  • Page 1 English INSTRUCTION MANUAL READ INSTRUCTIONS IN FULL BEFORE USE...
  • Page 2 English Important: This portable baby breathing monitor may not be able to detect all life-threatening events. SNUZA is a registered trademark of Snuza International (Pty) Ltd Unit 11 Roeland Square Roeland Street Cape Town 8001 South Africa Made in South Africa...
  • Page 3: Table Of Contents

    English CONTENTS 1. IMPORTANT – READ FIRST! ...................... 2. Introduction ............................3. Precautions and Contra-indications ..........................................5. Quick Start Guide ..........................6. Diaper (Nappy) Fitment ......................... 7. Fitting your Hero (Medical Device) ................... 8. Operation ............................... 8.1 Startup ..........................8.2 Monitoring .........................
  • Page 4: Important - Read First

    English 1. IMPORTANT – READ FIRST! It is essential that Hero sensor maintains constant contact with baby’s tummy. ALWAYS: Use the correct size diaper (nappy). If it is too loose, Hero may lose contact. Fold the top of the diaper over before attaching Hero Firmly clip Hero clip.
  • Page 5 English Ensure that Hero sits squarely on the diaper, and that the Breathing Sensor (1) is in contact with baby’s tummy. Press the ON/OFF button (6) to start monitoring. Always follow the above procedure to ensure that your baby’s breathing is being monitored, and to prevent false alarms caused by lack of contact with baby’s tummy.
  • Page 6: Introduction

    English 2. Introduction Intended use: Hero is a breathing monitor for the detection of infant breathing cessation. Intended user: This equipment is intended to be operated by the baby’s caregiver. The caregiver must have a good understanding of the instructions and safety measures contained in the user manual.
  • Page 7: Precautions And Contra-Indications

    8.3. Check your baby’s skin for red marks within an hour of attaching Hero . If the skin sensitivity persists, contact us via www. snuza.com. Keep out of reach of pets, pests and other children as damage to the device might cause it to not perform as intended.
  • Page 8 English Front Back Page 5...
  • Page 9: Quick Start Guide

    English Quick Start Guide 6 Diaper/Nappy Fitment). 5.2 Do not switch the monitor on when clipped onto your baby’s diaper. the ON/OFF button (6). 5.4 A start-up test will illuminate the indicator lights, vibrate the stimulator, play 3 audible tones, perform a battery test, and perform a sensor test. 5.5 When the sensor has been tested, a startup tone is heard.
  • Page 10: Diaper/Nappy Fitment

    A very ‘full’ or wet diaper may also cause it to gape. If baby sleeps on his or her side, Hero should be 7. FITTING YOUR HERO When attaching your Snuza monitor you should put your diaper and the baby’s skin. Make sure that the monitor is clipped securely onto the diaper and away from the baby’s skin when clipped...
  • Page 11: Operation

    English OPERA TION Startup Hero must not be switched on while clipped onto your baby’s diaper. The A startup-test will run which will test all indicator lights, the vibration stimulator, the audible alarm and the battery level. If a failure of any item is detected, the Self-Test Failure Alarm will sound and the device cannot be used.
  • Page 12: Monitoring

    English has commenced and Hero can be positioned on the baby’s diaper/nappy (see section 7 Fitting your Hero Warning: Ensure that the battery level is high enough for proper operation (i.e. green or orange light). Otherwise, DO NOT use the monitor. If the battery is too short beep every second.
  • Page 13: Clothing

    English The Alarm Indicator Light (4 Flashes red with If no movement is detected at all for 15 seconds, Hero vibrating stimulus gently vibrates in an attempt to rouse the baby. The Alarm and 5 short beeps Indicator Light (4 seconds, the Breathing Indicator Light (3 each subsequent breathing movement.
  • Page 14: Settings

    English SETTINGS 9.1 AudibleTick This feature causes Hero to emit an audible ‹tick› on each breathing movement. To enable the Audible tick, ensure that Hero is switched ON, then press the MODE/OFF button (7 1 second). To increase the volume of the tick, press the MODE/OFF button once again for medium button a fourth time –...
  • Page 15: Battery

    8 hours per day. 10.3 Replacing the Battery For all MD models: use a CR2 (CR15270) battery available from most retail outlets (please see our website www.snuza.com for information about batteries and details of your national distributor). 10.3.1...
  • Page 16 English 10.3.2 Remove the depleted battery by gently tapping Hero on your hand with the battery cover opening towards the ground. 10.3.3 5. Ensure that the polarity is correct (align the positive end ‘+’ of the battery with the ‘+’ symbol in the battery housing, as indicated). Replace the Battery Cover and ensure that it is correctly closed (Figure 6).
  • Page 17: Replacing The Battery

    12 TROUBLESHOOTING on baby’s diaper (see Section 5Diaper/Nappy Fitmentand Section 7 Fitting your Snuza Hero Battery failure. The Battery Indicator Light (2) will show red (see Section 9.3 Replacing the Battery). Please visit our website www.snuza.comfor further information.
  • Page 18: Disclaimer

    English 13 DISCLAIMER Customers use this monitor at their own risk and Snuza International (Pty) Ltd, its members, employees, agents, contractors, representatives, manufacturers, distributors and associates («SNUZA») shall not be held liable for any loss, or in connection with, the use or manufacture of the monitor, including loss,...