Care instructions ESTETICA E30
2 Reprocessing methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist element
Sterilization in a steam sterilizer in compliance with DIN
EN 13060
▶ Remove the taper sleeve from the cannula tube weld in sterile goods pack-
aging.
▶ Sterilise with a triple fractionated vacuum (B class steriliser); hold for 4
°
minutes; 134
C ± 1 °C; overpressure: 2.13 bar.
Damage caused by over-heating.
Damage to the cannula.
▶ Immediately remove the parts from the steriliser after the sterilisation cycle
is completed.
Storage
▶ Reprocessed products should be stored protected from dust with minimum
exposure to germs in a dry, dark and cool place.
2.4.5 Preparation of triple function handpiece
Note
Wet the O-rings of the gripping sleeves after sterilisation using silicone
grease (Mat. no. 1.000.6403) or KaVo Rota Spray 2 (Mat. no.
0.411.7510). After replacing the cannula, the air channel must be blown out
before the start of treatment to ensure that the air exiting during treatment
is dry.
Cleaning of triple-function handpiece
Automatic cleaning
Not applicable.
NOTICE
14 / 40