3. Specifications

The Ambu Res-Cue Pump is in conformity with the following standard: EN ISO 10079-2.
The Ambu Res-Cue Pump is in conformity with Council Directive 93/42/EEC
concerning Medical Devices
Vacuum (max): -450 mmHg
Peak free airflow: > 20 l/min
Disposable container volume: 300 ml
External diameter of cap (suction port): 17 mm
External diameter of catheter connection: 13 mm and 6 mm
Stroke Adjustment: Turn stroke adjustment knob 90° counter-clockwise to generate
50% vacuum (of original pressure).
Operating environmental temperature: -20 °C to + 50 °C (-4 °F to 122 °F)
Storage environmental temperature: -40 °C to + 60 °C (-40 °F to 140 °F)
Dimensions: 185 x 70 x 168 mm
Weight: 260 g

4. Description

Ambu Res-Cue Pump is a light portable suction unit that can be powered by one hand
and allows the other hand free for other important duties. This suction unit was designed
to provide simple operation and maintenance. The adjustable stroke knob will provide
different suction pressures.

5. Operating instructions

Select appropriate size tubing (according to the size of the patient) and attach it to the
adapter (if not attached already). Attach the adapter with tubing to the cap of the
container. The container should be attached to the pump.