Performance Tests
Multirate Optical Interfaces
Procedure 1310nm - Optical Power Output
1 Switch on the HP37718A/19A and recall the DEFAULT SETTINGS as
shown on Page 3-2.
2 Connect the HP37718A/19A Optical Out Port to the HP8153A (ensure
that all connections are tight and that the cable has no twists).
3
Setup the HP 8153A as follows:
a. Press
PARAM
b. Using
,
c. Press
PARAM
d. Using
,
e. Press
PARAM
f. Using
,
g. Press
PARAM
h. Using
,
i. Press the
ZERO
Meter is now ready.
4 Press
to select the Power Level measurement on the HP8153A.
MODE
5 Press
TRANSMIT
or [OC-1].
6 Check the HP8153A Power Meter reading is between 0 dBm and +3
dBm (Typically +1 dBm).
7 Press
TRANSMIT
or [OC-3].
8
Check the HP8153A Power Meter reading is between 0 dBm and +3
dBm (Typically +1 dBm).
9 Press
TRANSMIT
or [OC-12].
10 Check the HP8153A Power Meter reading is between 0 dBm and +3
dBm (Typically +1 dBm).
11 Press
TRANSMIT
OPT] or [OC-48].
12 Check the HP8153A Power Meter reading is between 0 dBm and +3
dBm (Typically +1 dBm).
key to display wavelength [λ]
and
keys, set the wavelength to 1310nm.
key to display Time [t]
and
keys, set the time to 200mS.
key to display REF.
and
keys, set the REF to 0.000dBm.
key to display CAL.
and
keys, set the CAL to 0.000dBm.
key on the Power Meter to calibrate - the Power
or
and select SIGNAL [STM-0 OPT]
SDH
SONET
or
and select SIGNAL [STM-1 OPT]
SDH
SONET
or
and select SIGNAL [STM-4 OPT]
SDH
SONET
or
and select SIGNAL [STM-16
SDH
SONET
3-73