Charging the Probe
WARNINGS!
• Use only cables, probes, chargers and accessories specified for use with Butterfly iQ™. Substi-
tution of non-approved accessories may cause the system to perform improperly or cause injury
to the patient or operator.
• If the probe seems unusually hot, produces an odor or smoke, or leaks, stop use immediately.
Unplug the probe from the mobile device or disconnect it from the wireless charger (if applica-
ble). Contact support at support.butterflynetwork.com.
• The probe is designed to remain sealed. Do not attempt to open the probe or tamper with the
device internals, including the battery. Doing so may cause injury to the patient or operator.
• The probe battery is not user-replaceable. Replacement of the battery by parties other than But-
terfly Support may result in a hazard such as higher temperatures, fire, or explosion.
• A non-medical grade power supply must be used outside of the patient environment so that it is
at least 1.5 meters from the patient.
CAUTIONS!
• The probe battery should be charged at least monthly to ensure proper functionality.
• If the probe will not power on after charging, it could indicate a battery failure. Contact Support
at support.butterflynetwork.com.
It is important to keep your probe charged. Charge your probe with the supplied battery charging accessories.
The battery charging accessories include the charging pad, charging cable, and wall adapter.
Place the probe on the charger in the orientation shown below
20
Figure 3. 3rd Generation Probe Charger
Charging the Probe