergency button Power switch MV series Speed volume Manual No.
Page 8
Goes up or down with force gauge. Mount the force gauge here. There are certain holes suitable Force gauge mount for Imada brand force gauges. Table Put a sample here or fix a fixture with GF-1 grip mount. Rear view (Page 6)
Page 9
Instruction Manual MX-500N / MV-500NII Series I/O Connector pin assignment -NG signal OK signal +NG signal Overload signal Signals ground Supplies Below supplies are included with this unit GF-1 grip mount Power supply cord Fuse This instruction manual Carton box Allen key for M3 Allen key for M4 Allen key for M8...
Instruction Manual MX-500N / MV-500NII Series Specifications Models MX-500N Series MV-500NII Series Normal MX-500N MV-500NII Speed/Length meter MX-500N-E MV-500NII-E Scale unit MX-500N-S MV-500NII-S Long stroke MX-500N-L MV-500NII-L Long stroke MX-500N-E-L MV-500NII-E-L Speed/Length meter Long Stroke MX-500N-S-L MV-500NII-S-L Scale unit Attach the following suffix after model name to specify speed range..
Page 11
Crosshead runs up and down with motor Adjustable upper/lower travel Adjustable speed Recommended IMADA brand digital and mechanical force gauges force gauge Except high-capacity models Length/Speed meter: “-E” suffix follows model name Scale unit: “-S” suffix follows with model name Options*2 Long stroke: “-L”...
Page 12
Instruction Manual MX-500N / MV-500NII Series Mounting force gauge Mounting digital/mechanical force gauge Screws (included) Screws (included) M6 flat head screws M6 flat head screws Flap nuts Crosshead Force gauge mount 1. Unscrew two flap-nuts, and release the force gauge mount with taking care not to slip the mount down.
Page 13
Instruction Manual MX-500N / MV-500NII Series Scale unit (only with MX/MV-500N-S) A scale unit included with MX/MV-500N-S series can display travel length of the crosshead so that you can measure elastic character and so on. For details, see an instruction manual for the scale unit. Scale unit: Mitutoyo Digimatic scale unit SDV-30E(Normal) SDV-45E(Long stroke) Scale unit Manual No.
Page 14
Instruction Manual MX-500N / MV-500NII Series Length/Speed Meter (only with MX/MV-500N-E) About… MX/MV-500N-E series has a built-in rotary-encoder which makes it possible to measure travel length and speed by sensing rotation of its feed screw. You can see those measurements with a display unit EN-00 included with. Zero button: Zero the display Length...
Page 15
Instruction Manual MX-500N / MV-500NII Series Procedures Screwing attachment(s) If needed, use optional attachment or standard attachment mounting on the sensing shaft of the force gauge. 1. When attach attachment on the table side, please attach GF-1 on a table. Optional attachment 2.
Page 16
Instruction Manual MX-500N / MV-500NII Series Operating MX series MX-500N series has three operation modes. This unit has “Overload stop” feature that always works while the gauge is connected by optional cable and it is powered on. Also “Force control” feature works when the gauge is connected.
Page 17
Instruction Manual MX-500N / MV-500NII Series Operation in Manual mode 1. Select Manual mode by button. toggles “FS Effective” LED lights up and off. When it is 2. Whenever pressing lights, “Speed knob” is not effective, but the “Fixed speed volume” is. You cannot toggle FS Effective while the crosshead runs.
Page 18
Instruction Manual MX-500N / MV-500NII Series Operation in Continuous cycle mode 1. Select Continuous cycle mode by button. toggles “FS Effective” LED lights up and off. When it is 2. Whenever pressing lights, “Speed knob” is not effective, but the “Fixed speed volume” is. You cannot toggle FS Effective while the crosshead runs.
Page 19
Instruction Manual MX-500N / MV-500NII Series Operation in Force control mode MX-500N series combined with Imada brand force gauge connected through an optional cable (CB-501) can get additional functions as below Forcing stopping the test stand by “Overload” signal from the force gauge. The signal is usually output when the applied force is beyond by a several percent of its capacity.
Page 20
Instruction Manual MX-500N / MV-500NII Series 1. Setup comparator high and low values of the force gauge. For details, please refer an instruction manuals included with the gauge. 2. Connect the gauge and the test stand through the cable (CB-501; sold separately), and turn it on.
Page 21
Instruction Manual MX-500N / MV-500NII Series Operating MV series Single/Continuous cycle operation and I/O connection with the force gauge are excluded for MV series. Operation Motion RUN LED Crosshead runs to the Dark→ Light direction of button figure at Press its maximum speed.
Please return the defective product to us with freight prepaid. 2. This warranty shall not be effective if the product has been subject to below: - Disassembling or modification by other parties than Imada - Misuse, overload or consumption items such as batteries or attachments.
Need help?
Do you have a question about the MX-500N Series and is the answer not in the manual?