C Commissioning Page - ABB Z-MT Commissioning And Calibration Procedures

Table of Contents

Advertisement

3.1C Commissioning Page

COMMISSIONING
Software Identification
ACJC const
Measure Temperature
or
ACJC Transistor
IC1
Fig. 3.3C Identifying the ACJC Transistor
4
Select the Commissioning Pag .
N o t e . It is possible to access the Electrical Calibration
down this page using Security Access at this point – see Section 4.
Advance to next parameter.
Software Identification
Shows the software type and issue number to be quoted on any
correspondence with the Company.
Advance to next parameter.
Automatic Cold Junction Compensation (ACJC) Constant
& ENT
The cold junction reference temperature (in °C) is shown on the upper display.
If the value is suspect it may be checked as follows:
a) Measure the temperature at the ACJC transistor (TR1) with a
mercury-in-glass thermometer – see Fig. 3.3C.
b) Adjust the displayed value to that of the measured temperature.
N o t e . Adjusting the ACJC constant affects the calibration.
Continued opposite.