Heat Disinfection Program with CleanCart
Performing
Note
• The program can be interrupted at any time when the program
is running; press and hold the Select key for 3 seconds. The
machine will then interrupt the procedure and finish off with
a drain sequence.
• The pick-up tubes must be rinsed/disinfected separately. See
"Pick-up Tubes" on page 8:41 for instructions.
Check before starting the procedure
Procedure
Use the Keypad next to the Information Display to select options and
to navigate within and between the menus. For details concerning the
keys included in the Key pad, see "Keypad" in chapter 3. part 1
HCEN12239 Revision 12.2010
Program version 3.xx
That the dialysis fluid tubes are connected to the safety couplings
and that the latches of the BiCart holder are closed.
That the concentrate connectors are placed in the corresponding
stand-by ports of the machine (blue and red), and that the yellow
disinfectant connector is placed in the parking port (marked P).
1. Press the Rinse/Disinfection button briefly.
2. The RINSE/DISINFECTION MENU will be displayed on
the Information Display.
3. Select HEAT.
4. Check that the key CleanCart
selected program.
If the key CleanCart is displayed, continue to point 5.
If not, do point 4.a and 4.b before continuing to point 5.
4.a.
4.b.
AK 96
®
Operator's Manual - Hygiene and Maintenance
Select CHANGE.
Select the key CleanCart.
cartridge -
®
Note
is displayed as the
8:13