If the button is not pressed within five seconds of arming, the erase oper-
ation will automatically disarm and memory will be maintained. For this
reason, if you do not intend to erase memory, simply wait until the RED
LED stops blinking at the fast rate.
Erasing the Memory using the Simple Logger II Control Panel:
1. Connect the instrument to the computer, then open the Simple Logger
II Control Panel.
2. Select Erase Memory from the Instrument Menu.
3. A dialog box will appear asking to confirm the erasure of the instru-
ment's memory. Select Yes to confirm or No to cancel the operation.
Erasing the memory will also clear any stored alarms.
4.7
Data Storage
The logger captures Trend measurements.
The following are definitions of terms used in this section:
Input Channel: Source for the measurement channel of the instrument.
Measurement Channel: Measurement of input. This can be a simple
direct measurement, the result of complex mathematical operations on a
single or multiple input, or other channels.
Sample Rate: The rate at which the instrument measures inputs.
Storage Rate: The rate at which channel measurements are stored.
4.7.1
Trend Measurements
The logger stores the measurement of each of the inputs. In addition, the
user can define the storage rate, recording period and measurement format
using the Configure Instrument dialog box in the Simple Logger II Control
Panel software. Trend measurements are stored at this fixed storage rate.
EXAMPLE: AC loggers store TRMS calculation of a single line cycle.
DC loggers store the conversion result of each input.