This function adjusts the balance for accurate weighing.
Perform a calibration i n the following cases.
q When the balance is first used.
q When the balance has been moved.
q When the ambient environment has changed.
q For regular calibration.

7-1. Calibration using a weight

1. Warm up the balance for at least half an hour with
nothing on the pan.
2. Press and hold the calibration (CAL) switch until
appears, and release the switch.
Cal
3. The balance displays
To change the calibration weight value, proceed to
step 4.
To use the calibration weight value in the balance
memory, proceed to step 5.
4. Press the
calibration weight value in "gram" that is stored in the
balance. Use the following keys to change the value.
SAMPLE key
RE-ZERO key
PRINT key
MODE key

7. CALIBRATION

Cal 0
.
SAMPLE
key. The display shows the
To select the digit blinking to
change.
To set the value of the digit
selected.
To store the value and return to
step 3.
To cancel the value and return to
step 3.
14
Press and hold
the CAL switch.
Cal
Release
the CAL switch.
Cal 0
01200. 0
Set the weight using
the relevant keys.
Cal 0

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the EK-i Series and is the answer not in the manual?

Questions and answers

Parris
February 27, 2025

6 lines when turned on. Does not go to 0.0 until i press MODE. Cannot tare device display shows -37.6g.

User image 67c0a03e68e8e
1 comments:
Mr. Anderson
February 27, 2025

If the A&D EK-i Series scale displays 6 lines and does not reset to 0.0 until the MODE button is pressed, and shows -37.6g when trying to tare, follow these steps:

1. Ensure that the weighing pan is empty.
2. Turn the power off and then turn it back on.
3. If the issue persists, check for any foreign objects or obstructions around the weighing pan and remove them.
4. If the problem continues, calibrate the balance to maintain accuracy.
5. If calibration does not resolve the issue, request service from the store where the balance was purchased or from a local A&D dealer.

If the error persists, professional service may be required.

This answer is automatically generated

Table of Contents

Save PDF