Apparent Resistance Of Supply Mains; Weight - Imaging Sciences i-Cat 17-19 Operator's Manual

Cone beam volumetric tomography and panoramic dental imaging system
Table of Contents

Advertisement

Apparent Resistance of Supply Mains

For the purpose of obtaining the apparent resistance of supply mains, resistance
is determined according to the following formula:
R =
Where:
U0 is the no-load Mains Voltage
U1 is the Mains Voltage under load.
I1 is the Mains Current under load.
Circuit Breaker
Assembly
100VAC
115VAC
200VAC
230VAC

Weight

Total Weight: 510 lbs. (231.3 kg)
Tube Head Pod: 35.5 lbs. (16.1 kg)
Receptor Pod: 57 lbs. (25.9 kg)
X-Ray Power Supply: 9 lbs. (4.1 kg)
990400 Rev C 2010 May 1
U0 - U1
I1
U0
100.3VAC
96.8VAC
115.4VAC
113.0VAC
200.4VAC
194.1VAC
230.8VAC
223.0VAC
U1
I1
5.22A
4.36A
2.62A
2.31A
Product Information
Apparent
Resistance
0.67ohms
0.55ohms
2.40ohms
3.37ohms
14-3