Maintenance & Troubleshooting
Issue: Pump will not rate entry above 500 ml/ hr
Solution: Reset factory default Maximal Rate. The Vet
Pro Pump comes with a factory default Maximal Rate of
500ml.
1. Turn pump ON
2. Press INFO key twice
3. Scroll down with the down-
to ADJUST SETTINGS.
4. Press START/OK key.
5. Scroll down with the down-
to MORE.
6. Press START/OK.
7. Using the numeric key pad, input Technician
Code 901
8. Press START/OK key.
9. Scroll down with the down-
to MAXIMAL RATE.
10. Press START/OK.
11.
Input 1200 ml.
12. Press START/OK key.
13. Scroll down with
keys (2, 0) to
RESTART PUMP. Confirm by
pressing START /OK.
A small numeral 3 will be displayed on the upper
right-hand corner of the LCD Display when the Factory
Defaults have been changed.
arrow key (0)
arrow key (0)
arrow key (0)
arrow