Error Codes (RH digital) Commissioning Technical data Control elements Standby ® ® RH digital RH basic A Main switch A Main switch B Adjustable safety circuit B Adjustable safety circuit F LED heating plate C Rotary knob, heater G Button heater On/Off D Rotary knob, motor H Button temperature “+“...
Support rod • H 16.1 Extension • H 38 Holding rod • H 44 Boss head clamp • ETS-D Contact thermometer • H 102 Protective cover (RH basic) • H 103 Protective cover (RH digital) See more accessories on www.ika.com...
- unsafe condition of container. 800 mm (min). • Process pathogenic materials only in closed vessels under a suitable fume hood. Please contact IKA ® if you have any questions. Only process media that will not react DANGER dangerously to the extra energy produ- ced through processing.
• Unpacking • Contents of package - Please unpack the device carefully - Magnetic stirrer with heating RH basic or RH digital - In the case of any damage a report must be sent immediately - Mains cable - Operating Instructions (post, rail or forwarder).
(B) to the clockwise stop. lower than the safe temperature limit. • Use the temperature rotary knob (C) [RH basic] or with the Factory setting: about 360 °C. button (H or I) [RH digital] to set the target temperature to the desired ”Safe Temp”...
- Software Version (second displayed segment when the device is on) The device is maintenance-free. It is subject only to the natural - Item and designation of the spare part please see www.ika.com. wear and tear of components and their statistical failure rate.
Failure in safety circuit test - Switch off and switch on again after about 1 minute. ® When fault is indicated again, please contact with IKA service. - Set a higher safe temperature limit. Surface temperature (Temperature of contr- Heating off - Switch off device until the surface temperature of the heating plate is lower than ol sensor) of the heating plate is higher than the the selected safe temperature limit.
Need help?
Do you have a question about the RH basic and is the answer not in the manual?