FUNCTION SETTINGS
1. Press [ON/OFF] to turn on the scale, the display will show flash digits and
then --------, press [PRINT], the display will show b xxx BAU value (press
to select among 1200, 2400, 4800, 9600), our default setting is 9600.
[UNIT]
Press [ZERO] to enter Parity mode, the display will show
[UNIT] to select odd,even,none (odd stands for 7 data bits with odd parity, even
stands for 7 data bits with even parity, none stands for 8 data bit without parity)our
default setting is none.
Press [ZERO] to enter Print mode, the display will show n xxx Mod, press [UNIT]
to select ON, OFF, KEY, STB (OFF stands for serial data output disabled, KEY
stands for demand mode by press
scale is stable, ON stands for continuous print, our default setting is STB. Press
[ZERO] again to return to weighing mode.
2. Press [ON/OFF] to turn on the scale, the display will show flash digits and
then ----------, press [UNIT] to enter unit selecting function, the display will show
ON X (X stands for the units: g, oz, lb, dwt...tola...) press
(activate the selected unit) or OFF (inactivate the selected unit), press [ZERO] again
to return to weighing mode.
3. Switch the power to "ON" and than press [ON/OFF] to turn on the scale。The
display shows F-E-9 to 0 and than "- - - - - - -"|,press [PCS] key to enter ZERO
track function setting,display shows ZRD x.xd,press [UNIT] key to select ZERO
track value(0.0d-0.5d-1.0d-1.5d-2.0d); press[ZERO]key to enter ZERO track
function setting,display shows ZRT x.xd,press [UNIT] key to select ZERO track
time(05. to 4.0 seconds; press [ZERO] key to enter ZERO setting,display shows
RZD XXX, press [UNIT] key to select ZERO value(0.8d-1.8d-2.8d-3.8d);press
[ZERO] key to enter FIL setting, press [UNIT] key to select filter range amend
(0-1-2-3-4-5-6-7),"2"is default setting; press [ZERO] key to select zero range when
turn on the scale, press [UNIT] key to select percent (PZR 2, PZR 5, PZR 10.
PZR 20); press [ZERO] key to set zero range, press [UNIT] key to select percent
(KZR 3,4,5,10); press [ZERO] key to set TARE range, press [UNIT] key select
percent (KTR 50 or KTR 100); press [ZERO] key to enter overload protection
setting, press [UNIT] key to select overload protection value (9d OVR-1 OVR-2
OVR-10 OVR); press [ZERO] key to set beep function, press [UNIT] key to choose
turn on or turn off the beep, "b2 ON" means activate the beep, "b2 OFF" means
inactivate the beep; press [ZERO] key to unite the function of TARE and ZREO,
press [UNIT] to switch between them. "ZT ON" means unite function activate,
"ZT OFF" means unite function inactivate. press [ZERO] key to enter auto off
function setting,display shows A xxx,press [UNIT] key to select auto off time
(3,5,10,off). Press [ZERO] to select backlight, the display will show L Aut, press
[UNIT] to select X value (X stands for AUT, OFF, ON), OFF stands for inactivate
backlight function, ON stands for the backlight is always make function. AUTO
stands for auto backlight.
4. Press [ZERO] to return to weighing mode.
P XXX PAR
stands for automatic print when
[PRINT], STB
to select ON
[UNIT]