MODULE 4: POWER AND CIRCUIT BOARDS
Power and Board Diagnostic Information:
•
Remove SIB fuses with a thin, small tip to prevent holder damage
•
Power OFF to remove or connect any SIB connection to prevent fuse, board, or voltage loss
•
Card Cage Boards:
•
i2000
®
and ci8200
SR
boards that are identical, and therefore, can be swapped for troubleshooting purposes
•
Some older styles of Indexer Boards contain jumpers for configuration prior to installation
•
Indexer Boards contains a 5VDC test point
•
Controller Board contains two reset switches for an alternate method to reboot and initialize
the RSH
•
When installing new controller boards, M&D 6008 Controller Configuration is required for
downloading new bootcode data to board
•
UDB contains switches that must be set during board installation
•
Three motors receive 36VDC:
•
X Motor
•
Transport Carrier Assembly
•
Carrier Positioner Assembly
•
To assess fuse failure on SIB, test each end of fuse to ground or remove and test continuity
4-20
FOR INTERNAL USE ONLY
®
Systems contain multiple Indexer, Motor Driver, and Controller (CPU)
®
ARCHITECT
RSH Service Training Guide