Summary of Contents for Palisades Dental Impact Air 45 9
Page 1
PALISADES DENTAL HIGH SPEED HANDPIECE Care and Use of the Impact Air 45 ® Push Button (Model 9) High Speed Handpiece www.palisadesdental-llc.com...
Page 2
FOR PROFESSIONAL USE ONLY THE IMPACT AIR 45 HIGH SPEED HANDPIECE IS A PRESCRIPTION ® DEVICE AND IS INTENDED FOR USE ONLY BY LICENSED DENTAL PRACTITIONERS. BEFORE OPERATING HANDPIECE, CAREFULLY READ AND FOLLOW THESE INSTRUCTIONS. SAVE THESE INSTRUCTIONS FOR FUTURE REFERENCE. OBSERVE ALL CAUTIONS AND WARNINGS DURING USE.
BEFORE OPERATING THE HANDPIECE, CAREFULLY READ AND FOLLOW ALL OF THESE INSTRUCTIONS. SAVE THESE INSTRUCTIONS FOR FUTURE REFERENCE. TO MINIMIZE RISK OF SERIOUS INJURY, OBSERVE ALL WARNINGS AND CAUTIONS. The Impact Air 45 High Speed Handpiece is an air-powered dental ®...
Page 4
After each use, the handpiece should be cleaned as follows: a. To clean the Exterior of the Handpiece: •Wipe with damp sponge or brush to remove debris. b. To disassemble and clean the Interior of the Head: •Make sure a bur is in place prior to disassembly. •Fit the head cap wrench, provided with the handpiece, into the slots of the head cap.
Page 5
Diagram B). Gently clean the exterior of the chuck with I.P.A. and a small bristle brush. Gently clean the interior of the chuck with I.P.A. and a pipe cleaner. f. To clean the Spindle: •Insert a pipe cleaner dampened with I.P.A. into the spindle hole as far as possible and twist (refer to Diagram B).
Page 6
of the head and tighten with the head cap wrench. Do not force the head cap. If resistance is experienced, refer to the troubleshooting section of this booklet. WARNING: BE SURE TO TIGHTEN HEAD CAP SECURELY WITH HEAD CAP WRENCH SO IT DOES NOT LOOSEN DURING OPERATION.
Page 7
5.Lubrication handpiece should be properly lubricated BEFORE The Impact Air 45 ® each standard autoclaving cycle with Impact Air 45 Autoclavable ® Lubricant. Proper lubrication is the most critical factor in handpiece bearing life. For maximum bearing life, use only Impact Air 45 ®...
Page 8
6.Burs USE COMPATIBLE BURS: The Impact Air 45 handpiece is intended for use with friction-grip burs, ® including surgical length burs up to 25mm in length, that conform to I.S.O. and A.D.A. shank diameter standards (.0625 - .0630 in./1.59-1.60 mm). USE OF INCOMPATIBLE BURS may damage push-button chuck or other parts, and result in bur slippage.
7.Operation •When attaching a standard handpiece to the coupling, be sure that all tubes are fully inserted with the gasket flush to the connector and maintained in position prior to tightening the coupler sleeve/nut onto the handpiece. •A Quick Disconnect handpiece is designed to connect to one of our Quick Disconnect Couplers, Item #627, #628, and #629.
Page 10
•INSPECT interior of head carefully to ensure that all turbine components, such as the two washers and the front, blue O-ring, have been removed. Clean interior of the head with Isopropyl Alcohol (I.P.A.) using a bristle brush and/or swab. After cleaning, ensure that the interior of the head is dry, clean and free of debris.
Page 11
WARNING: BE SURE TO TIGHTEN HEAD CAP SECURELY WITH HEAD CAP WRENCH SO IT DOES NOT LOOSEN DURING OPERATION. OPERATING HANDPIECE WITH LOOSE HEAD CAP COULD CAUSE PERSONAL INJURY. •After installation, proceed with sterilizing the handpiece in accordance with the instructions listed in Section 4 - Sterilization of this booklet. 9.To Replace the Chuck (refer to Diagrams A &...
Page 12
•Remove new chuck from package, being careful not to remove or reposition curved spring washers. Insert end of new chuck into turbine cartridge and push with finger until fully inserted. •Insert the bur back into turbine by placing the turbine – chuck side down –...
Palisades Dental parts ONLY, as descibed in these instructions. If your handpiece requires service, maintenance or repair, send to Palisades Dental ONLY for evaluation at the address on the back of this booklet. DO NOT SEND TO OTHER CENTERS. IMPACT AIR 45 HIGH ®...
Page 14
•Head Cap threads may be cross-threaded. Realign threads and tighten head cap. •If head cap will not tighten, send your handpiece to Palisades Dental ONLY for service. •Handpiece may need to be cleaned. Follow Air Leak Detected cleaning instructions listed in Section 3 - Cleaning of this booklet.
Page 15
Air Leak Detected (cont’d.) •Bearings may be worn or blown. Replace Turbine or return the handpiece to Palisades Dental ONLY for service. •If the air leak persists after replacing the above listed parts and tightening the connections, return the handpiece to Palisades Dental ONLY for service.
Page 16
Align and tighten connections. If your problem cannot be resolved using the Troubleshooting tips, send your handpiece to Palisades Dental ONLY for service at the address listed on the back of this booklet. DO NOT SEND TO OTHER CENTERS. IMPACT AIR 45 ®...
•Please complete the repair card enclosed with these instructions (additional copies can be downloaded from our website at: www.palisadesdental-llc.com), or include a note, with a brief description of the problem you are having, as well as your return address and phone number. 11.Parts and Accessories WE RECOMMEND having a bottle of Impact Air 45 Autoclavable Lubricant, as...
Page 18
•The handpiece must not have been previously altered, repaired or serviced by anyone other than Palisades Dental. •Failure to use ONLY Palisades Dental parts in the handpiece will void the warranty. •Failure to follow the maintenance guidelines in the booklet will void the warranty.
Page 19
PALISADES DENTAL 111 CEDAR LANE ENGLEWOOD, NJ 07631 PHONE: 1-800-664-8000 TEL: 201-569-0050 FAX: 201-569-0250 www.palisadesdental-llc.com Rev. 07/18...